Skip to content

IOL Master 5.5: Software versions 7.5 and 7.7 recallZ-0357-2016 · Carl Zeiss Meditec AG · initiated Oct 27, 2015

Class II Terminated All recalls posted in December 2015

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Carl Zeiss Meditec AG or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

IOL Master software versions 7.5 and 7.7 calculation printouts and exported reports can contain the wrong IOL power data.

FDA root cause category: Software design

What the company is doing

Urgent Medical Device Correction letters will be mailed on Friday, November 13, 2015 as the IOL SW update material is now available. Letter informs customers of the issue and provide instructions for identifying the software version and for installing the software update with illustrations. All customers are requested to complete and return the confirmation form to Stericycle as soon as the upgrade is complete or it has been determined that your unit is not running the affected version of the software.

Recall details

Recall number
Z-0357-2016
Recall class
Class II
Status
Terminated
Initiated by the company
Oct 27, 2015
Posted by the FDA
Dec 1, 2015
Recalling firm
Carl Zeiss Meditec AG
Quantity
49
Product code
HJO
FDA submissions
K101182, K122418, K993357
Distribution
Nationwide Distribution.

Product description

IOL Master 5.5: Software versions 7.5 and 7.7; Ophthalmic: used to obtain ocular measurements and perform calculations to allow physicians to determine appropriate IOL power and type for implantation.

Code information (models, lots, serial numbers)

Software versions 7.5 and 7.7: Serial numbers:  1040306  1045450  1035312  1034011  1035310  1043145  1039588  1041805  1040321  1041803  1045443  1036716  1035318  1038800  1034016  1041838  1041833  1039587  1035291  1040313  1038795  1034051  1036553  1040314  1036539  1036702  1043952  1039614  1036541  1043121  1036786  1036713  1045455  1034801  1034792  1034071  1043117  1034795  1045451  1039611  1043941  1036693  1041836  1036532  1043132  1043134  1034073  1034822  1040315.

Other devices under the same FDA submission

1 device cleared under K101182, K122418, K993357 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 72564).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page