GE Healthcare, Aestiva MRI, 1006-9310-000 recallZ-0679-2016 · GE Medical Systems, LLC · initiated Dec 15, 2015

Class II Terminated All recalls posted in January 2016

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from GE Medical Systems, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The Drive Gas Check Valve of certain Anesthesia products could become stuck in a fixed open position, which could cause pressure to build in the mechanical ventilation cycle. If this issue is left unresolved, it could result in excessive or prolonged pressure in the patient breathing circuit during ventilation potentially resulting in barotrauma.

FDA root cause category: Nonconforming Material/Component

What the company is doing

GE Healthcare sent an Urgent Medical Device Correction letter dated December 15, 2015, to all affected consignees. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction and Contact Information. Customers with questions were instructed to contact GE Healthcare Service at 1-800-437-1171 or their local Service Representative.

Recall details

Recall number
Z-0679-2016
Recall class
Class II
Status
Terminated
Initiated by the company
Dec 15, 2015
Posted by the FDA
Jan 21, 2016
Recalling firm
GE Medical Systems, LLC
Quantity
84
Product code
BSZ
FDA submissions
K050055
Distribution
Worldwide Distribution - US (nationwide including PR except NH, VT, & WY. Internationally to ALBANIA, ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BANGLADESH, BELGIUM, BOLIVIA, BOSNIA, BRAZIL, CAMBODIA, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, CZECH REPUBLIC, DENMARK, ECUADOR, EGYPT, ESTONIA, FINLAND, FRANCE, GEORGIA, GERMANY, CHANA, GIBRALTAR, GREECE, GUATEMALA, HONG KONG, HUNGARY, ICELAND, INDIA, INDONESIA, IRAQ, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KAZAKHSTAN KENYA, REPUBLIC OF KOREA, KUWAIT, LATVIA, LITHUANIA,MALAYSIA, MALDIVES, MARTINIQUE, MEXICO, MOROCCO, NETHERLANDS, NEW ZEALAND, NICARAGUA, NORWAY, OMAN, PAKISTAN, PERU, PHILIPPINES, POLAND, QATAR, REUNION, ROMANIA, RUSSIAN FEDERATION SAUDI ARABIA, SERBIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SPAIN, SRI lanka, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, VENEZUELA, VIETNAM.

Product description

GE Healthcare, Aestiva MRI, 1006-9310-000. This version of the Datex-Ohmeda 7900 ventilator is used in Datex-Ohmeda Aestiva/5 anesthesia systems. It is a microprocessor based, electronically controlled, pneumatically driven ventilator that provides patient ventilation during surgical procedures. The Aestiva/5 MRI anesthesia system provides the functional feature set offered by the conventional Aestiva/5 to the clinician with the added ability to be used in the MR environment.

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

AMTU01048 AMTU01023 AMTU00163 AMTU01050 AMTU00172 AMTU00175 AMTU00176 AMTU01034 AMTU01047 AMTU01055 AMTU01001 AMTU01002 AMTU01021 AMTU01044 AMTU01042 AMTU01015 AMTU00178 AMTU00179 AMTU00162 AMTU01000 AMTU01012 AMTU01041 AMTU00167 AMTU01019 AMTU00177 AMTU01049 AMTU00171 AMTU00164 AMTU01066 AMTU01006 AMTU00168 AMTU01061 AMTU01060 AMTU01056 AMTU01057 AMTU01073 AMTU00154 AMTU00166 AMTU00174 AMTU01005 AMTU01035 AMTU00150 AMTU01024 AMTU01025 AMTU01026 AMTU01027 AMTU01037 AMTU00152 AMTU01040 AMTU01022 AMTU00157 AMTU00158 AMTU00159 AMTU00160 AMTU00170 AMTU01018 AMTU01009 AMTU01032 AMTU01046 AMTU01045 AMTU01020 AMTU01008 AMTU00165 AMTU01013 AMTU01014 AMTU01033 AMTU01016 AMTU01036 AMTU01038 AMTU00151 AMTU00169 AMTU01003 AMTU00161 AMTU01010 AMTU01064 AMTU01011 AMTU01004 AMTU01017 AMTU01031 AMTU01053 AMTU01058 AMTU01059 AMTU01039 AMTU01007

Other devices under the same FDA submission

1 device cleared under K050055 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 72943).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page