Why it was recalled
Insufficient seal strength on sterile barrier packaging. The seal could be jeopardized causing the sterile barrier to be compromised and the product could become nonsterile. If the product becomes non-sterile and is used or implanted, it may lead to an infection.
FDA root cause category: Equipment maintenance
What the company is doing
Zimmer Biomet sent an Urgent Medical Device Recall Notice dated April 1, 2016, to all affected customers. Customers were instructed to locate and return the affected product, and implanting surgeons were instructed to monitor patients. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Questions related to this notice should be directed to (574) 372-1570, Monday through Friday, 8 a.m. to 5 p.m.
Recall details
- Recall number
- Z-1819-2016
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Apr 1, 2016
- Posted by the FDA
- May 26, 2016
- Recalling firm
- Biomet, Inc.
- Product code
- HSB
- FDA submissions
- K050118, K063570, K091976, K871539
- Distribution
- Worldwide Distribution - US (Nationwide) and globally to Argentina, Australia, Brazil, Canada, Chile, Colombia, Costa Rica, Ecaudor, The Netherlands, Japan, Korea, Malaysia, Mexico, Peru.
Product description
Various trauma and sports medicine instruments and implants. Rod, fixation, intramedullary and accessories.
Code information (models, lots, serial numbers)
Part #'s 28234 28338 28811 28813 28821 29260 29822 805065080 14-400428 14-400436 14-405032 14-405040 14-440118 14-440121 14-441280 14-440066S Lot #'s 590970 591080 591090 665210 692040 692250 692280 692290 695900 610470 665230 665240 692440 692450 692480 692620 692660 692670 692690 721680 721720 721730 729080 722730 621840 621860 678300 664740 664820 664870 664880 503760 503770 693840 476230
Other devices under the same FDA submission
23 devices cleared under K050118, K063570, K091976, K871539 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- PHOENIX ANKLE NAIL SYSTEM 14-440221Biomet Orthopedics, LLC · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES4 reports
- PHOENIX NAIL SYSTEM 14-405080Biomet Orthopedics, LLC · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES3 reports
- PHOENIX NAIL SYSTEM 14-405065Biomet Orthopedics, LLC · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES2 reports
- PHOENIX ANKLE NAIL SYSTEM 14-440115Biomet Orthopedics, LLC · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES2 reports
- PHOENIX ANKLE NAIL SYSTEM 14-440215Biomet Orthopedics, LLC · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES2 reports
- PHOENIX ANKLE NAIL SYSTEM 14-440218Biomet Orthopedics, LLC · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES2 reports
- PTN Peritrochanteric Nail 28338Biomet Orthopedics, LLC · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 report
- PHOENIX NAIL SYSTEM 14-405024Biomet Orthopedics, LLC · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 report
- PHOENIX NAIL SYSTEM 14-405030Biomet Orthopedics, LLC · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 report
- PHOENIX NAIL SYSTEM 14-405032Biomet Orthopedics, LLC · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 report
- PHOENIX NAIL SYSTEM 14-405034Biomet Orthopedics, LLC · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 report
- PHOENIX NAIL SYSTEM 14-405040Biomet Orthopedics, LLC · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 74038).
- Z-1817-2016 · TerminatedVarious trauma and sports medicine instruments and implants. Plate, fixation, bone.
- Z-1822-2016 · TerminatedVarious trauma and sports medicine instruments and implants. Pin, fixation, smooth.
- Z-1823-2016 · TerminatedVarious trauma and sports medicine instruments and implants. Screw, fixation, bone.
- Z-1826-2016 · TerminatedVarious trauma and sports medicine instruments and implants. Spinal interlaminal fixation orthosis.
- Z-1828-2016 · TerminatedVarious trauma and sports medicine instruments and implants. Fastener, fixation, nondegradable, soft tissue.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page