Medtronic SynchroMed II Implantable Drug Infusion System.… recallZ-0788-2017 · Medtronic Neuromodulation · initiated Oct 3, 2016
Class I Terminated All recalls posted in December 2016
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Neuromodulation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Medtronic is following up to a May 2013 communication regarding the Priming Bolus function for the SynchroMed Infusion System. Medtronic is updating the Model 8870 software application card (to version AAU01) and the SynchroMed pump labeling to address the priming bolus issue.
FDA root cause category: Under Investigation by firm
What the company is doing
Medtronic Representatives began visiting consignees on 10/3/2016 to replace previous version Model 8870 software application cards, contained within the model 8840 N Vision Clinician Programmer, with the updated version (new version is AAU01). Beginning October 20, 2016, Implanting and Following physicians were sent an Urgent Medical Device Correction letter dated September 2016, . This letter to physicians is informing them of the updated Model 8870 software application card, and changes to SynchroMed infusion system labeling. A response form is requested to be returned to Medtronic. Customers with questions were instructed to contact Medtronic Neuromodulation Technical Services at 1-800-707-0933.
Recall details
- Recall number
- Z-0788-2017
- Recall class
- Class I (most serious)
- Status
- Terminated
- Initiated by the company
- Oct 3, 2016
- Posted by the FDA
- Dec 19, 2016
- Recalling firm
- Medtronic Neuromodulation
- Quantity
- 22,298 software cards
- Product code
- LKK
- FDA submissions
- P860004
- Distribution
- Worldwide Distribution - (US Nationwide) Internationally to Germany, Greece, Australia, United Kingdom, Switzerland, Italy, Norway, France, Canada, and Malta.
Product description
Medtronic SynchroMed II Implantable Drug Infusion System. This system includes: Model 8870 Software Application Card, Model 8840 N'Vision Clinician Programmer, and Model 8637 SynchroMed II Implantable Drug Infusion Pump (supplied in 20 ml or 40 ml reservoir size.) Product Usage: The SynchroMed II Programmable Pump is indicated when patient therapy requires the chronic infusion of the following drugs or fluids: Chronic intraspinal (epidural and intrathecal) infusion of preservative-free morphine sulfate sterile solution in the treatment of chronic intractable pain, Chronic intrathecal infusion of preservative-free ziconotide sterile solution for the management of severe chronic pain, Chronic intrathecal infusion of Lioresal Intrathecal (baclofen Injection ) in the management of serve spasticity, Chronic Intravascular Infusion of Floxuridine (FUDR) or methotrexate for the treatment of primary or metastatic cancer
Code information (models, lots, serial numbers)
software application cards with versions other than AAU01
Other devices under the same FDA submission
51 devices cleared under P860004 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- SYNCHROMED® II 8637-40Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE1 recall7 deaths4,929 reports
- SYNCHROMED® II 8637-20Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE2 recalls5 deaths4,303 reports
- SYNCHROMED ® II 8637-40Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE1 recall6 deaths4,119 reports
- SYNCHROMED® II 8637-40Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE2 recalls10 deaths3,054 reports
- SYNCHROMED ® II 8637-20Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE1 recall4 deaths3,084 reports
- SYNCHROMED® II 8637-20Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE2 recalls1 death2,588 reports
- SynchroMed® II 8637-40Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE1 recall6 deaths2,291 reports
- SYNCHROMED® II 8637-20Medtronic, Inc. · Pump, infusion, implanted, programmable1 recall3 deaths2,242 reports
- SYNCHROMED® II 8637-40Medtronic, Inc. · Pump, infusion, implanted, programmable1 recall4 deaths1,964 reports
- SynchroMed® II 8637-20Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE1 recall4 deaths1,920 reports
- SYNCHROMED™ III 8667-20Medtronic, Inc. · Pump, infusion, implanted, programmable1 death1,937 reports
- SYNCHROMED ® II 8637-20Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE2 deaths1,766 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page