Aestiva MRIRecalls & reported problems

Datex-Ohmeda Inc · AESTIVA MRI · Class 2

All models of Aestiva (2)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Aestiva MRI anesthesia machine supplies gas and breathing support during surgery. Many reports describe power problems, including failure to power up or run on battery, and some describe insufficient gas delivery or loss of mechanical ventilation, often found during pre-use checks. The 158 reports include no deaths and no injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
154
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 154 adverse event reports that list this exact device (DI 00840682102339), received between Apr 2, 2019 and Jun 23, 2026 — classified as 154 malfunctions.

The most frequently reported specific device problem is “Power Problem” (33 reports).

26 other recalls cover devices cleared under the same FDA submission (K050055, K172045, K993410); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 4 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 5 reportsApr 2024: 3 reportsMay 2024: 10 reportsJun 2024: 4 reportsJul 2024: 5 reportsAug 2024: 8 reportsSep 2024: 3 reportsOct 2024: 4 reportsNov 2024: 3 reportsDec 2024: 0 reportsJan 2025: 5 reportsFeb 2025: 3 reportsMar 2025: 12 reportsApr 2025: 4 reportsMay 2025: 5 reportsJun 2025: 5 reportsJul 2025: 4 reportsAug 2025: 3 reportsSep 2025: 4 reportsOct 2025: 4 reportsNov 2025: 3 reportsDec 2025: 7 reportsJan 2026: 6 reportsFeb 2026: 8 reportsMar 2026: 8 reportsApr 2026: 3 reportsMay 2026: 9 reportsJun 2026: 3 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports20242025202612
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Power Problem33
  • Insufficient Flow or Under Infusion20
  • Mechanical Problem19
  • Therapeutic or Diagnostic Output Failure16
  • Failure to Power Up15
  • Failure to Run on Battery12
  • Infusion or Flow Problem9
  • Gas/Air Leak7

Patient problems

  • No Clinical Signs, Symptoms or Conditions150
  • No Patient Involvement4

Recent adverse event reports

The 25 most recent of 154 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 23, 2026Insufficient Flow or Under Infusion

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION DURING PRE-USE CHECKOUT RESULTING IN INSUFFICIENT O2. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25598598 ↗
  • MalfunctionJun 19, 2026Power Problem

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN LOSS OF POWER AND SUBSEQUENT LOSS OF MECHANICAL VENTILATION DURING PRE-USE CHECK. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25565226 ↗
  • MalfunctionJun 9, 2026Insufficient Flow or Under Infusion

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN INSUFFICIENT AGENT DELIVERY DURING PRE-USE CHECKOUT THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25450361 ↗
  • MalfunctionMay 27, 2026Therapeutic or Diagnostic Output Failure

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN LOSS OF MECHANICAL VENTILATION DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25301488 ↗
  • MalfunctionMay 22, 2026Failure to Power Up

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN THE DEVICE COULDN¿T POWER UP. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25257047 ↗
  • MalfunctionMay 21, 2026Power Problem

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN POWER SUPPLY. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25250501 ↗
  • MalfunctionMay 21, 2026Electrical /Electronic Property Problem

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN MIXER ISSUE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25250494 ↗
  • MalfunctionMay 21, 2026Failure to Run on Battery

    IT WAS REPORTED THAT THERE WAS A BATTERY FAILURE THAT COULD CAUSE LOSS OF MECHANICAL VENTILATION DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25241506 ↗
  • MalfunctionMay 20, 2026Leak/Splash

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN A LEAK DURING PRE-USE CHECKOUT . THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25231130 ↗
  • MalfunctionMay 18, 2026Mechanical Problem

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION DURING PRE-USE CHECKOUT RESULTING IN LOSS OF VENTILATION. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25202726 ↗
  • MalfunctionMay 15, 2026Power Problem

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN MACHINE POWER FAILURE. THERE WAS NO PATIENT INVOLVEMENT REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25182527 ↗
  • MalfunctionMay 15, 2026Insufficient Flow or Under Infusion

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN INSUFFICIENT O2 DURING A CASE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25181156 ↗
  • MalfunctionApr 29, 2026Therapeutic or Diagnostic Output Failure

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN LOSS OF MECHANICAL VENTILATION DURING A CASE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25035814 ↗
  • MalfunctionApr 9, 2026Therapeutic or Diagnostic Output Failure

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN LOSS OF MECHANICAL VENTILATION DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24827018 ↗
  • MalfunctionApr 9, 2026Electrical /Electronic Property Problem

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION DUE TO FAULTY POWER CORD THAT COULD CAUSE LOSS OF MECHANICAL VENTILATION DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24825312 ↗
  • MalfunctionMar 30, 2026Insufficient Flow or Under Infusion

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN INSUFFICIENT OXYGEN OR FRESH GAS FLOW DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24728572 ↗
  • MalfunctionMar 26, 2026Excess Flow or Over-Infusion

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN DELIVERY OF HIGH N20 DURING A CASE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24700767 ↗
  • MalfunctionMar 20, 2026Insufficient Flow or Under Infusion

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN INSUFFICIENT O2 DURING SERVICE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24648471 ↗
  • MalfunctionMar 13, 2026Failure to Power Up

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN INABILITY OF UNIT TO POWER UP. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24589829 ↗
  • MalfunctionMar 11, 2026Insufficient Flow or Under Infusion

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN INSUFFICIENT O2 DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24568133 ↗
  • MalfunctionMar 9, 2026Power Problem

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN POWER FAILURE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24551595 ↗
  • MalfunctionMar 3, 2026Electrical /Electronic Property Problem

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN STANDBY SWITCH. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24497938 ↗
  • MalfunctionMar 2, 2026Power Problem

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN LOSS OF POWER, AND SUBSEQUENT LOSS OF MECHANICAL VENTILATION DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24483300 ↗
  • MalfunctionFeb 27, 2026Gas/Air Leak

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN A LEAK THAT COULD CAUSE LOSS OF MECHANICAL VENTILATION DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24467704 ↗
  • MalfunctionFeb 25, 2026Excess Flow or Over-Infusion

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN HIGHER N2O DELIVERY DURING SERVICE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24443021 ↗

FDA clearance and classification

Cleared through 510(k) K172045 on Nov 3, 2017 (decision: Substantially Equivalent). Submitted by Datex-Ohmeda, Inc. as “Aestiva 7900, Aestiva MRI, Aestiva 7100, Aestiva 7100 Compact, Aespire 100, Aespire 7100, Aespire 7900, Aespire View, Avance, Avance CS2, Aisys, Aisys CS2”. The FDA review took 120 days. FDA record ↗

Cleared through Special 510(k) K050055 on Feb 4, 2005 (decision: Substantially Equivalent). Submitted by Datex-Ohmeda, Inc. as “Ge Datex-Ohmeda Aestiva/5 Mri Anesthesia System”. The FDA review took 24 days. FDA record ↗

Cleared through 510(k) K993410 on Apr 25, 2000 (decision: Substantially Equivalent). Submitted by Datex-Ohmeda, Inc. as “Dates-Ohmeda Aestiva Smartvent Mri”. The FDA review took 196 days. FDA record ↗

The FDA regulates product code BSZ (gas-machine, anesthesia) under 21 CFR 868.5160 as a Class II device. 21 CFR 868.5160 ↗

The FDA lists this type of device as life-sustaining or life-supporting.

Recalls under the same FDA submission

These recalls involve devices cleared under K050055, K172045, K993410. One submission often covers many models, so check the product and lot details in each record.

  • Z-3192-2026 · Jul 30, 2026 · Open, Classified
    Et Control, Software option for anesthesia gas machine to achieve and maintain targeted end tidal oxygen and anesthetic agents, Field replaceable unit (FRU) part numbers: 1011-8000
  • Z-3189-2026 · Jul 30, 2026 · Open, Classified
    Aisys, Datex-Ohmeda anesthesia system, Part Number 1011-9050-000
  • Z-3191-2026 · Jul 30, 2026 · Open, Classified
    Aisys CS2 with Et Control, Datex-Ohmeda anesthesia system, Part Number 1011-9000-000
  • Z-3190-2026 · Jul 30, 2026 · Open, Classified
    Aisys CS2, Datex-Ohmeda anesthesia system, Part Number 1011-9055-000
  • Z-0082-2025 · Jul 12, 2024 · Open, Classified
    Aisys CS2. This anesthesia gas machine is intended to provide general inhalation anesthesia and ventilatory support to patients and are to be used only by medical professionals tra
  • Z-0084-2025 · Jul 12, 2024 · Open, Classified
    Auxiliary Common Gas Outlet (ACGO) conversion/install kits sold under product names CONV KIT SCGO TO ACGO and ACGO ASSY FIELD INSTL. These kits are not affected specifically, howev
  • Z-0074-2025 · Jul 12, 2024 · Open, Classified
    Aespire 7100, 7100 (Variant), and Aespire 7100 S/5 Protiva configurations. These anesthesia gas machines are intended to provide general inhalation anesthesia and ventilatory suppo
  • Z-0075-2025 · Jul 12, 2024 · Open, Classified
    Aespire 100. This anesthesia gas machine is intended to provide general inhalation anesthesia and ventilatory support to patients and are to be used only by medical professionals t
  • Z-0081-2025 · Jul 12, 2024 · Open, Classified
    Aisys. This anesthesia gas machine is intended to provide general inhalation anesthesia and ventilatory support to patients and are to be used only by medical professionals trained
  • Z-0077-2025 · Jul 12, 2024 · Open, Classified
    Aespire View. This anesthesia gas machine is intended to provide general inhalation anesthesia and ventilatory support to patients and are to be used only by medical professionals
  • Z-0076-2025 · Jul 12, 2024 · Open, Classified
    Aespire 7900. This anesthesia gas machine is intended to provide general inhalation anesthesia and ventilatory support to patients and are to be used only by medical professionals
  • Z-0080-2025 · Jul 12, 2024 · Open, Classified
    Avance CS2 and Avance CS2 Pro. These anesthesia gas machines are intended to provide general inhalation anesthesia and ventilatory support to patients and are to be used only by me
  • Z-0078-2025 · Jul 12, 2024 · Open, Classified
    Avance. This anesthesia gas machine is intended to provide general inhalation anesthesia and ventilatory support to patients and are to be used only by medical professionals traine
  • Z-1389-2022 · Jun 1, 2022 · Terminated
    GE Healthcare Aisys CS2 Anesthesia System, Part number 1011-9050-000; and Avance CS2 and Avance CS2 Pro Anesthesia System, Part number 1009-9050-000.
  • Z-1427-2022 · May 27, 2022 · Open, Classified
    GE Healthcare Avance CS2, Avance CS2 Pro Anesthesia Systems, Part #1009-9050-000; and Aisys CS2 Anesthesia Systems, Part # 1011-9050-000.

…and 11 more.

Common questions about the Aestiva MRI

Has the Aestiva MRI been recalled?

No FDA recall record in our data names this exact device identifier (DI 00840682102339), based on the FDA recall file dated Nov 27, 2026. However, 26 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.

What problems have been reported with the Aestiva MRI?

The FDA's MAUDE database lists 154 adverse event reports for this device, received between Apr 2, 2019 and Jun 23, 2026. In those reports, the most common specific device problems are “Power Problem” (33), “Insufficient Flow or Under Infusion” (20) and “Mechanical Problem” (19). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Aestiva MRI?

None of the reports are classified as deaths, 0 as injuries and 154 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Aestiva MRI FDA approved?

It was cleared through 510(k) K993410 on Apr 25, 2000; its record lists 3 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the Aestiva MRI MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Safe: it poses no known hazards in any MRI environment.

Who makes the Aestiva MRI, and is it still sold?

The FDA's device database lists Datex-Ohmeda Inc as the company responsible for it (brand Aestiva, model MRI). Its record shows it as no longer in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.