Skip to content

Product 4 consists of all products under product code HWC… recallZ-1102-2016 · Zimmer Manufacturing B.V. · initiated Jan 11, 2016

Class II Terminated All recalls posted in March 2016

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Zimmer Manufacturing B.V. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.

FDA root cause category: Packaging

What the company is doing

Zimmer Biomet sent an Urgent Medical Device Recall Letter, dated January 11, 2016, to Distributors, Sales Representatives, Operation Managers, and Risk Managers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to 1.Review the notification and ensure that relevant personnel are aware of the contents. 2. Assist your Zimmer Biomet sales representative with the quarantine of any affected product. 3. Your Zimmer Biomet sales representative will remove the recalled product from your facility. 4. Complete and return the attached Certificate of Acknowledgment form to [email protected]. 5. If after reviewing this notification you have further questions or concerns please call Customer Service at 1-800-348-2759, or contact your Zimmer Biomet Sales Representative.

Recall details

Recall number
Z-1102-2016
Recall class
Class II
Status
Terminated
Initiated by the company
Jan 11, 2016
Posted by the FDA
Mar 12, 2016
Recalling firm
Zimmer Manufacturing B.V.
Quantity
154
Product code
HWC
FDA submissions
K112885
Distribution
US Nationwide Distribution

Product description

Product 4 consists of all products under product code HWC, and same usage: Item no: 47482002603 2.0 X 26 CORT SCREW CRUCI 47482002803 2.0 X 28 CORT SCREW CRUCI 47482003203 2.0 X 32 CORT SCREW CRUCI 47482003403 2.0 X 34 CORT SCREW CRUCI 47482003603 2.0 X 36 CORT SCREW CRUCI 47482003803 2.0 X 38 CORT SCREW CRUCI 47482004003 2.0 X 40 CORT SCREW CRUCI 47482004203 2.0 X 42 CORT SCREW CRUCI 47482004403 2.0 X 44 CORT SCREW CRUCI 47482004603 2.0 X 46 CORT SCREW CRUCI 47482004803 2.0 X 48 CORT SCREW CRUCI 47482005003 2.0 X 50 CORT SCREW CRUCI 47482702000 2.7 X 20 CORT SCREW HEX Product Usage: A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Code information (models, lots, serial numbers)

lot no.61951256 62230459 61907760 61907764 61926199 62261661 61926201 61907767 61907769 61907771 61909381 62230462 61907772 61196133 61907773 62020888 61920808

Other devices under the same FDA submission

3 devices cleared under K112885 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 72900).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page