Skip to content

Exactech Optetrak Logic CC 15mm Tibial Augment Screws recallZ-2704-2017 · Exactech, Inc. · initiated Apr 7, 2017

Class II Terminated All recalls posted in June 2017

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Exactech, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Missing Screw in carton. Nine (9) serial numbers were identified as being inadvertently packaged as one per carton when two per carton are required.

FDA root cause category: Process control

What the company is doing

Consignees were initially contacted by phone on 04/07/2017. Consignees were sent a recall notice by mail, dated 04/13/2017. The notice requested that consignees immediately cease distribution/use of product, that the recall Extend the information to accounts in possession of the product Identify/quarantine any subject devices in your inventory Complete & return the attached Recall Inventory Response Form to Exactech within 5 business days at 2411 NW 66th Ct, Gainesville, FL 32653 or faxed to 352-337-3915. Questions from customers are to be directed to Kaya Davis at [email protected] or 800-392-2832.

Recall details

Recall number
Z-2704-2017
Recall class
Class II
Status
Terminated
Initiated by the company
Apr 7, 2017
Posted by the FDA
Jun 19, 2017
Recalling firm
Exactech, Inc.
Quantity
9 devices
Product code
JWH
FDA submissions
K150890
Distribution
CA and NY.

Product description

Exactech Optetrak Logic CC 15mm Tibial Augment Screws

Code information (models, lots, serial numbers)

Catalog No. 02-012-50-9015; Serial #'s 4739530, 4739531, 4739532, 4739533, 4739534, 4739535, 4739536, 4739537, and 4739538.

Other devices under the same FDA submission

71 devices cleared under K150890 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page