CADD Administration Set, 21-7339-24, Rx Only, Sterile EO recallZ-1767-2017 · Smiths Medical ASD Inc. · initiated Mar 17, 2017

Class II Terminated All recalls posted in March 2017

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Smiths Medical ASD Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Smiths Medical has become aware that the drawing on the Instructions for Use (IFU) on certain yellow striped CADD Administration Sets (part number 21-7339-24), incorrectly represent the actual filter size of the product. Although the text description on the IFU correctly identifies the product to include a 0.2 micron filter, the illustration on the IFU identifies the filter size as 2.0 microns.

FDA root cause category: Packaging process control

What the company is doing

Smiths Medical sent an "Urgent Medical Device Field Safety Notice" dated March 13, 2017. The letter was addressed to "Risk Manager". It provided a list of the Affected devices, Risk to Health, Instructions to Distributors & Customers. Advised consignees to notify their customers. There is no need to return the devices to Smiths Medical. Requested consignees to complete the Urgent Medical Device Field Safety Notice (FSN) Response Form and return to [email protected]. For questions contact Stericycle at 844-265-7407. For further questions regarding this recall , please call (844) 265-7407.

Recall details

Recall number
Z-1767-2017
Recall class
Class II
Status
Terminated
Initiated by the company
Mar 17, 2017
Posted by the FDA
Mar 30, 2017
Recalling firm
Smiths Medical ASD Inc.
Quantity
38,415 (38145 US; 2268 OUS)
Product code
FPA
FDA submissions
K040636
Distribution
US:AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MS, MT, NC, NE, NJ, NM, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WV, WY. OUS: Canada

Product description

CADD Administration Set, 21-7339-24, Rx Only, Sterile EO.

Code information (models, lots, serial numbers)

46X817, 46X829, 46X899, 46X828, 46X896, 46X990, 46X831, 46X897, 46X1001.

Other devices under the same FDA submission

18 devices cleared under K040636 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page