Intra operative MRI Systems: Neuro II-SE , Neuro III-SV… recallZ-1098-2016 · Deerfield Imaging · initiated Feb 10, 2016

Class II Terminated All recalls posted in March 2016

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Deerfield Imaging or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Emergency helium venting lines (i.e. quench lines) may not have been properly installed.

FDA root cause category: Device Design

What the company is doing

Consignees were sent a IMRIS "Urgent Field Safety Notice" letter dated February 10, 2016. The letter described the problem and the product involved in the recall. The firm also stated that it is important that customers review the portion of the quench line vent described and ensure that it is installed to the specifications supplied as part of the Planning Guide. Customers were reminded that it is important to have an emergency plan in place in the event Helium gas escapes into the magnet room or any other area. For questions customers can contact their local customer support hotline or +1-866-475-0525.

Recall details

Recall number
Z-1098-2016
Recall class
Class II
Status
Terminated
Initiated by the company
Feb 10, 2016
Posted by the FDA
Mar 11, 2016
Recalling firm
Deerfield Imaging
Quantity
35 USA, 14 OUS
Product code
LNH
FDA submissions
K061916, K071099, K083137, K133692
Distribution
Worldwide Distribution -- US, including the states of CA, CT, FL, GA, IL, IN, MD, MA, MN, MO, NH, NC, OH, OR, PA, TX, UT, VA, and WI; and ,countries of CHINA, CANADA, AUSTRALIA, FRANCE, JAPAN, and GERMANY.

Product description

Intra operative MRI Systems: Neuro II-SE , Neuro III-SV, iMRI 1.5T A, and iMRI 3T S. Indicated for use as a magnetic resonance diagnostic device (MRDD).

Code information (models, lots, serial numbers)

70036, 10076-21, M00004, 10073-01, 23529, 40481, M00152, M00005, 30445, M00169, 060014, 10043-00, 30823, 10004680, 10050-01, 40662, 30935, 31084, 31203, 10002221, 40510, 40613, 30719, 31046, 30910, 30832, 30623, 31213, 10004152, 10072-01, 31250, 40815, 31049, 30068, 40412, M0038, M00106, M00367, 10035-01, 40475, 40495, 10007498, 10005946, 40263, 40688, 10058-00, 40619, 31085, and 10005639.

Other devices under the same FDA submission

7 devices cleared under K061916, K071099, K083137, K133692 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.