Legion(R) Tib Cone Impactor Heads recallZ-2625-2017 · Smith & Nephew, Inc. · initiated May 30, 2017
Class II Terminated All recalls posted in June 2017
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Smith & Nephew, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The incorrect Loctite adhesive was used to assemble the impactor heads.
FDA root cause category: Process control
What the company is doing
The firm notified the consignee by email on 05/30/2017. The notice directed the consignee to inspect its inventory and locate any unused devices from the listed product and batch numbers, and quarantine them immediately. Distributors were directed to notify their customers of the field action and ensure the recall directions were followed.
Recall details
- Recall number
- Z-2625-2017
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- May 30, 2017
- Posted by the FDA
- Jun 12, 2017
- Recalling firm
- Smith & Nephew, Inc.
- Quantity
- 46 units
- Product code
- JWH
- FDA submissions
- K162775
- Distribution
- US distribution to TN
Product description
LEGION(R) TIB CONE IMPACTOR HEADS: a) REF 71441972, ID 24MM b) REF 71441971, ID 22MM c) REF 71441973, ID 26MM d) REF 71441974, ID 28MM e) REF 71441975, ID 30MM
Code information (models, lots, serial numbers)
Lot numbers: a) 16HGA0040 b) 16JGA0003 c) 16JGA0005 d) 16JGA0006 e) 16JGA0007
Other devices under the same FDA submission
12 devices cleared under K162775 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Legion 71442011Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER13 reports
- Legion 71431909Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER12 reports
- Legion 71431908Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER10 reports
- Legion 71431911Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER5 reports
- Legion 71441977Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER3 reports
- Legion 71935387Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER3 reports
- Legion 71441976Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER2 reports
- Legion 71442013Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER2 reports
- Legion 71441978Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 report
- Legion 71441979Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 report
- Legion 71935386Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 report
- Legion 71935409Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 77413).
- Z-2624-2017 · TerminatedLEGION(R) FEM CONE IMPACTOR HEADS: a) REF 71441976, ID 18MM b) REF 71441977, ID 20MM c) REF 71441978, ID 22MM
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page