Battery Pack Used with the Capnostream"20 and… recallZ-1577-2016 · Medtronic · initiated Apr 13, 2016
Class I Terminated All recalls posted in May 2016
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Medtronic Respiratory & Monitoring Solutions Announces a Voluntary Recall of a Battery Pack Used with the Capnostream"20 and Capnostream"20p Patient Monitor Due to the Battery Pack Might Experience Thermal Runaway with the Potential of Fire Risk to the Monitor.
FDA root cause category: Process change control
What the company is doing
Medtronic sent an Urgent Medical Device Recall letter dated April 2016, by postal mail or courier to all affected customers. The letter instructed users to remove and properly dispose of current batteries. If product was further distributed customers were instructed to forward the recall letter. Customers were asked to completed and return the Acknowledgement and Receipt Form by fax or email. Customers with questions should contact Technical Services at 800-635-5267, option 1, ten 1 and 1 or by email at [email protected]. For questions regarding this recall call 303-876-8859.
Recall details
- Recall number
- Z-1577-2016
- Recall class
- Class I (most serious)
- Status
- Terminated
- Initiated by the company
- Apr 13, 2016
- Posted by the FDA
- May 12, 2016
- Recalling firm
- Medtronic
- Quantity
- 9,817 batteries (8,170 with monitors and 1,647 as spare parts)
- Product code
- CCK
- FDA submissions
- K123690
- Distribution
- Nationwide Distribution to AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY. Product was also distributed worldwide. There is 1 VA customer.
Product description
Battery Pack Used with the Capnostream"20 and Capnostream"20p Patient Monitor, Battery Packs Model Number - 016400, 010520. Battery Pack used with the Capnostream"20 and Capnostream"20p Patient Monitor. The battery pack powers the monitor when AC power is not available or when monitoring is required during intra-hospital transfer of patients. The Capnostream"20 and Capnostream"20p is a combined capnograph/pulse oximeter monitor and its accessories are intended to provide professionally trained health care providers with continuous, non-invasive measurement and monitoring of carbon dioxide concentration of the expired and inspired breath and respiration rate, and with continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate. It is intended for use with neonatal, pediatric, and adult patients in hospitals, hospital-type facilities, and intra-hospital transport environments
Code information (models, lots, serial numbers)
Product # 016400; All SNs, irrespective of production date. Product # 010520; All SNs with production date starting 14.14
Other devices under the same FDA submission
3 devices cleared under K123690 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Microstream CS08651-03Oridion Medical 1987 Ltd. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE1 death15 reports
- Capnostream20p DLCS08651-03Oridion Medical 1987 Ltd. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE4 reports
- Oridion RS08251Oridion Medical 1987 Ltd. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE2 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page