MyCareLink Smart recallZ-0400-2018 · Medtronic Inc., Cardiac Rhythm and Heart Failure · initiated Jul 14, 2017
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Inc., Cardiac Rhythm and Heart Failure or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Patients monitored on two (2) or more implanted Medtronic heart devices in the Medtronic CareLink Network may have potential impact on the ability to remotely monitor the patient's heart devices. Potential impacts could lead to missed CareAlert notifications or device reports.
FDA root cause category: Error in labeling
What the company is doing
Customers were notified via letter on 07/14/2017. Medtronic communicated that an updated Clinician Instructions for Use has been released with the following updated information: In patients with more than one Medtronic implanted hear device, device data may not be able to transmit from any of the implanted devices to the physician via the Monitor; The patient could potentially transmit device data unintentionally from one of the implanted heart devices when attempting to transmit from another device implanted in the same patient; and a Medtronic Programmer can be used to obtain data from each implanted heart device. Instructions included to ensure all appropriate personnel are notified of this correction.
Recall details
- Recall number
- Z-0400-2018
- Status
- Terminated
- Initiated by the company
- Jul 14, 2017
- Recalling firm
- Medtronic Inc., Cardiac Rhythm and Heart Failure
- Quantity
- 1,519,984 enrollments total
- Product code
- NVZ
- FDA submissions
- P890003
- Distribution
- Domestic: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, Internationally: Austria, Belgium, Brunei, China, Czech Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Japan, Kazakhstan, Latvia, Lithuania, Malaysia, Malta, Netherlands, Norway, Philippines, Poland, Portugal, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, UAE, United Kingdom
Product description
MyCareLink Smart Patient Monitors. It is intended for use with a compatible Medtronic patient implanted heart device. The reader is a portable electronic device that communicates with an implanted heart device
Code information (models, lots, serial numbers)
Model 25000, UPN/GTIN 00643169527546, 00643169569744, 00643169666337, 00643169697843¿, 00643169862876, 00643169697850, 00643169742246, 00643169862883, 00643169666375, 00643169666382, 00643169666344, 00643169666351, 00643169666368, 00643169666320, 00643169590397, 00643169527584, 00643169527591, 00643169527553, 00643169527560, 00643169527577, 00643169947108
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- MyCareLink Smart™ 25000Medtronic, Inc. · CONTAINER, SPECIMEN, STERILE1 recall5 reports
- MyCareLink Smart™ 25000Medtronic, Inc. · CONTAINER, SPECIMEN, STERILE1 recall0 reports
Product lines
Other devices under the same FDA submission
81 devices cleared under P890003 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- NA 229047Medtronic, Inc. · Pacemaker/icd/crt non-implanted components1 recall662 reports
- Medtronic CareLink® Programmer 2090Medtronic, Inc. · Pacemaker/icd/crt non-implanted components3 recalls559 reports
- Medtronic CareLink® Programmer 2090WMedtronic, Inc. · Programmer, pacemaker3 recalls489 reports
- CareLink Encore™ 29901AMedtronic, Inc. · Pacemaker/icd/crt non-implanted components2 recalls474 reports
- Medtronic CareLink® Programmer 2090Medtronic, Inc. · Pacemaker/icd/crt non-implanted components3 recalls422 reports
- CareLink Encore™ 29901Medtronic, Inc. · implantable pacemaker Pulse-generator3 recalls265 reports
- Medtronic CareLink® Programmer 2090WMedtronic, Inc. · Programmer, pacemaker3 recalls208 reports
- CareLink SmartSync™ 24970AMedtronic, Inc. · TESTER, PACEMAKER ELECTRODE FUNCTION308 reports
- Medtronic CareLink® Programmer 2090Medtronic, Inc. · Pacemaker/icd/crt non-implanted components3 recalls151 reports
- Medtronic CareLink® Programmer 2090Medtronic, Inc. · Programmer, pacemaker2 recalls198 reports
- Medtronic CareLink® Programmer 2090Medtronic, Inc. · Programmer, pacemaker3 recalls108 reports
- Medtronic CareLink® Programmer 2090AA4Medtronic, Inc. · Pacemaker/icd/crt non-implanted components3 recalls82 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 77964).
- Z-0399-2018 · TerminatedMyCareLink Patient Monitors. It is intended for used with a compatible Medtronic patient implanted heart device. The monitor is an external electronic device that interfaces with …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page