Product 47 consists of all product under product code: LPH… recallZ-1145-2016 · Zimmer Manufacturing B.V. · initiated Jan 11, 2016
Class II Terminated All recalls posted in March 2016
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Zimmer Manufacturing B.V. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.
FDA root cause category: Packaging
What the company is doing
Zimmer Biomet sent an Urgent Medical Device Recall Letter, dated January 11, 2016, to Distributors, Sales Representatives, Operation Managers, and Risk Managers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to 1.Review the notification and ensure that relevant personnel are aware of the contents. 2. Assist your Zimmer Biomet sales representative with the quarantine of any affected product. 3. Your Zimmer Biomet sales representative will remove the recalled product from your facility. 4. Complete and return the attached Certificate of Acknowledgment form to [email protected]. 5. If after reviewing this notification you have further questions or concerns please call Customer Service at 1-800-348-2759, or contact your Zimmer Biomet Sales Representative.
Recall details
- Recall number
- Z-1145-2016
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Jan 11, 2016
- Posted by the FDA
- Mar 12, 2016
- Recalling firm
- Zimmer Manufacturing B.V.
- Quantity
- 30
- Product code
- LPH
- FDA submissions
- K934765
- Distribution
- US Nationwide Distribution
Product description
Product 47 consists of all product under product code: LPH and same usage: Item no: 611004022 MODULAR CUP 10 DEGREE LIN 611004428 MODULAR CUP 10 DEGREE LIN 611004628 MODULAR CUP 10 DEGREE LIN 611004828 MODULAR CUP 10 DEGREE LIN 611004832 MODULAR CUP 10 DEGREE LIN 611005028 MODULAR CUP 10 DEGREE LIN 611005032 MODULAR CUP 10 DEGREE LIN 611005628 MODULAR CUP 10 DEGREE LIN 611005632 MODULAR CUP 10 DEGREE LIN 611005828 MODULAR CUP 10 DEGREE LIN 611005832 MODULAR CUP 10 DEGREE LIN 611006028 MODULAR CUP 10 DEGREE LIN 611006032 MODULAR CUP 10 DEGREE LIN 611006228 MODULAR CUP 10 DEGREE LIN 611006232 MODULAR CUP 10 DEGREE LIN 611006428 MODULAR CUP 10 DEGREE LIN 611006432 MODULAR CUP 10 DEGREE LIN 611006628 MODULAR CUP 10 DEGREE LIN 611006832 MODULAR CUP 10 DEGREE LIN 611007028 MODULAR CUP 10 DEGREE LIN 612004626 MODULAR CUP 20 DEGREE LIN 612004628 MODULAR CUP 20 DEGREE LIN 612004828 MODULAR CUP 20 DEGREE LIN 612004832 MODULAR CUP 20 DEGREE LIN 612005028 MODULAR CUP 20 DEGREE LIN 612005032 MODULAR CUP 20 DEGREE LIN 612005628 MODULAR CUP 20 DEGREE LIN 612005828 MODULAR CUP 20 DEGREE LIN 612005832 MODULAR CUP 20 DEGREE LIN 612006028 MODULAR CUP 20 DEGREE LIN 612006032 MODULAR CUP 20 DEGREE LIN 610504428 MODULAR CUP NEUTRAL LINER 610504832 MODULAR CUP NEUTRAL LINER 610505028 MODULAR CUP NEUTRAL LINER 610505032 MODULAR CUP NEUTRAL LINER 610505632 MODULAR CUP NEUTRAL LINER 610505832 MODULAR CUP NEUTRAL LINER 610506032 MODULAR CUP NEUTRAL LINER 610506232 MODULAR CUP NEUTRAL LINER For use in total hip arthroplasty
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
lot no.: 61890322 62213292 62215053 62277522 62402117 61921671 62207256 62212215 62241126 62251893 62268498 62281865 62396132 62402115 62024672 62176459 62210646 62215051 62221907 62241128 62241129 62266480 62277523 62402113 62402114 62406796 62202293 62218187 62255246 62281866 62354788 62396145 62406802 61882636 61890326 61942630 61943368 61957706 61957708 61957709 61968019 61983748 61983749 62109144 62109145 62143889 62163554 62173374 62176463 62188368 62197359 62197360 62197361 62204040 62204041 62204042 62204632 62204633 62204634 62207259 62210647 62210648 62213295 62221909 62221910 62226817 62233573 62238101 62241131 62241135 62241136 62241137 62241139 62241142 62241144 62241145 62251896 62251897 62251898 62251899 62251901 62255247 62266481 62266482 62277524 62284060 62352923 62358047 62383751 62383752 62383753 62383754 62386984 62386985 62402212 62402217 61882637 61885966 61889305 61976198 62147389 62170705 62191282 62197358 62199480 62199481 62210649 62210650 62218188 62218189 62218190 62236577 62236578 62241147 62251895 62258978 62260179 62260180 62352924 62358045 62388057 62402126 62402213 61890327 61968020 62170708 62188373 62207257 62207258 62226818 62241149 62241150 62251902 62258976 62268506 62383758 62383759 62396119 62396120 62396121 61980814 62170709 62202294 62202295 62236582 62258974 62260181 62260182 62352926 62358046 62383761 62396133 61976203 62170710 62176464 62191285 62193578 62221912 62226823 62241152 62250619 62251903 62265335 62354787 62396122 62396123 62403713 62406785 61885968 61980816 62167449 62198732 62202296 62218192 62241154 62266483 62354789 62389394 62389395 62402214 61919746 62170712 62207260 62226819 62226820 62251112 62265336 62383760 62402140 62406770 61882639 61980817 62167448 62218193 62266484 62396135 62403714 62406808 61882640 62199479 62233574 62258975 62396149 61980818 62238099 62354790 62402121 62406806 62221914 62402116 62255245 62402122 62268507 62406810 62188377 62258972 62260183 62402118 62183219 62193583 62354791 62406803 61815148 62080364 62154619 62180252 62193580 62193581 62226821 62226822 62241155 62241156 62251113 62268508 62316411 62383755 62383757 62396134 62402112 62406794 62402123 62213294 62215054 62396142 62406786 62354792 62406804 62233572 62406807 62163545 62395778 62202292 62241114 62198725 62226816 62241115 62268495 62389251 62389252 62198722 62218184 62268496 62395770 62406805 62218185 62277521 62395775 62406812 62395776 62281868
Other devices under the same FDA submission
8 devices cleared under K934765 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Trilogy® 6200-56-22Zimmer, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED5 reports
- Trilogy® 6200-50-22Zimmer, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED3 reports
- Trilogy® 6200-52-22Zimmer, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED2 reports
- Trilogy® 6200-60-22Zimmer, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED2 reports
- Trilogy® 6200-54-22Zimmer, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 report
- Trilogy® 6200-62-20Zimmer, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 report
- Trilogy® 6200-66-22Zimmer, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 report
- Trilogy® 6105-50-32Zimmer, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 72900).
- Z-1101-2016 · TerminatedProduct 3 consists of all products under product code JDI, and same usage: Item no:781803800 12/14 UNIPOLAR FEMORAL HE 781804000 12/14 UNIPOLAR FEMORAL HE 781804100 12/14 UNIPOLAR …
- Z-1102-2016 · TerminatedProduct 4 consists of all products under product code HWC, and same usage: Item no: 47482002603 2.0 X 26 CORT SCREW CRUCI 47482002803 2.0 X 28 CORT SCREW CRUCI 47482003203 2.0 X …
- Z-1103-2016 · TerminatedProduct 5 consists of all products under product code HWC, and same usage: Item no: 47482701001 2.7MM CORT. SCREW 10MM, S 47482701201 2.7MM CORT. SCREW 12MM, S 47482701401 2.7M…
- Z-1105-2016 · TerminatedProduct 7 consists of all products under product code HWC, and same usage: Item no: 47234801435 3.5MM CORT SCR X 14MM 47234801635 3.5MM CORT SCR X 16MM 47234806535 3.5MM CORT SCR…
- Z-1106-2016 · TerminatedProduct 8 consists of all products under product code HWC, and same usage: Item no: 47483501201 3.5MM CORT. SCREW 12MM LN 47483501401 3.5MM CORT. SCREW 14MM LN 47483501601 3.5MM C…
- Z-1108-2016 · TerminatedProduct 10consists of all product under product code HWC, and same usage: Item no: 47484502601 4.5 X 26 CORT SCREW SELF Product Usage: A smooth or threaded metallic bone fixat…
- Z-1109-2016 · TerminatedProduct 11 consists of all product under product code JDI , and same usage: Item no: 902602500 6 DEGREE COCR FEM HEAD 26 902602600 6 DEGREE COCR FEM HEAD 26 902602700 6 DEGREE COC…
- Z-1110-2016 · TerminatedProduct 12 consists of all product under product code JDI , and same usage: Item no: 902602935 6 DEGREE COCR FEM HEAD 28 902602835 6 DEGREE COCR FEM HEAD 28 902603235 6 DEGREE COC…
- Z-1111-2016 · TerminatedProduct 13 consists of all product under product code HWC , and same usage: Item no: 47486510002 6.5 X 100 CANC SCREW, 32M Product Usage: A smooth or threaded metallic bone fixat…
- Z-1114-2016 · TerminatedProduct 16 consists of all product under poduct code: KWY and same usage: Item no: 500103822 BIPOLAR LINER 38MM OD X 2 500103922 BIPOLAR LINER 39MM OD X 2 500104022 BIPOLAR LINER…
- Z-1115-2016 · TerminatedProduct 17 consists of all product under product code: LPH and same usage: Item no: 625004515 BONE SCREW 4.5X15 SELF-TA 625004520 BONE SCREW 4.5X20 SELF-TA 625004525 BONE SCREW …
- Z-1116-2016 · TerminatedProduct 18 consists of all product under product code: HWC and same usage: Item no: 47248001441 CANC SCR 4.0MM X 14MM FUL 47248001641 CANC SCR 4.0MM X 16MM FUL 47248001840 CANC S…
- Z-1119-2016 · TerminatedProduct 21 consists of all product under product code: JDI and same usage: Item no: 811400218 CPT 12/14 COCR REVISION S 811400318 CPT 12/14 COCR REVISION S 811400420 CPT 12/14 CO…
- Z-1120-2016 · TerminatedProduct 22 consists of all product under product code: JDI and same usage: Item no: 811400218 CPT 12/14 COCR REVISION S 811400318 CPT 12/14 COCR REVISION S 811400420 CPT 12/14 C…
- Z-1123-2016 · TerminatedProduct 25 consists of all product under product code: LPH and same usage: Item no: 234703104 DISTAL MEDIAL FEMORAL CON 234703206 DISTAL MEDIAL FEMORAL CON 234703208 DISTAL MEDIAL…
- Z-1126-2016 · TerminatedProduct 28 consists of all product under product code: JDI and same usage: Item no: 32902605000 FEMORAL HD 28MM MEDIUM 32902604000 FEMORAL HD 28MM SHORT 32902604035 FEMORAL HD 28M…
- Z-1127-2016 · TerminatedProduct 29 consists of all product under product code: LPHand same usage: Item no: 32902603800 FEMORAL HEAD +0.5 X 22 MM 32902604810 FEMORAL HEAD +3.5 X 22 MM 32902604845 FEMORAL…
- Z-1131-2016 · TerminatedProduct 33 consists of all product under product code: JDI and same usage: Item no: 662406515 HGP II ACETABULAR CUP BON 662406520 HGP II ACETABULAR CUP BON 662406525 HGP II ACETA…
- Z-1133-2016 · TerminatedProduct 35 consists of all product under product code: JWH and same usage: Item no: 522007300 I/B II KNEE FLUTED ROD, 1 for use in total knee arthroplasty
- Z-1137-2016 · TerminatedProduct 39 consists of all product under product code: JDI and same usage: Item no: 32902605800 ION IMPLANTED FEM 22MM LO 32902604900 ION IMPLANTED FEM 26MM ME 32902603900 ION IM…
- Z-1146-2016 · TerminatedProduct 48 consists of all product under product code: JWH and same usage: Item no: 597206526 NEXGEN ALL-POLY PATELLA, for use in total knee arthroplasty
- Z-1147-2016 · TerminatedProduct 49 consists of all product under product code: JWH and same usage: Item no: 597206529 NEXGEN ALL-POLY PATELLA, 597206532 NEXGEN ALL-POLY PATELLA, 597206535 NEXGEN ALL-POL…
- Z-1149-2016 · TerminatedProduct 51 consists of all product under product code: JWH and same usage: Item no: 598603701 NEXGEN OPTION STEMMED TIB 598603702 NEXGEN OPTION STEMMED TIB 598604701 NEXGEN OPTIO…
- Z-1151-2016 · TerminatedProduct 53 consists of all product under: product code: JWH and same usage: Item no: 597002501 NEXGEN PRECOAT PEG TIB PL 597002502 NEXGEN PRECOAT PEG TIB PL 597003501 NEXGEN PR…
- Z-1152-2016 · TerminatedProduct 54 consists of all product under product code: JWH and same usage: Item no: 596009900 NEXGEN TAPER PLUG for use in total knee arthroplasty
- Z-1154-2016 · TerminatedProduct 56 consists of all product under product code: HWC and same usage: Item no: 47234702032 PERI SCR 3.5MM X 32MM STE 47234702018 PERI SCR. 3.5MM X 18MM ST 47234702020 PERI SC…
- Z-1157-2016 · TerminatedProduct 59 consists of all product under product code: JWH and same usage: Item no: 42540000026 PSN ALL POLY PAT PLY 26MM 42540000029 PSN ALL POLY PAT PLY 29MM 42540000032 PSN ALL…
- Z-1159-2016 · TerminatedProduct 61 consists of all product under product code: JWH and same usage: Item no: 598003712 STD PRECOAT TIBIA PLATE S for use in total knee arthroplasty
- Z-1161-2016 · TerminatedProduct 63 consists of all product under product code: LPH and same usage: Item no: 620004020 TRILOGY ACET SHELL 40MM O 620004220 TRILOGY ACET SHELL 42MM O 620004420 TRILOGY ACET …
- Z-1162-2016 · TerminatedProduct 64 consists of all product under product code: JDI and same usage: Item no: 902605101 UNIPOLAR ENDO FEM HEAD 51MM For use in hemi hip arthroplasty
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.