Synthes 0.76 mm Drill Bit/Stryker J-Latch/with 5mm… recallZ-0681-2017 · Synthes (USA) Products LLC · initiated Nov 1, 2016
Class II Terminated All recalls posted in December 2016
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Synthes (USA) Products LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Package may be incorrectly labeled as containing 0.76 mm Drill Bit, but in fact contain a 1.1 mm diameter Drill Bit.
FDA root cause category: Under Investigation by firm
What the company is doing
DePuy Synthes sent an Urgent Notice to customers affected by the recall to inform them of the potential impacts, and actions to be taken for the return of the identified units.
Recall details
- Recall number
- Z-0681-2017
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Nov 1, 2016
- Posted by the FDA
- Dec 1, 2016
- Recalling firm
- Synthes (USA) Products LLC
- Quantity
- 84 units
- Product code
- HWE
- FDA submissions
- K043310
- Distribution
- Distributed Nationwide
Product description
Synthes 0.76 mm Drill Bit/Stryker J-Latch/with 5mm Stop/44.5 mm Non-Sterile
Code information (models, lots, serial numbers)
Model/Catalog # 316.15 Lot/Serial Number U172728
Other devices under the same FDA submission
66 devices cleared under K043310 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- NA 05.001.010Synthes GmbH · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE770 reports
- NA 05.001.040Synthes GmbH · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE61 reports
- NA 05.001.039Synthes GmbH · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE50 reports
- NA 05.001.046Synthes GmbH · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)41 reports
- NA 05.001.037Synthes GmbH · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)26 reports
- MatrixMANDIBLE 03.503.476Synthes GmbH · DRILL, BONE, POWERED26 reports
- NA 310.510Synthes GmbH · DRILL, BONE, POWERED17 reports
- NA 317.82Synthes GmbH · DRILL, BONE, POWERED16 reports
- NA 05.001.049Synthes GmbH · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)13 reports
- NA 05.001.123Synthes GmbH · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE12 reports
- MatrixMANDIBLE 03.503.479Synthes GmbH · DRILL, BONE, POWERED11 reports
- Matrix 03.511.248Synthes GmbH · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE10 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page