HeartMate II LVAS with Pocket Controller recallZ-2051-2017 · Thoratec Corporation · initiated Mar 30, 2017
Class I Terminated All recalls posted in May 2017
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Thoratec Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
St. Jude Medical is providing all users of their HeartMate II LVAS with Pocket Controller with new software and updates to hardware to make the exchange to a backup controller easier. This is related to recall Z-1227/1230-2014.
FDA root cause category: Device Design
What the company is doing
Urgent Medical Device Correction letters, dated March 29, 2017, were sent to customers. The letters provided information on what the current correction will involve to ensure patients are successfully able to exchange their pocket controller in an emergency situation. Patients will receive the updates at scheduled appointment. Questions regarding the notice, should be directed to Abbott HeartLine at 800-456-1477; or, your Abbott Sales Representative.
Recall details
- Recall number
- Z-2051-2017
- Recall class
- Class I (most serious)
- Status
- Terminated
- Initiated by the company
- Mar 30, 2017
- Posted by the FDA
- May 24, 2017
- Recalling firm
- Thoratec Corporation
- Quantity
- 24,077 active units in US
- Product code
- DSQ
- FDA submissions
- P060040
- Distribution
- Worldwide Distribution -- US and Austria, Belgium, Brazil, Canada, Cayman Islands, Chile, Columbia, Croatia, Czech Republic, Denmark, Egypt, Estonia, France, Germany, Greece, Hong Kong, Hungary, India, Iran, Ireland, Israel, Italy, Kazakhstan, Kuwait, Latvia, Lebanon, Liechtenstein, Lithuania, Malaysia, Orman, Palastine, Poland, Qatar, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, Spain, Sweden, Switzerland, Thailand, The Netherlands, Turkey, United Arab Emirates, and United Kingdom.
Product description
HeartMate II LVAS with Pocket Controller; 107801 - HMII PUMP & POCKET CTRL ONLY UDI:00813024011286 Intended to provide hemodynamic support in patients with end-stage, refractory left ventricular heart failure; intended for use inside or outside the hospital.
Code information (models, lots, serial numbers)
All Serial numbers.
Other devices under the same FDA submission
7 devices cleared under P060040 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- HeartMate™ 107754Thoratec Corporation · Ventricular (assist) bypass2 recalls447 reports
- HeartMate Touch™ HMT1100Thoratec Corporation · Ventricular (assist) bypass1 recall50 reports
- HeartMate™ 107760Thoratec Corporation · Ventricular (assist) bypass1 recall6 reports
- HeartMate Touch™ HMT1100-RThoratec Corporation · Ventricular (assist) bypass1 recall3 reports
- HeartMate™ 106128Thoratec Corporation · Ventricular (assist) bypass1 recall0 reports
- HeartMate Touch™ HMT2100Thoratec Corporation · Ventricular (assist) bypass10 reports
- HeartMate™ 1340Thoratec Corporation · Ventricular (assist) bypass1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 76978).
- Z-2052-2017 · TerminatedHeartMate II LVAS with Pocket Controller; 106015 - HMII LVAD SEALED KIT-PCTRL, NA UDI: 00813024011224 Intended to provide hemodynamic support in patients with end-stage, refracto…
- Z-2053-2017 · TerminatedHeartMate II LVAS with Pocket Controller; 106762 - HMII POCKET CONTROLLER, NA; UDI: 00813024011286 Intended to provide hemodynamic support in patients with end-stage, refractory …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗