Synthes Trauma Ex-Fix System (Small, Medium, Distraction… recallZ-2451-2015 · Synthes (USA) Products LLC · initiated Jul 29, 2015
Class II Terminated All recalls posted in August 2015
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Synthes (USA) Products LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Labeling changes were made related to MR conditions that align with standard definitions for MR Safe, MR Conditional, and MR Unsafe. This recall identifies additional Trauma External Fixation System part numbers that should be considered MR Conditional. One serious injury reported.
FDA root cause category: Labeling design
What the company is doing
Urgent recall letters, dated July 31, 2015, were sent to end users and sales consultants to notifiy them about the issue, the associated risks, and to provide instructions for returning the affected devices and response form.
Recall details
- Recall number
- Z-2451-2015
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Jul 29, 2015
- Posted by the FDA
- Aug 24, 2015
- Recalling firm
- Synthes (USA) Products LLC
- Quantity
- 5306080
- Product code
- KTT
- FDA submissions
- K030390, K031428, K033158, K082650, K122455
- Distribution
- Worldwide Distribution-including US (nationwide) and Canada.
Product description
Synthes Trauma Ex-Fix System (Small, Medium, Distraction Osteogenesis (DO) and Large) ; (S) intended to stabilize and provide treatment for fractures of the small bones, such as the hand, wrist, forearm, foot, and ankle; (M) for the construction of an external fixation frame for the treatment of pediatric and adult fractures; )L) for use to provide treatment for long bone and pelvic fractures that require external fixation; and (DO) for fracture fixation (open and closed),pseudoarthrosis or non-unions of long bones, limb lengthening by epiphyseal or metaphysical distraction, correction of bony or soft tissue deformities and correction of segmental bony or soft tissue defects in adult and pediatric patients.
Code information (models, lots, serial numbers)
Show the full code information (4k characters)
All Lot Numbers of Part Numbers: 390.002 390.003 390.004 390.005 390.007 390.008 390.009 390.010 390.011 390.013 394.79 394.791 394.792 394.793 394.80 394.81 394.82 394.83 394.84 394.85 394.86 394.87 394.88 394.89 394.90 394.91 394.92 03.301.010S 03.301.011S 03.301.012S 03.301.013S 292.40 292.41 390.026 390.027 390.028 390.029 390.030 390.031 390.032 390.033 390.034 390.035 390.036 390.037 395.690 395.691 395.692 395.693 395.779 395.780 395.782 395.784 395.786 395.788 395.792 395.796 395.797 395.798 03.311.010 03.311.011 03.311.012 03.311.013 03.311.015 03.311.020 03.311.022 03.311.023 03.311.024 03.311.025 03.311.031 03.311.032 03.311.033 03.311.036 03.311.037 03.311.038 03.311.041 03.311.042 03.311.043 03.311.046 03.311.047 03.311.048 03.311.050 03.311.051 03.311.054 03.311.055 03.311.056 03.311.058 03.311.059 03.311.060 03.311.061 03.311.061.10 03.311.062 03.311.070 03.311.071 03.311.081 03.311.082 03.311.084 03.311.090 03.311.091 03.311.092 03.311.106 03.311.108 03.311.110 03.311.112 03.311.115 03.311.120 03.311.125 03.311.130 03.311.135 03.311.140 03.311.171 03.311.172 03.311.173 03.311.174 03.311.175 03.311.200 03.311.201 03.311.202 03.311.203 03.311.204 03.311.205 03.311.212 03.311.213 03.311.214 03.311.215 03.311.220 03.311.230 03.311.240 03.311.250 03.311.308 03.311.310 03.311.311 03.311.312 03.311.313 03.311.314 03.311.315 03.311.316 03.311.318 03.311.320 03.311.322 03.311.324 03.311.344 03.311.346 03.311.348 03.311.350 03.311.373 03.311.375 03.311.376 03.311.378 03.311.380 03.311.391 03.311.392 03.311.396 03.311.397 03.311.406 03.311.412 03.311.418 03.311.425 03.311.450 03.311.451 03.311.808 03.311.810 03.311.811 03.311.812 03.311.813 03.311.814 03.311.816 03.311.818 03.311.820 03.311.844 03.311.846 03.311.848 03.311.850 03.311.873 03.311.874 03.311.875 03.311.876 03.311.878 03.311.880 03.311.891 03.311.892 03.311.896 03.311.897 03.311.910 03.311.914 03.311.916 03.311.918 03.311.940 03.311.944 03.311.946 03.311.948 03.311.960 03.311.964 03.311.966 03.311.968 03.311.970 03.311.980 03.311.984 03.311.986 03.311.988 03.311.990 292.75 293.62 293.67 293.72 293.790 293.840 293.890 293.940 294.48 294.52 294.53 294.54 294.55 294.56 294.57 294.65 294.66 294.67 294.68 294.71 294.72 294.73 294.74 294.75 294.76 294.767 294.768 294.769 294.771 294.772 294.773 294.774 294.775 294.776 294.777 294.778 294.779 294.782 294.783 294.784 294.785 294.786 294.788 294.792 294.793 294.794 294.795 294.796 294.798 294.950 494.769 494.771 494.772 494.774 494.775 494.776 494.777 494.778 494.779 494.782 494.783 494.784 494.785 494.786 494.788 494.792 494.793 494.794 494.795 494.796 494.798 294.46SHA 294.52SHA 294.53SHA 294.54SHA 294.55SHA 294.56SHA 294.57SHA 294.67SHA 294.68SHA 294.73SHA 294.74SHA 294.75SHA 294.769SHA 294.76SHA 294.771SHA 294.772SHA 294.775SHA 294.776SHA 294.777SHA 294.778SHA 294.779SHA 294.782SHA 294.783SHA 294.784SHA 294.785SHA 294.786SHA 294.788SHA 294.792SHA 294.793SHA 294.794SHA 294.795SHA 294.796SHA 294.798SHA 494.771SHA 494.772SHA 494.775SHA 494.776SHA 494.777SHA 494.778SHA 494.779SHA 494.782SHA 494.783SHA 494.784SHA 494.785SHA 494.786SHA 494.788SHA 390.041 390.042 395.60 395.61 395.63 395.64 395.65 395.680 395.681 395.682 395.685 395.686 395.687 395.688 390.006 390.012 690.315 390.051 394.055 690.451 03.390.052S 03.390.053S 03.390.054S 03.390.055S 03.390.056S 03.390.057S SD311.202 SD311.203 SD311.204 SD311.205 SD357.100 293.35 293.36 293.40 293.41 293.42 293.44 293.45 293.46 293.47 293.48 293.49 293.50 293.51 293.52 293.53 293.54 293.58 293.59 293.63 293.68 293.69 293.73 293.74 293.78 293.83 293.88 293.93 294.30 294.43 294.44 294.45 294.455 294.46 293.48L 293.48X 294.30SHA 294.43SHA 294.44SHA 294.45SHA 395.62 395.66 395.67 690.978 03.304.220S 03.304.222S 03.304.320S 03.304.322S
Other devices under the same FDA submission
43 devices cleared under K030390, K031428, K033158, K082650, K122455 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- NA 390.005Synthes GmbH · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT21 reports
- NA 390.008Synthes GmbH · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT, METAL COMPOSITE21 reports
- NA 294.56Synthes GmbH · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT, METAL COMPOSITE8 reports
- NA 390.003Synthes GmbH · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT, METAL COMPOSITE8 reports
- NA 394.85Synthes GmbH · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT, METAL COMPOSITE8 reports
- NA 395.126Synthes GmbH · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT8 reports
- NA 395.111Synthes GmbH · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT7 reports
- NA 390.051Synthes GmbH · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT, METAL COMPOSITE6 reports
- NA 394.86Synthes GmbH · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT6 reports
- NA 395.109Synthes GmbH · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT6 reports
- NA 294.786Synthes GmbH · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT5 reports
- NA 395.107Synthes GmbH · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT5 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page