Synthes TI Matrix Pre-Bent Maxillary Plate / 2mm… recallZ-2272-2015 · Synthes (USA) Products LLC · initiated Jun 24, 2015
Class II Terminated All recalls posted in July 2015
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Synthes (USA) Products LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The affected part and lot numbers of the TI Matrix Pre-Bent Maxillary Plates may potentially have the incorrect laser etch denoting the orientation.
FDA root cause category: Process control
What the company is doing
DePuy Synthes sent an Urgent Notice Medical Device Recall letter dated June 24, 2015, to all affected customers. Customers were asked to check their inventory and return affected devices with the completed response form. Customers with questions were instructed to call 610-719-5450 or contact their DePuy Synthes Sales Consultant.
Recall details
- Recall number
- Z-2272-2015
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Jun 24, 2015
- Posted by the FDA
- Jul 28, 2015
- Recalling firm
- Synthes (USA) Products LLC
- Quantity
- 31
- Product code
- JEY
- FDA submissions
- K083388
- Distribution
- Nationwide Distribution including CA, UT, CO, NJ, WA, VT, KY, WY, CT, NY, MS, and MD.
Product description
Synthes TI Matrix Pre-Bent Maxillary Plate / 2mm ADVMNT/0.8mm THK/L and Synthes TI Matrix Pre-Bent Maxillary Plate / 2mm ADVMNT/0.8mm THK/R. The Synthes Matrix ORTHOGNATHIC Fixation System is intended for use in selective trauma of the midface and craniofacial skeleton; craniofacial surgery; reconstructive procedures; and selective orthognathic surgery of the maxilla, mandible and chin in adults and children.
Code information (models, lots, serial numbers)
Part Numbers: 04.511.381 04.511.382 Lot Numbers: 8672528 8673158 & 8672260
Other devices under the same FDA submission
38 devices cleared under K083388 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Matrix 04.511.225.01Synthes GmbH · PLATE, BONE62 reports
- MatrixMIDFACE 04.503.206.01Synthes GmbH · PLATE, BONE30 reports
- MatrixMIDFACE 04.503.226.01Synthes GmbH · SCREW, FIXATION, INTRAOSSEOUS21 reports
- MatrixMIDFACE 04.503.225.01Synthes GmbH · PLATE, BONE18 reports
- Matrix 04.511.205.01Synthes GmbH · PLATE, BONE18 reports
- Matrix 04.511.206.01Synthes GmbH · PLATE, BONE11 reports
- MatrixMIDFACE 04.503.204.01Synthes GmbH · SCREW, FIXATION, INTRAOSSEOUS8 reports
- MatrixMIDFACE 04.503.224.01Synthes GmbH · PLATE, BONE8 reports
- Matrix 04.511.325Synthes GmbH · PLATE, BONE6 reports
- MatrixMIDFACE 04.503.376Synthes GmbH · PLATE, BONE3 reports
- Matrix 04.511.204.01Synthes GmbH · PLATE, BONE3 reports
- Matrix 04.511.307Synthes GmbH · PLATE, BONE3 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page