Skip to content

Alaris PC Unit, Model 8015 recallZ-2671-2017 · CareFusion 303, Inc. · initiated Jun 12, 2017

Class II Terminated All recalls posted in June 2017

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from CareFusion 303, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

BD initiated the recall of Alaris PC unit model 8015 after the firm identified five scenarios which can result in the occurrence of Systems Error Code 255-16-275 and can potentially result in interrupted infusions.

FDA root cause category: Software design

What the company is doing

BD sent an Medical Device Safety Notification Letter dated June 12, 2017 , to customers to inform them that BD is issuing this Safety Notification letter to inform you of reports of customers experiencing a System Error (error code: 255-16-275) with Alaris System PC unit model 8015 that can result in interruption of infusions. Customers are informed of the issue, potential risk, and required actions to be taken by users and by BD. Customers with questions with the safety notification are instructed to call (888) 562-6018 or email [email protected]. Customers with adverse event reports are instructed to call (888) 812-3266 or email [email protected]. Customers with technical questions are instructed to call (888) 812-3229 or email [email protected]. For further questions, please call (858) 617-5925.

Recall details

Recall number
Z-2671-2017
Recall class
Class II
Status
Terminated
Initiated by the company
Jun 12, 2017
Posted by the FDA
Jun 19, 2017
Recalling firm
CareFusion 303, Inc.
Quantity
568,283 units
Product code
FRN
FDA submissions
K051641
Distribution
Worldwide Distribution - USA (nationwide) and to the countries of : Europe, Australia, New Zealand, South Africa, Greater Asia, Middle East, and Canada.

Product description

Alaris PC Unit, Model 8015

Code information (models, lots, serial numbers)

Notification will be distributed to all Alaris PC Unit model 8015 customers with software versions greater than 9.12.

Other devices under the same FDA submission

1 device cleared under K051641 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.