GE Healthcare, Optima XR200amx, Mobile Digital Ready… recallZ-0770-2016 · GE Medical Systems, LLC · initiated Dec 23, 2015
Class II Terminated All recalls posted in February 2016
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from GE Medical Systems, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential safety latch mechanism failure. The safety latch, within the column that supports the horizontal arm, is a backup in the unlikely event the primary safety cable breaks. A fall of the arm assembly could result in bodily harm to a person.
FDA root cause category: Device Design
What the company is doing
Consignees were sent on 12/23/2015 a GE Healthcare "Urgent Medical Device Correction" letter GEHC Ref# 10906 dated December 18, 2015. The letter was addressed to Director of Biomedical Engineering, Director of Radiology and Chief of Cardiology. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction and Contact Information. Customers may continue to use the system. The firm will correct the affected devices and no charge to the customer; and a GE Healthcare representative will contact customers to arrange for the correction. For questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Recall details
- Recall number
- Z-0770-2016
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Dec 23, 2015
- Posted by the FDA
- Feb 8, 2016
- Recalling firm
- GE Medical Systems, LLC
- Quantity
- 895
- Product code
- IZL
- FDA submissions
- K142383
- Distribution
- Worldwide Distribution -- US, including Washington D.C. Puerto Rico, Virgin Islands, and Guam; and the countries of Afghanistan, Albania, Anguilla, Antigua and Barbuda, Armenia, Australia, Austria, Bahamas, Bahrain, Belgium, Bermuda, Bolivia, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Denmark, Dominican Rep., Ecuador, Egypt, Ethiopia, Finland, France, Germany, Greece, Guadeloupe, Guatemala, Guyana, Honduras, Hong Kong, Hungary, Iceland, India, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea (Republic of), Kuwait, Lebanon, Malaysia, Martinique, Mexico, Netherlands, New Zealand, Nigeria, Norway, Pakistan, Panama, Papua New Guinea, Paraguay, Philippines, Poland, Portugal, Romania, Russian Federation, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Taiwan, Thailand, Trinidad , Turkey, United Arab Emirates, United Kingdom, Uruguay, Venezuela, and Vietnam.
Product description
GE Healthcare, Optima XR200amx, Mobile Digital Ready Radiographic System Designed to generate diagnostic radiographic images.
Code information (models, lots, serial numbers)
Show the full code information (11k characters)
Lot or serial numbers include, but not limited to: 00001031869WK9, 00001031925WK9, 00000081156HL2, GEGR2170, 00001033726WK9, 00000119120HL4, 00000084104HL9, 00001033405WK0, 00000085648HL4, 00001033548WK7, 1035415WK7, 124175HL1, 122205HL8, 1036207WK7, 00000086387HL8, 00000104902HL2, 00000087904HL9, 00000087902HL3, 00000106045HL8, 00000105791HL8, 00000083607HL2, 00000090594HL3, 00000090597HL6, 00001030061WK4, 00001030140WK6, 00001030142WK2, 114185HL2, 00001031534WK9, 00001031533WK1, 00001031696WK6, 00000092062HL9, 00001028273WK9, 00000106054HL0, 00000106067HL2, 00000105797HL5, 00000106074HL8, 00000105801HL5, 00000104928HL7, 00000093370HL5, 00000089482HL4, 00000085430HL7, 00000091765HL8, 120821HL4, 00000086171HL6, 00000089492HL3, 00000089863HL5, 00000104898HL2, 00000108421HL9, 00000092383HL9, 00000093388HL7, 00001032768WK2, 00000104929HL5, 00000104507HL9, 00000105458HL4, 00000090610HL7, 00000086388HL6, 00000083023HL2, 00000085899HL3, 00000080694HL3, 00000102536HLO, 00000092384HL7, 115224HL8, 00000090608HL1, 00000091221HL2, 00000087939HL5, 00000087912HL2, 00000086382HL9, 00000091420HL0, 00001031277WK5, 00000085023HL0, 00000100184HL1, 00000082246HL0, 00000089865HL0, 00001029027WK8, 00000108416HL9, 00001029021WK1, 00000117417HL6, 00000090227HL0, 00000090593HL5, 00000089495HL6, 00000087096HL4, 00000101153HL5, old-pos, 00000108581HL0, 00000110652HL5, 00000081168HL7, 00000107465HL7, GEGR4022, 1035828WK1, 00000108579HL4, 00000110665HL7, 00000098205HL8, 00000117423HL4, 00000086166HL6, 00000082242HL9, 427332GE2, 00001033782WK2, 119096hl6, 00000120673HL9, 00000083021HL6, 00000085425HL7, 00000086976HL8, 00000086973HL5, 00000094572BT5, 1035839WK8, 00000085428HL1, 120672HL1, 00000091392HL1, 00001028490WK9, 00000115766HL8, 00000118540HL4, 125111BT5, 00000115498HL8, 00000115243HL8, 00000117429HL1, 00000084397HL9, 00000092632HL9, ASTE-51009, 00000117424HL2, 436814GE8, 00000102543HL6, 00001030548WK0, 00000076405HL0, 00000081685HL0, 00000098618HL2, 00000099170HL3, 00000124188HL4, 00000120637HL4, 00000109390HL5, 00000082459HL9, 115234HL7, 121174HL7, 00000121174HL7, 00000095904HL9, 00000084417HL5, 00000084715HL2, 00000129117HL8, 00000079283HL8, 00000093377HL0, 00000087075HL8, 00000091398HL8, 00000077470HL3, 00000127206HL1, 00000127656HL7, 00000090970HL5, 00000106059HL9, 00000091417HL6, 00000082723HL8, 00000091391HL3, 00000115804HL7, 115804HL7, 00000121166HL3, 00000091395HL4, 00000091395-HL, 00000092613HL9, 00001033068WK6, 00001033184WK1, 00000224760HP9, 00000124012HL6, 00000083008HL3, 00000090611HL5, 124188HL4, 00000124172HL8, 00000085894HL4, 113140HL8, 00000099165HL3, 00000085061HL0, 00000103016HL2, 00000107468HL1, 00000092939HL8, 00000084920HL8, 00000082209HL8, 00000087913HL0, 00000084913HL3, 00000104588HL9, 00000110117HL9, 00000092386HL2, 00000103787HL8, 00000086372HL0, 00001033549WK5, 00000367990GE9, 00000224083HP6, 00000091430HL9, 00000105788HL4, 00000122010HL2, 00000083031HL5, 00000102081HL7, 00000103001HL4, 00000103624HL3, 00000084710HL3, 00000089497HL2, 00000111797HL7, 00000118470HL4, 00000089862HL7, 00000115239HL6, 00000091401HL0, 00000081812HL0, 00000090219HL7, 00000090600HL8, 00000085058HL6, 00000083025HL7, 00000103785HL2, 00000104899HL0, 00000093380HL4, 122030HL0, 124198HL3, 00000085900HL9, 00000089484HL0, 00000086391HL0, 00000077913HL2, 00000090216HL3, 00000085001HL6, 00000122237HL1, 00000090224HL7, 00000111923HL9, 00000127665HL8, 110713HL5, 122943HL4, 9102015, 00000109394HL7, 00000086385HL2, 00000085651HL8, 00000092074HL4, 00000081153HL9, 00000103017HL0, 00000106082HL1, 00000083594HL2, 00000085060HL2, 00000091431HL7, 00000081482HL2, GEGR4180, 00000084103HL1, 00000085038HL8, 00001028496WK6, 00000128722HL6, 126695HL6, 00000110102HL1, 1036298WK6, 00000117869HL8, 00000117879HL7, 1036347WK1, 00000083005HL9, 00000090596HL8, 00000079636HL7, 00001029965WK9, 00001029991WK5, 00000098216HL5, 00000085035HL4, 00000082472HL2, 00000083028HL1, 00000089499HL8, 00000098621HL6, 00001029907WK1, 00000090223HL9, 120204HL3, 00000085672HL4, 00000108111HL6, 124225HL4, 00000089867HL6, 00000090975HL4, 00000091400HL2, 00000102078HL3, 00000093375HL4, 00000083542HL1, 00000083535HL5, 00000087087HL3, 00000102535HL2, 00000083004HL2, 00000122257HL9, 00000126924HL0, 00000087910HL6, 00000242375HP4, 00000108521HL6, 00000112082HL3, 00000086371HL2, 00000090971HL3, 00000103006HL3, 125693hl2, 00000093386HL1, 00000090617HL2, 00000102548HL5, 00000093396HL0, 00000084717HL8, 00000092625HL3, 00000091764HL1, 00000091042HL2, 00000103002HL2, 00000077774HL8, 00000090601HL6, 00000090615HL6, 00000118288HL0, 00000091390HL5, 00000091751HL8, 00000092393HL8, 00001033037WK1, 00000126235HL1, 126240HL1, 00000083018HL2, 00000075960HL5, 00000104939HL4, 00000091388HL9, 00000078813HL3, 00000103791HL0, 00000085022HL2, 00000084149HL4, 00000092631HL1, 00000108112HL4, 00000105478HL2, 155565BT5, 00000124479HL7, 00000109399HL6, 00000092619HL6, 00000117420HL0, 00000118469HL6, 00001034445WK5, GEGR2374, 00000098981HL4, 00000106825HL3, 00000124483HL9, 00000090222HL1, 00000089870HL0, 00000087091HL5, 00000091426HL7, 00000085649HL2, 00000099169HL5, 00000108542HL2, 00000111406HL5, 113160HL6, 00000115245HL3, 00000129119HL4, 00000087916HL3, 00000089858HL5, 447838GE4, 126692HL3, 00001033676WK6, 121209HL1, 00000114102HL7, 00000110516HL2, 00000090602HL4, 00001033419WK1, 00000108115HL7, 00000115770HL0, 00000088767HL9, 00000090604HL0, 00000109414HL3, 00000102534HL5, 00000102540HL2, 00000087908HL0, 00001031809WK5, 00000102533HL7, 00000077481HL0, 00000100743HL4, 00000090598HL4, 00000098197HL7, 11111, 433352GE2, GEGR1161, 00000110115HL3, 00001033815WK0, 00001032543WK9, 142890BT3, 00000127669HL0, 00000086376HL1, 00000105466HL7, 124179HL3, 00001033552WK9, 00001033664WK2, 00000103620HL1, 00000087905HL6, 00000126700HL4, 00000110686HL3, 117857HL3, 00000123741HL1, GEGR3739, 00000102547HL7, 114157HL1, 113106HL9, 00000105395BT8, 00000099158HL8, 00000110647HL5, 00000127662HL5, 00000128717HL6, 00000129108HL7, 00000110711HL9, 00000098210HL8, 00000112080HL7, 00000077925HL6, 00000085041HL2, 143250BT9, 00000091404HL4, 00000086977HL6, 1032498WK6, 00000105483HL2, 00000110518HL8, 00000084425HL8, 00000121177HL0, 00000109405HL1, 00000082719HL6, 00000084431HL6, 00000129968BT4, 00001033375WK5, 00000090217HL1, 00000368036GE0, 00000088644HL0, 000000GEGR1831, 00001030910WK2, 00000107470HL7, 00001033118WK9, 00000103018HL8, 00000087917HL1, 00001035261WK5, 00001033783WK0, 00001034203WK8, 00000085643HL5, 00001029480WK9, 00000117870HL6, 00001033042WK1, 00000097748HL8, 00001033078WK5, 00001034960WK3, 00000082722HL0, 00000082481HL3, 00000081503HL5, 113158HL0, UA46929-7, 00000105465HL9, 00000085032HL1, 101081HL8, 00000090225HL4, 00000082470HL6, GEGR3975, 00000083013HL3, 00000081489HL7, 00000082724HL6, 00000083009HL1, 112117HL7, 00000082249HL4, 00000082473HL0, 432199GE8, 124174HL4, 00000090226HL2, 00000083620HL5, 00000084112HL2, 00000080780HL0, 00000087090HL7, 00000089494HL9, 00000091748HL4, 00000082478HL9, 00000081493HL9, 00000082479HL7, 00000083002HL6, 00000087067HL5, 86394HL4, 00000090218HL9, 00001031046WK4, 00000084399HL5, 00000113144HL0, 441284GE7, 00000075948HL0, 93847HL2, 00000077476HL0, 00000112114HL4, 000000GEGR3075, 00000092623HL8, 00000089487HL3, 00001029478WK3, 00001030966WK4, 00000099163HL8, 00000122236HL3, 00001035840WK6, 00001029898WK2, 00000087069HL1, 00000091746HL8, 00000093384HL6, 00001032380WK6, 00001032661WK9, 00000092063HL7, 00000117846HL6, 00000114155HL5, 00000126931HL5, 00000094216BT9, 00000121182HL0, 00000368089GE9, 00000112112HL8, 00000100421HL7, 00000118554HL5, 00000113164HL8, 00000122252HL0, 00000112144HL1, 00000112134HL2, 00001029448WK6, 00000091389HL7, 00000110090HL8, 00000087071HL7, 00000108557HL0, 00000108561HL2, 00000120657HL2, 00000117854HL0, 00000117208HL9, 00000116104BT1, 00000117180HL0, 00000124166HL0, 00000090973HL9, 00000103012HL1, 00000102542HL8, 000000GEGR1014, 00000090230HL4, 00000107462HL4, 00000120190HL4, 00000090233HL8, 00000089857HL7.00000122908HL7, 00000367818GE2, 00000100749HL1, 00000076461HL3, 00001033939WK8, 00001030742WK9, 00001034715WK1, 00001034381WK2, 00001035412WK4, 00000344268GE8, 00001033916WK6, 00000092626HL1, 00000084903HL4, 00001030752WK8, 00000108121HL5, 00001035159WK1, 00001034204WK6, 00000124499HL5, 00000124498HL7, 00000089875HL9, 00000101696HL3, 00001034447WK1, 00000349207GE1, 00000101707HL8, 00000108578HL6, 00000085586HL6, 00000098221HL5, 00000115768HL4, 00000115765HL0, 00000115781HL7, 00000423355GE7, 00000125012HL5, 00000125018HL2, 00000125016HL6, 00000102537HL8, 00000114141HL5, 1036323WK2, 00000102532HL9, 00000104508HL7, 00001028494WK1, 00000115782HL5, 00001031399WK7, 00001031418WK5, 00001035241WK7, 00001035277WK1, 00001035593WK1, 00001031365WK8, 00001033143WK7, 00001028928WK8, 00001034961WK1, 00001030753WK6, 00001030817WK9, 00001030818WK7, 00001031764WK2, 00001028862WK9, 00001029023WK7, 00001030336WK0, 00001030363WK4, 00001035640WK0, 00001030937WK5, 00001031753WK5, 00001033101WK5, 00001029026WK0, 00001030297WK4, 00001035028WK8, 00001030547WK2, 00001032502WK5, 00001032346WK7, 00001032382WK2, 00001032383WK0, 00001033203WK9, 00001030298WK2, 00001031420WK1, 00001031050WK6, 00001031217WK1, 00001031695WK8, 00001027717WK6, 00001033117WK1, 00000109402HL8, 00000103014HL7, 00000095323HL2, 00000082461HL5, 00000097930HL2, 00000092942HL2, 00000092069HL4, 00000122008HL6, 00000092621HL2, 00001029996WK4, 00001029997WK2, 00001029998WK0, 00001029999WK8, 00001030158WK8, 00001030159WK6, 00001030161WK2, 00001030163WK8, 00001030164WK6, 00001030165WK3, 00001030166WK1, 0001030201WK6,, 00001028226WK7, 00001030623WK1, 00001030263WK6, 00000093372HL1, 00001032505WK8, 00000090231HL2, 00000095901HL5, 00000107461HL6, 00000111795HL1, 00000084157HL7, 00000115500HL1, 00000126225HL2, 00000111426HL3, 00000103015HL4, 00000113113HL5, 00000112126HL8, 00000126703HL8, 00000126230HL2, 00000126231HL0, 00000126232HL8, 00000112152HL4, 00000093385HL3, 00000122248HL8, 00000098614HL1, 00000110655HL8, 00000091050HL5, 00000126704HL6, 00000091405HL1, 00000093381HL2, 00000093382HL0, 00000101687HL2, 00000088760HL4, 00000087903HL1, 00000081806HL2, 00000108585HL1, 00000108584HL4, 00000106693HL5, 00000105476HL6, 00000105790HL0, 00000105787HL6, 00000106070HL6, 00000082236HL1, 00000086975HL0, 00000085667HL4, 00000100432HL4, 00000089486HL5, 00000077770HL6, 00000119116HL2, 00000085657HL5, 00000086847HL1, 00000108119HL9, 00000108528HL1, 00000084405HL0, 00000109417HL6, 00000082464HL9, 00000120649HL9, 00001035441WK3, 00000092398HL7, 00000092397HL9, 00000085037HL0, 00000081686HL8, 00000081157HL0, 00000099159HL6, 00000082732HL9, 00000102069HL2, 00000914006HL9, 00000116634HL7, 00000115775HL9, 00000116625HL5, 00000106692HL7, 00000086162HL5, 00000099166HL1, 00000126927HL3, 00001035264WK9, 00000100199HL9, 00000101299HL6, 00000104506HL1, 00000124178HL5, 00000104509HL5, 00000115238HL8, 00000120179HL7, 00000098906HL1, and 00001034958WK7.
Other devices under the same FDA submission
2 devices cleared under K142383 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Optima XR200amxGE Medical Systems, L.L.C. · System, x-ray, mobile2 recalls1 report
- Optima XR220amxGE Medical Systems, L.L.C. · System, x-ray, mobile2 recalls0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 72974).
- Z-0768-2016 · TerminatedGE Healthcare, Optima XR220amx, Mobile Digital Ready Radiographic System Designed to generate diagnostic radiographic images.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page