INFX-8000V Bi-Plane X-Ray Interventional System recallZ-0752-2016 · Toshiba American Medical Systems Inc · initiated Aug 14, 2015

Class II Terminated All recalls posted in February 2016

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Toshiba American Medical Systems Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

When a fontal DA (Digital Angiography) acquisition was done, scattered x-ray came into the dose meter on the lateral side. As a result, the dose meter sent a minus value to the software. Consequently, the software defined the data as an "abnormal value" and it stopped displaying dose data and the dose data was lost. The following message was displayed, "Dose meter abnormal, Dose info disabled".

FDA root cause category: Radiation Control for Health and Safety Act

What the company is doing

Toshiba's planned action to bring into compliance: 1.Affected customers will be notified using the Customer Notification Letter you provided, 2. revised system software will be installed on the systems to prevent this occurrence from happening, 3. the corrections will be made free of charge, and 4. the corrections will be completed by April 4, 2016. For further questions, please call 1- 800 521-1968.

Recall details

Recall number
Z-0752-2016
Recall class
Class II
Status
Terminated
Initiated by the company
Aug 14, 2015
Posted by the FDA
Feb 9, 2016
Recalling firm
Toshiba American Medical Systems Inc
Quantity
4
Product code
OWB
FDA submissions
K120073, K133535, K143225, K152696
Distribution
US Nationwide Distribution to OH and NY

Product description

INFX-8000V Bi-Plane X-Ray Interventional System X-ray systems

Code information (models, lots, serial numbers)

Serial Numbers ACA1492002, AAA1492001, W4B1442020, and W4B1432018.

Other devices under the same FDA submission

1 device cleared under K120073, K133535, K143225, K152696 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.