Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 05055662931105), received between Jan 17, 2020 and Mar 2, 2021 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Fracture” (4 reports).
2 other recalls cover devices cleared under the same FDA submission (K141784); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Fracture4
Patient problems
- Insufficient Information2
- No Clinical Signs, Symptoms or Conditions2
- No Information2
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 2, 2021Fracture; Fracture
REPORTEDLY, DURING A QUADS REPAIR SURGICAL PROCEDURE, THE DRILL BIT FRACTURED DURING USE TO DRILL SOFT BONE. THE DRILL TIP WAS RETAINED WITHIN THE PATIENT, AND WAS UNABLE TO BE RETRIEVED. NO ADDITIONAL COMPLICATIONS WERE REPORTED FOR THE PATIENT AT THE TIME OF REPORT.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 11397749 ↗MalfunctionDec 15, 2020Fracture; Fracture
DRILL BROKEN DURING A TOTAL HIP REPLACEMENT PROCEDURE; THE DRILL TIP COULDN'T BE FOUND AND SUBSEQUENTLY, AN X-RAY WAS CALLED INTO THEATRE TO IDENTIFY THE DRILL TIP. THE FRAGMENT WAS FOUND NEAR THE STEM, SO IT WOULD CAUSE MORE DAMAGE TRYING TO RETRIEVE IT. NO OTHER CLINICAL INFORMATION WAS AVAILABLE.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 11012540 ↗MalfunctionMar 24, 2020Fracture; Fracture
THE DRILL FRACTURED DURING SURGERY. A SECOND DRILL OF THE SAME PRODUCT CODE WAS USED TO COMPLETE THE SURGERY. THE SURGERY WAS COMPLETED SUCCESSFULLY WITH MINOR DELAY CAUSED. MOST OF THE DRILL TIP (THE FLUTED SECTION) REMAINS INSIDE THE PATIENT. THE SURGEON DECIDED NOT TO ATTEMPT TO REMOVE THE BROKEN DRILL SECTION. NO ADDITIONAL PATIENT INJURY OR COMPLICATIONS WERE REPORTED.
Patient problems: No Information
MAUDE report 9871745 ↗MalfunctionJan 17, 2020Fracture; Fracture
IT WAS REPORTED THAT DURING A LEFT TIBIAL TUBERCLE TRANSFER AND PATELLAR STABILISATION PROCEDURE THAT AS THE SURGEON WAS DRILLING THE TIBIA WITH THE DRILL BIT, THE DRILL BIT SNAPPED AS IT CONTACTED THE DISTAL CORTEX OF THE TIBIA. X-RAYS WERE DONE AND THE END OF THE DRILL WAS LEFT IN THE TIBIA AFTER FURTHER ASSESSMENT BY THE CONSULTANT, AS IT WAS IRRETRIEVABLE. THE PATIENT WAS NOTED TO HAVE HARD BONE.
Patient problems: No Information
MAUDE report 9602143 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K141784. One submission often covers many models, so check the product and lot details in each record.
- Z-0230-2026 · Sep 17, 2025 · Open, ClassifiedFuse ULTRA Foot Plating System Instructions for Use. Models: OS200400, OS200401, OS328017, OS200020L, OS200025L, OS200120L, OS200125, OS200125L, OS200135, OS200324, OS200325, OS
- Z-0026-2016 · Jul 14, 2015 · TerminatedOrtho OS Solutions UltOS 3.5mm locking screw 12mm to 50mm Ortho Solutions LTD, Unit 5 West Station Business Park, Spital Road, Maldon, Essex CM96FF Rx Only a) Sterile b) Non-Steril
Other Bit, Drill devices
- QUICKSET 2274-25-500Depuy Orthopaedics, Inc. · BIT, DRILL1,985 reports
- QUICKSET 2274-35-500Depuy Orthopaedics, Inc. · BIT, DRILL770 reports
- QUICKSET 2274-30-500Depuy Orthopaedics, Inc. · BIT, DRILL761 reports
- NA 03.037.022Synthes GmbH · BIT, DRILL339 reports
- NA 323.062Synthes GmbH · BIT, DRILL309 reports
- QUICKSET 2274-56-000Depuy (Ireland) · BIT, DRILL275 reports
- QUICKSET 2274-45-500Depuy Orthopaedics, Inc. · BIT, DRILL211 reports
- LCS 2366-83-000Depuy (Ireland) · BIT, DRILL207 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page