Dialog+ Hemodialysis machines: Conductivity Sensor (BIC)… recallZ-1536-2016 · B Braun Medical Inc · initiated Apr 1, 2016
Class I Terminated All recalls posted in May 2016
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from B Braun Medical Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential leakage of conductivity sensors of the Dialog Dialysis machines.
FDA root cause category: Process control
What the company is doing
B. Braun mailed an Urgent Medical Device Correction notice to affected customers informing them of the Dialog+Hemodialysis System due to the potential for Bicarbonate and End conductivity cells (3456102A and 3456103A respectively) to exhibit cracks. Notice provided customers with actions for Qualified Service Technicians, Clinical Personnel Administering Dialysis treatment and adverse event reporting. All Dialog+Hemodialysis machines will be inspected at the customer's site by the customer Biomed, who has been trained and qualified by B. Braun Avitum. The inspection of each machine will be performed in accordance to a Field Service Information (FSI). Customers with questions were instructed to contact the Technical Support Team at 800-621-0445 or the Customers Service Team at 800-848-2066. For questions regarding this recall call 610-997-4717.
Recall details
- Recall number
- Z-1536-2016
- Recall class
- Class I (most serious)
- Status
- Terminated
- Initiated by the company
- Apr 1, 2016
- Posted by the FDA
- May 3, 2016
- Recalling firm
- B Braun Medical Inc
- Quantity
- 147
- Product code
- KDI
- FDA submissions
- K083460, K963440
- Distribution
- Nationwide Distribution to ND, OH, CA, VA, FL, LA, SC, IL, UT, AZ, KY, IA, WI, OK, TN, SD, AL, NJ, TX, NM, NY, KS, WA, MO, CO, PA, CT, DE, GA, MD, VA, Puerto Rico, and St. Thomas USVI
Product description
Dialog+ Hemodialysis machines: Conductivity Sensor (BIC) Item 3456103A The Dialog+ can be used for implementing and monitoring hemodialysis treatments for patients with acute or chronic kidney failure. The system can be used in hospital, health center and outpatient dialysis center settings when prescribed by a Physician. The following types of renal replacement therapy can be carried out with the system: Hemodialysis (HD) with or without phases of pure ultrafiltration, High flux hemodialysis and low flux hemodialysis.
Code information (models, lots, serial numbers)
Item # 3456103A
Other devices under the same FDA submission
4 devices cleared under K083460, K963440 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Dialog+® 710200LB.Braun Avitum AG · Dialyzer, high permeability with or without sealed dialysate system24 deaths42 reports
- Dialog+® 710200SB.Braun Avitum AG · Dialyzer, high permeability with or without sealed dialysate system11 deaths17 reports
- Dialog+® 710200KB.Braun Avitum AG · Dialyzer, high permeability with or without sealed dialysate system3 reports
- Dialog+® 710200UB.Braun Avitum AG · Dialyzer, high permeability with or without sealed dialysate system1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 73744).
- Z-1531-2016 · TerminatedDialog+ Hemodialysis machines: Dialog A + HE/BIC Product Code 710200L Serial # >200318 and < 2039222 The Dialog+ can be used for implementing and monitoring hemodialysis treatments…
- Z-1532-2016 · TerminatedDialog+ Hemodialysis machines: Dialog A + 1 Blood Pump 120V Product Code 710200K Serial # >202169 and < 203249 The Dialog+ can be used for implementing and monitoring hemodialysis …
- Z-1533-2016 · TerminatedDialog+ Hemodialysis machines: Dialog + 120 V DF+Adimea+WAN-BSL Product Code # 710200S Serial # >200805 and <204001 The Dialog+ can be used for implementing and monitoring hemodial…
- Z-1534-2016 · TerminatedDialog+ Hemodialysis machines: Dialog + 120 V DF+Adimea+WAN-BSL-BIC Product Code # 710200U Serial # 202136 and < 203837 The Dialog+ can be used for implementing and monitoring hemo…
- Z-1535-2016 · TerminatedDialog+ Hemodialysis machines: Conductivity Sensor (BIC) Item # 3456102A The Dialog+ can be used for implementing and monitoring hemodialysis treatments for patients with acute or …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page