Skip to content

BD Vacutainer EDTA Lavender Top Tubes: Catalog Numbers: 1.… recallZ-1139-2018 · Becton Dickinson & Company · initiated Mar 22, 2018

Class I Terminated All recalls posted in March 2018

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Becton Dickinson & Company or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

BD Vacutainer EDTA Lavender, Tan, and Pink Top Tubes and BD Vacutainer Lithium Heparin Green Top Tubes can cause an underestimation of lead in blood samples when used with Magellan Diagnostics LeadCare assays, employing the Anodic Stripping Voltammetry (ASV) methodology, or any other assay employing ASV methodology.

FDA root cause category: Device Design

What the company is doing

On March 22, 2018, BD issued a press release. BD intends to distribute URGENT MEDICAL DEVICE CORRECTION notices to their customers in the upcoming days via courier service. The URGENT MEDICAL DEVICE CORRECTION notification has been posted on BD's website along with a link to their press release. BD has updated the IFU for BD Vacutainer EDTA and BD Vacutainer Lithium Heparin Tubes to include the noted precaution for use of Lavender, Tan, Pink and Green Top tubes on Magellan Diagnostics Leadcare assays using ASV methodology. **Distributors are advised to Take the Following Actions: 1) This is a Medical Device Correction and there is no need to return or discard product. The product can continue to be used with other non-ASV blood lead level test technologies such as GFAAS and ICP-MS and all assays which do not employ ASV methodology. 2) If you have distributed this product, please identify your customers and notify them of this recall by sharing a copy of the correction notice. 3) Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the covered product. 4) If you would like BD to conduct the notification to your customers, please email your customer list within 3 business days to [email protected]. Please include the contact name, address, phone number, email address, and/or fax numbers for each customer. Distributors who require further assistance, please contact BD at 1-888-237-2762 (select Option #3 and then Option # 4) between 8AM and 5 PM, CT Monday through Friday. **Medical Directors, Risk Managers, Medical Device Safety Officers & Lab Managers are advised to Take the Following Actions: 1) This is a Medical Device Correction and there is no need to return or discard product. The product can continue to be used with other non-ASV blood lead level test technologies such as GFAAS and ICP-MS and all assays which do not employ ASV methodology. 2) Continue to follow FDA's safety

Recall details

Recall number
Z-1139-2018
Recall class
Class I (most serious)
Status
Terminated
Initiated by the company
Mar 22, 2018
Posted by the FDA
Mar 22, 2018
Recalling firm
Becton Dickinson & Company
Quantity
1,116,012,050
Product code
JKA
FDA submissions
K981013
Distribution
Distributed in 46 states: AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, ND, NE, NH, NJ, NM, NY, NV, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and Puerto Rico.

Product description

BD Vacutainer EDTA Lavender Top Tubes: Catalog Numbers: 1. 366401 (UDI: 50382903664011); 2. 366450 (UDI: 50382903664509); 3. 366643 (UDI: 50382903666435); 4. 367835 (UDI: 50382903678353); 5. 367841 (UDI: 50382903678414); 6. 367844 (UDI: 50382903678445); 7. 367856 (UDI: 50382903678568); 8. 367861 (UDI: 50382903678612); 9. 367862 (UDI: 50382903678629); 10. 367863 (UDI: 50382903678636); 11. 368021 (UDI: 50382903680219); 12. 368047 (UDI: 50382903680479); 13. 368054 (UDI: 50382903680547); 14. 368661 (UDI: 50382903686617); 15. 364300 (UDI: pre-UDI)

Code information (models, lots, serial numbers)

All lots. Catalog Number (Shelf Life): 1. 366401 (24 months); 2. 366450 (24 months); 3. 366643 (12 months); 4. 367835 (16 months); 5. 367841 (15 months); 6. 367844 (16 months); 7. 367856 (16 months); 8. 367861 (16 months); 9. 367862 (16 months); 10. 367863 (17 months); 11. 368021 (16 months); 12. 368047 (15 months); 13. 368054 (16 months); 14. 368661 (17 months); 15. 364300 (Requested)

Other devices under the same FDA submission

1 device cleared under K981013 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 79223).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page