Various trauma and sports medicine instruments and implants recallZ-1826-2016 · Biomet, Inc. · initiated Apr 1, 2016
Class II Terminated All recalls posted in May 2016
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Biomet, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Insufficient seal strength on sterile barrier packaging. The seal could be jeopardized causing the sterile barrier to be compromised and the product could become nonsterile. If the product becomes non-sterile and is used or implanted, it may lead to an infection.
FDA root cause category: Equipment maintenance
What the company is doing
Zimmer Biomet sent an Urgent Medical Device Recall Notice dated April 1, 2016, to all affected customers. Customers were instructed to locate and return the affected product, and implanting surgeons were instructed to monitor patients. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Questions related to this notice should be directed to (574) 372-1570, Monday through Friday, 8 a.m. to 5 p.m.
Recall details
- Recall number
- Z-1826-2016
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Apr 1, 2016
- Posted by the FDA
- May 26, 2016
- Recalling firm
- Biomet, Inc.
- Product code
- KWP
- FDA submissions
- K083071
- Distribution
- Worldwide Distribution - US (Nationwide) and globally to Argentina, Australia, Brazil, Canada, Chile, Colombia, Costa Rica, Ecaudor, The Netherlands, Japan, Korea, Malaysia, Mexico, Peru.
Product description
Various trauma and sports medicine instruments and implants. Spinal interlaminal fixation orthosis.
Code information (models, lots, serial numbers)
Part #'s 904756 Lot #'s 625590
Other devices under the same FDA submission
5 devices cleared under K083071 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- VERTEX® Reconstruction System 6958726Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERLAMINAL2 reports
- VERTEX® Reconstruction System 6958734Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERLAMINAL1 report
- VERTEX® Reconstruction System 6958734PTMedtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERLAMINAL1 report
- VERTEX® Reconstruction System 7753520Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERLAMINAL1 report
- VERTEX® Reconstruction System 7751101Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERLAMINAL1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 74038).
- Z-1817-2016 · TerminatedVarious trauma and sports medicine instruments and implants. Plate, fixation, bone.
- Z-1819-2016 · TerminatedVarious trauma and sports medicine instruments and implants. Rod, fixation, intramedullary and accessories.
- Z-1822-2016 · TerminatedVarious trauma and sports medicine instruments and implants. Pin, fixation, smooth.
- Z-1823-2016 · TerminatedVarious trauma and sports medicine instruments and implants. Screw, fixation, bone.
- Z-1828-2016 · TerminatedVarious trauma and sports medicine instruments and implants. Fastener, fixation, nondegradable, soft tissue.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page