Better-Bladder cardiopulmonary bypass blood reservoir recallZ-0817-2018 · Circulatory Technology Inc · initiated Sep 22, 2017

Class II Terminated All recalls posted in March 2018

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Circulatory Technology Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The device is used as part of extracorporeal membrane oxygenation (ECMO) cardiopulmonary bypass circuits and may collapse during use which can increase resistance to flow in the venous line and cause a drop in circuit blood flow.

FDA root cause category: Process control

What the company is doing

The firm is requesting that consignees immediately inspect inventory, and quarantine all units having the aforementioned lots. These units must be returned to CTI. CTI will provide UPS shipping labels for returning the units and will provide replacements or credit for the returned units. Should the inventory consist of only the affected lots then, until the consignee receives replacement BB14s from an unaffected lot, the firm recommends using the BB14 from the affected lots only if, prior to its use, they pull on the pigtail of the Better-Bladder to test its bond integrity prior to its use. If the pigtail dislodges then return the unit for credit or replacement. The consignees should complete and return the enclosed response form via email to [email protected] as soon as possible.

Recall details

Recall number
Z-0817-2018
Recall class
Class II
Status
Terminated
Initiated by the company
Sep 22, 2017
Posted by the FDA
Mar 1, 2018
Recalling firm
Circulatory Technology Inc
Quantity
840
Product code
DTN
FDA submissions
K981284
Distribution
**Requested update** Distributed in 19 states: AL, AR, CA, FL, GA, IL, IN, KY, MD, MO, NJ, NY, OH, OR, PA, TN, TX, VA, WA, and the District of Columbia.

Product description

Better-Bladder cardiopulmonary bypass blood reservoir; Models: BB14 (individual, sterile, pouched) and BB14NS (sold bulk to kit manufacturers)

Code information (models, lots, serial numbers)

**Recall expanded to add additional lot numbers highlighted with "**": BB14 Lot Numbers: 5290-S19785, 5290-S19786, 5290-S19872, **5290-S19925; BB14NS Lot Numbers: 17062201, 17062202, 17062203, and **170905

Other devices under the same FDA submission

2 devices cleared under K981284 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 79196).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page