Why it was recalled

Potential for failure of the device safety interlock. The safety interlock prevents an empty single-use loading unit from being fired a second time.

FDA root cause category: Under Investigation by firm

What the company is doing

All consignees were notified via Federal Express or certified mail on September 13, 2017, and the letter informs customers of the potential for failure of the device safety interlock and the actions they should take. Customers are requested to acknowledge receipt and understanding of the Urgent Field Corrective Action Notice. All in control inventory and product returned from customers will be scrapped.

Recall details

Recall number
Z-0636-2018
Recall class
Class II
Status
Terminated
Initiated by the company
Sep 13, 2017
Posted by the FDA
Feb 16, 2018
Recalling firm
Covidien LLC
Quantity
3,816
Product code
GDW
FDA submissions
K111825
Distribution
Nationally

Product description

Endo GIA Auto Suture Universal Articulating Loading Unit for use in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection and creation of anastomosis.

Code information (models, lots, serial numbers)

N7F1058KX, N7F1121KX, N7F1056KX, N7F1057KX