Skip to content

Platinium CRT DF1 1711, Model Number TDF033U, UDI/GTIN… recallZ-3230-2017 · Sorin Group Italia SRL - CRF · initiated Jul 24, 2017

Class II Terminated All recalls posted in September 2017

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Sorin Group Italia SRL - CRF or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

There is a possibility of overconsumption of certain PLATINIUM Implantable Cardiac Defibrillators (ICDs) and Cardiac Resynchronization Therapy Defibrillators (CRT-Ds) following an ElectroStatic Discharge (ESD) during the implant surgery or a Magnetic Resonance Imaging (MRI) scan, and what actions to take.

FDA root cause category: Software design

What the company is doing

LivaNova issued a communication to physicians treating patients implanted with the products in scope in the form of a customer letter beginning July 24, 2017. The letter instructs users to do the following: 1. In order to mitigate the potential risks associated with both triggering events (ESD at implant or MRI scan), LivaNova recommends physicians follow the patients at the periodicity already stated in the implant manual, especially: " Before the patient is discharged and at each subsequent follow-up, it is advisable to check the battery status and the occurrence of system warnings; " It is recommended that a routine follow-up examination be done one month after discharge, and then every three months until the device nears the replacement date.

Recall details

Recall number
Z-3230-2017
Recall class
Class II
Status
Terminated
Initiated by the company
Jul 24, 2017
Posted by the FDA
Sep 28, 2017
Recalling firm
Sorin Group Italia SRL - CRF
Quantity
154 units
Product code
NIK
FDA submissions
P060027
Distribution
US Nationwide Distribution

Product description

Platinium CRT DF1 1711, Model Number TDF033U, UDI/GTIN 08031527015538, Implantable cardioverter defibrillator, Biventricular or ventricular antitachycardia pacing, Dual or single chamber arrhythmia detection Product Usage: PLATINIUM VR is indicated for use in patients who are at high risk of sudden cardiac death due to ventricular tachyarrhythmias and who have experienced one of the following situations: - Survival of at least one episode of cardiac arrest (manifested by the loss of consciousness) due to ventricular tachyarrhythmia, - Recurrent, poorly tolerated sustained ventricular tachycardia (VT).

Code information (models, lots, serial numbers)

All Serial Numbers

Other devices under the same FDA submission

3 devices cleared under P060027 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 77983).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page