Skip to content

MEDTECH ROSA Brain 3.0 recallZ-1754-2017 · Zimmer Biomet, Inc. · initiated Feb 27, 2017

Class II Terminated All recalls posted in March 2017

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Zimmer Biomet, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Unapproved change made by the supplier.

FDA root cause category: Component change control

What the company is doing

Zimmer Biomet sent an Urgent Medical Device Recall Correction letter on February 27, 2017,to all affected customers informing them of the following: Description of the issue: A robot arm file parameter linked to arm calibration was modified by the robot arm supplier without notification to MEDTECH. As a result, once the robot arm is on trajectory, an error leading to a device reboot may decrease the devices applicative accuracy, one of the essential performance requirements, to a level above the specification. This may happen after a shutdown of the device, (voluntary or due to an error), without a return of the robot arm position to a Parking position. In the worst case scenario, this may cause an incorrect positioning of the instruments by the ROSA devices. If this is not detected by the user this could ultimately lead to ineffective treatment, serious injury or even death of the patient. MEDTECH did not record, in normal conditions of use, any customer reports linked to this issue. Required actions by users: Pending the implementation of a permanent correction, MEDTECH is recommending the following actions to all users of the ROSA device: 1- After a shutdown of the device, (voluntary or due to an error), without a return of the robot arm in Parking position, ROSA device must be switched off with the I/O switch button on the device rear panel. 2- Restart the device after 10 seconds, and after full extinction of emergency button light. If the Reset button at the device rear panel is red, restart the ROSA PC by pushing the Reset button. 3- If robot position is NOT Home , load patient folder and main software interface will appear on the screen. a. Go to the Guidance menu, select trajectories and follow instructions. b. Click on Start button. c. Start recovery procedure and follow all instructions displayed on the screen until robot arm position is Home. d. Close the patient folder e. Click on Exit and follow all ins

Recall details

Recall number
Z-1754-2017
Recall class
Class II
Status
Terminated
Initiated by the company
Feb 27, 2017
Posted by the FDA
Mar 27, 2017
Recalling firm
Zimmer Biomet, Inc.
Quantity
16
Product code
HAW
FDA submissions
K151359
Distribution
Nationwide Distribution to CA, FL, IL, KY, MA, MN, NC, NY, OH, PA, and District of Columbia.

Product description

MEDTECH ROSA Brain 3.0 The device is intended for the spatial positioning and orientation of instrument holders or tool guides to be used by neurosurgeons to guide standard neurosurgical instruments (biopsy needle, stimulation or recording electrode, endoscope). The device is indicated for any neurological procedure in which the use of stereotactic surgery may be appropriate.

Code information (models, lots, serial numbers)

Serial #'s: BR16004, BR16005, BR16006, BR16009, BR16010, BR16011, BR16012, BR16013, BR16014, BR16015, BR16016, BR16017, BR16018, BR16021, BR16023 and BR16025.

Other devices under the same FDA submission

1 device cleared under K151359 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page