Adaptor for Small Battery Drive and Small Battery Drive II recallZ-2784-2016 · The Anspach Effort, Inc. · initiated Jun 24, 2016

Class I Terminated All recalls posted in September 2016

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from The Anspach Effort, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

There was a potential for the Adaptor and Light Adaptor for Small Battery Drive and Small Battery Drive II to generate excessive internal pressure that may cause the products to burst.

FDA root cause category: Device Design

What the company is doing

An Urgent Notice for this Recall was provided to the consignees on 6/24/2016. They were to review their inventory and return the products with instructions. IF YOU HAVE ANY PRODUCT: To immediately quarantine product so it's not used; ensure all in their facility read the Urgent Notice for the Recall; Call the DePuy Synthes Customer Support to obtain a Return Materials Authorization (RMA) Number; to complete the Verification Section indicating the product subject to recall was located; to return the Verification Section with the product to Product Returns, The Anspach Effort Inc., 4500 Riverside Drive, Palm Beach Gardens, FL 33410; Send a copy of the completed Verification Section to DePuy Synthes, Customer Quality Dept by: FAX (561) 627-2682 or Scan/email: [email protected]; Keep the notice visibly posted for awareness until all products have been returned; maintain a copy of the notice subject to this action and keep a copy for your records. The return documentation acknowledges your receipt of the medical device removal information. IF YOU DO NOT HAVE the identified product: complete the Verification Section of the recall, by checking the box indicating no affected product located; Include your name title, address, telephone #, signature and date; Send a copy of the completed Verification Section to DePuy Synthes Customer Quality Dept by: Fax: (561) 627-2682, or Scan/e-mail: [email protected]; This return document acknowledges your receipt of medical device removal information; ADVISORY: For Alternative products: This product does not have any inventory for replacement. Since the recall of this device may have an impact on your surgical schedule, DePuy Synthes Power Tools would like to propose alternative products to assist you with a replacement for the Recalled Adaptor (05.100.024) or Light Adaptor (05.001.108) . The products are the Small Battery Drive II Hand-piece (532.110) can also be operated

Recall details

Recall number
Z-2784-2016
Recall class
Class I (most serious)
Status
Terminated
Initiated by the company
Jun 24, 2016
Posted by the FDA
Sep 27, 2016
Recalling firm
The Anspach Effort, Inc.
Quantity
1,833 devices (05.001.024 and 05.001.108)
Product code
HWE
FDA submissions
K043310
Distribution
Worldwide Distribution - US Nationwide and including the countries of Canada, Austria, Australia, Belgium, Brazil, Chile, China, Colombia, Costa Rica, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Libya, Malaysia, Mexico, Nepal, Netherlands, New Zealand, Norway, Poland, Portugal, Romania, Russia, Saudi Arabia, Singapore, Slovenia, Slovakia, Spain, South Africa, South Korea, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, and Vietnam, and areas of Middle East, Africa, Asia & Pacific.

Product description

Adaptor for Small Battery Drive and Small Battery Drive II. Product Usage: The Light Adaptor for Small Battery Drive (SBD) /Small Battery Drive II (SBD II) is designed for use in traumatology and orthopedic surgery of the skeleton, i.e. drilling, reaming or cutting bone. The Adaptors (05.001.024 and 05.001.108) can be used with one of the electrical consoles belonging to the Electric Pen Drive (EPD) to operate the Small Battery Drive and Small Battery Drive II.

Code information (models, lots, serial numbers)

Model/Catalog: Adapter for Small Battery Drive and Small Battery Drive II for Electric Pen Drive 05.001.024; Lot/Set/Serial: All Manufactured Lots/Serial Numbers.

Other devices under the same FDA submission

66 devices cleared under K043310 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 74835).