Ortho OS Solutions UltOS 3.5mm locking screw 12mm to 50mm… recallZ-0026-2016 · Ortho Solutions Inc · initiated Jul 14, 2015
Class II Terminated All recalls posted in October 2015
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Ortho Solutions Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Ortho Solutions Inc is initiating a recall on behalf of UltOS 3.5mm locking screw. The locking thread feature of the screw was not manufactured to specification. As a consequence there is a risk that a screw may not lock onto the plate at the time of primary insertion into the plate and the bone. In the event of this occurring the non functioning screw would need to be replaced with an alternativ
FDA root cause category: Nonconforming Material/Component
What the company is doing
The firm, Ortho Solutions Group, sent an "Urgent Medical Device Recall" notifications and "Effectiveness or Recall Declaration Forms" dated 14 July 2015. The notification informed the consignee of the issue, clinical risks, how to identify the affected product, and the actions the consignees were requested to take. Requested actions included reviewing product on-hand, ceasing distribution of affected product and arrangement for product to be returned to Ortho Solutions, Inc. within 2 working days. Consignees were also requested to fill out the Recall Declaration Form and return it to Ortho Solutions, Inc. via fax to: 201-335-0759 by 16 July 2015. If you should have any queries relating to this matter or require any additional information, contact General Manager at 201-906-6175.
Recall details
- Recall number
- Z-0026-2016
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Jul 14, 2015
- Posted by the FDA
- Oct 6, 2015
- Recalling firm
- Ortho Solutions Inc
- Quantity
- 875 units
- Product code
- HWC
- FDA submissions
- K141784
- Distribution
- US Distribution to states of: CA, CO, GA, IL, LA, NC, NY, TX and WA.
Product description
Ortho OS Solutions UltOS 3.5mm locking screw 12mm to 50mm Ortho Solutions LTD, Unit 5 West Station Business Park, Spital Road, Maldon, Essex CM96FF Rx Only a) Sterile b) Non-Sterile Orthopaedic Bone Screw
Code information (models, lots, serial numbers)
Part Number (Lot Number) infomration is as follows: OS421712-NS (lot 7891), OS421714-NS (lot 7891) OS421716-NS (lot 7891, 7520), OS421718-NS (lot 7891, 7520) OS421720-NS (lot 7891), OS421722-NS (lot 7891, 7520) OS421724-NS (lot 7891, 7520), OS421726-NS (lot 7891) OS421728-NS (lot 7891), OS421730-NS (lot 7891) OS421735-NS (lot 7520, 7891) ,OS421740-NS (lot 7520, 7891) OS421745-NS (lot 7520, 7891) ,OS421750-NS (lot 7520, 7891) OS421716 (lot 7519) ,OS421718 (lot 7519)
Other devices under the same FDA submission
5 devices cleared under K141784 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- ORTHO SOLUTIONS ULTOS PLATING SYSTEM OS200125Ortho Solutions Holdings Limited · Bit, Drill4 reports
- ORTHO SOLUTIONS ULTOS PLATING SYSTEM OS292200Ortho Solutions Holdings Limited · Plate, Fixation, Bone1 report
- ORTHO SOLUTIONS ULTOS PLATING SYSTEM OS200025LOrtho Solutions Holdings Limited · Plate, Fixation, Bone1 report
- ORTHO SOLUTIONS ULTOS PLATING SYSTEM OS422702-NSOrtho Solutions Holdings Limited · Plate, Fixation, Bone1 report
- ORTHO SOLUTIONS ULTOS PLATING SYSTEM OS422716-NSOrtho Solutions Holdings Limited · Plate, Fixation, Bone1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.