Platinium DR DF4 1540, Model Number TDF036U, UDI/GTIN… recallZ-3232-2017 · Sorin Group Italia SRL - CRF · initiated Jul 24, 2017
Class II Terminated All recalls posted in September 2017
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Sorin Group Italia SRL - CRF or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There is a possibility of overconsumption of certain PLATINIUM Implantable Cardiac Defibrillators (ICDs) and Cardiac ResynchronizationTherapy Defibrillators (CRT-Ds) following an ElectroStatic Discharge (ESD) during the implant surgery or a Magnetic Resonance Imaging (MRI) scan, and what actions to take.
FDA root cause category: Software design
What the company is doing
LivaNova issued a communication to physicians treating patients implanted with the products in scope in the form of a customer letter beginning July 24, 2017. The letter instructs users to do the following: 1. In order to mitigate the potential risks associated with both triggering events (ESD at implant or MRI scan), LivaNova recommends physicians follow the patients at the periodicity already stated in the implant manual, especially: " Before the patient is discharged and at each subsequent follow-up, it is advisable to check the battery status and the occurrence of system warnings; " It is recommended that a routine follow-up examination be done one month after discharge, and then every three months until the device nears the replacement date.
Recall details
- Recall number
- Z-3232-2017
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Jul 24, 2017
- Posted by the FDA
- Sep 28, 2017
- Recalling firm
- Sorin Group Italia SRL - CRF
- Quantity
- 86 units
- Product code
- MRM
- FDA submissions
- P980049
- Distribution
- US Nationwide Distribution
Product description
Platinium DR DF4 1540, Model Number TDF036U, UDI/GTIN 08031527015477, Implantable cardioverter defibrillator, Biventricular or ventricular antitachycardia pacing, Dual or single chamber arrhythmia detection Product Usage: PLATINIUM VR is indicated for use in patients who are at high risk of sudden cardiac death due to ventricular tachyarrhythmias and who have experienced one of the following situations: - Survival of at least one episode of cardiac arrest (manifested by the loss of consciousness) due to ventricular tachyarrhythmia, - Recurrent, poorly tolerated sustained ventricular tachycardia (VT).
Code information (models, lots, serial numbers)
All Serial Numbers
Other devices under the same FDA submission
4 devices cleared under P980049 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- CPR4 B10031Microport Crm Srl · Defibrillator, Implantable, Dual-Chamber22 reports
- Paradym RF DR 9550Microport Crm Srl · DEFIBRILLATOR, IMPLANTABLE, DUAL-CHAMBER9 reports
- PLATINIUM DR 1510Microport Crm Srl · DEFIBRILLATOR, IMPLANTABLE, DUAL-CHAMBER3 reports
- PLATINIUM VR 1210Microport Crm Srl · DEFIBRILLATOR, IMPLANTABLE, DUAL-CHAMBER2 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 77983).
- Z-3228-2017 · TerminatedPlatinium VR 1210, Model Number TDF031U, UDI/GTIN 08031527015514, Implantable cardioverter defibrillator, Biventricular or ventricular antitachycardia pacing, Dual or single cham…
- Z-3229-2017 · TerminatedPlatinium DR 1510, Model Number TDF032U, UDI/GTIN 08031527015521, Implantable cardioverter defibrillator, Biventricular or ventricular antitachycardia pacing, Dual or single cham…
- Z-3230-2017 · TerminatedPlatinium CRT DF1 1711, Model Number TDF033U, UDI/GTIN 08031527015538, Implantable cardioverter defibrillator, Biventricular or ventricular antitachycardia pacing, Dual or single…
- Z-3231-2017 · TerminatedPlatinium VR DF4 1240, Model Number TDF035U, UDI/GTIN 08031527015460, Implantable cardioverter defibrillator, Biventricular or ventricular antitachycardia pacing, Dual or single …
- Z-3233-2017 · TerminatedPlatinium CRT-D 1741, Model Number TDF037U, UDI/GTIN 08031527015484, Implantable cardioverter defibrillator, Biventricular or ventricular antitachycardia pacing, Dual or single c…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page