Norian Drillable Inject 10 CC-Sterile recallZ-1944-2016 · Synthes (USA) Products LLC · initiated May 11, 2016
Class II Terminated All recalls posted in June 2016
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Synthes (USA) Products LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
DePuy Synthes is initiating a voluntary medical device recall of unexpired and unopened part and lot numbers for the Norian Drillable Injects-Sterile due to the rotary pouch within the referenced lots and associated part numbers may potentially be labeled with the incorrect powder volume/size; however, the rotary pouch contains the correct powder volume/size according to the outer label.
FDA root cause category: Under Investigation by firm
What the company is doing
DePuy Synthes sent an Urgent Medical Device Recall letter dated May 9, 2016, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers with the affected products were asked to call DePuy Synthes at 1-800-479-6329 to obtain a Return Authorization (RA) Number and replacement product(s), Complete the Verification Section (page 3 of this letter) and Return the Verification Section (page 3 of this letter) with the product to: Credit/Returns, DePuy Synthes, 1101 Synthes Avenue, Monument, CO 80132. Send a copy of the completed Verification Section by:Fax: 855-695-8597 or Scan/email: [email protected]. If they had any questions they were asked to call 610-719-5450. For questions regarding this recall call 610-719-5443.
Recall details
- Recall number
- Z-1944-2016
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- May 11, 2016
- Posted by the FDA
- Jun 9, 2016
- Recalling firm
- Synthes (USA) Products LLC
- Quantity
- 2098 units
- Product code
- MQV
- FDA submissions
- K073303, K102722
- Distribution
- Worldwide Distribution - Nationwide to AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD., ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, UT, VA, VT, WA, WI, WV, WY and Internationally to Canada...
Product description
Norian Drillable Inject 10 CC-Sterile; catalog # 07.704.010S Intended for bony voids or defects of the extremities and pelvis that are not instrinsic to the stability of the bony structure.
Code information (models, lots, serial numbers)
Catalog ID 07.704.003S, 07.704.005S and 07.704.010S Lot Numbers: DSC1679, DSC3869, DSC7712, DSC8448, DSC9141, DSD0256, DSD2012, DSD2506, DSD3236, DSC1747, DSC1748, DSC2468, DSC3590, DSC3874, DSC6847, DSC7713, DSC8449, DSC9067, DSC9142, DSD0257, DSD0919, DSD2013, DSD2380, DSD3237. DSC1676, DSC2469, DSC3591, DSC3879, DSC8441, DSC9143, DSD0055, DSD0466, DSD1956, DSD2381
Other devices under the same FDA submission
5 devices cleared under K073303, K102722 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Norian 07.704.005SSynthes GmbH · FILLER, BONE VOID, CALCIUM COMPOUND25 reports
- Norian 07.704.010SSynthes GmbH · FILLER, BONE VOID, CALCIUM COMPOUND15 reports
- Norian 07.704.003SSynthes GmbH · FILLER, BONE VOID, CALCIUM COMPOUND7 reports
- Norian 07.704.110SSynthes GmbH · FILLER, BONE VOID, CALCIUM COMPOUND5 reports
- Norian 07.704.105SSynthes GmbH · FILLER, BONE VOID, CALCIUM COMPOUND4 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 74172).
- Z-1942-2016 · TerminatedNorian Drillable Inject 3 CC-Sterile; catalog # 07.704.003S Intended for bony voids or defects of the extremities and pelvis that are not intrinsic to the stability of the bony str…
- Z-1943-2016 · TerminatedNorian Drillable Inject 5 CC-Sterile; catalog # 07.704.005S Intended for bony voids or defects of the extremities and pelvis that are not instrinsic to the stability of the bony st…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page