Medtronic RestoreSensor, Model 37714 recallZ-1926-2016 · Medtronic Neuromodulation · initiated May 3, 2016
Class II Terminated All recalls posted in June 2016
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Neuromodulation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Loss of therapy during recharging of device. Reportedly, a charging session was terminated prior to obtaining recharge threshold voltage of 3.615, which triggered rapid battery depletion state. As a result, the device depleted to 1.925 volts in one to two days rather than the typical 30 days. Insufficient coupling between recharger and device during recharge session was found to be key factor.
FDA root cause category: Under Investigation by firm
What the company is doing
Consignees were sent on 5/3/2016 a Medtronic "Urgent Medical Device Safety notification" letter dated May 2016. The letter was addressed to "Healthcare Professional". The letter described the problem and the product involved in the recall. The letter listed the recommendation to patients to follow current recharge instructions, paying particular attention to Charging Efficiency and Battery Charge Level indicators on the recharger. Consignees were also requested to complete and return the "Physician Reply Form:. For questions contact Medtronic Neuromodulation Technical services at 1-800-707-0933 weekdays 7 am - 6 pm CST.
Recall details
- Recall number
- Z-1926-2016
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- May 3, 2016
- Posted by the FDA
- Jun 7, 2016
- Recalling firm
- Medtronic Neuromodulation
- Quantity
- 60,941
- Product code
- LGW
- FDA submissions
- P840001
- Distribution
- Worldwide Distribution
Product description
Medtronic RestoreSensor, Model 37714; and, RestoreSensor SureScan, (MRI), Model 97714, Implantable Neurostimulators. Used to treat chronic, intractable pain of the trunk and/or limbs-including unilateral or bilateral pain.
Code information (models, lots, serial numbers)
All lot/serial numbers are affected for both Model numbers 37714 and 97714.
Other devices under the same FDA submission
116 devices cleared under P840001 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Intellis™ AdaptiveStim® 97715Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF2 deaths23,976 reports
- RestoreSensor™SureScan® 97714Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF6 deaths17,903 reports
- Intellis™ AdaptiveStim® 97715Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF11,487 reports
- RestoreSensor 37714Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1 death4,116 reports
- PrimeAdvanced® SureScan® 97702Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1 death2,263 reports
- RestoreUltra® SureScan® 97712Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1,887 reports
- Inceptiv™ 977119Medtronic, Inc. · Stimulator, spinal-cord, totally implanted for pain relief1,339 reports
- Vanta™ AdaptiveStim™ 977006Medtronic, Inc. · Stimulator, spinal-cord, totally implanted for pain relief3 recalls1,151 reports
- Intellis™ 97716Medtronic, Inc. · Stimulator, spinal-cord, totally implanted for relief of pain due to diabetic neuropathy1,125 reports
- Intellis™ 97716Medtronic, Inc. · Stimulator, spinal-cord, totally implanted for relief of pain due to diabetic neuropathy635 reports
- RestoreAdvanced® SureScan® 97713Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF632 reports
- Inceptiv™ 977119Medtronic, Inc. · Stimulator, spinal-cord, totally implanted for pain relief513 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page