Skip to content

Product 61 consists of all product under product code: JWH… recallZ-1159-2016 · Zimmer Manufacturing B.V. · initiated Jan 11, 2016

Class II Terminated All recalls posted in March 2016

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Zimmer Manufacturing B.V. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.

FDA root cause category: Packaging

What the company is doing

Zimmer Biomet sent an Urgent Medical Device Recall Letter, dated January 11, 2016, to Distributors, Sales Representatives, Operation Managers, and Risk Managers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to 1.Review the notification and ensure that relevant personnel are aware of the contents. 2. Assist your Zimmer Biomet sales representative with the quarantine of any affected product. 3. Your Zimmer Biomet sales representative will remove the recalled product from your facility. 4. Complete and return the attached Certificate of Acknowledgment form to [email protected]. 5. If after reviewing this notification you have further questions or concerns please call Customer Service at 1-800-348-2759, or contact your Zimmer Biomet Sales Representative.

Recall details

Recall number
Z-1159-2016
Recall class
Class II
Status
Terminated
Initiated by the company
Jan 11, 2016
Posted by the FDA
Mar 12, 2016
Recalling firm
Zimmer Manufacturing B.V.
Quantity
39
Product code
JWH
FDA submissions
K933785
Distribution
US Nationwide Distribution

Product description

Product 61 consists of all product under product code: JWH and same usage: Item no: 598003712 STD PRECOAT TIBIA PLATE S for use in total knee arthroplasty

Code information (models, lots, serial numbers)

lot no.: 61678240 61838251 61838252 61838253 61838254 61901158 61901159 61925016 61926797 61958844 61958845 61999206 61999207 62037903 62037904 62037905 62070148 62070149 62141260 62170961 62191179 62220224 62232384 62245104 62284609 62298550 62298551 62313021 62319305 62326309 62359946

Other devices under the same FDA submission

89 devices cleared under K933785 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 72900).