BD Vacutainer Lithium Heparin Green Top Tube recallZ-1142-2018 · Becton Dickinson & Company · initiated Mar 22, 2018
Class I Terminated All recalls posted in March 2018
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Becton Dickinson & Company or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
BD Vacutainer EDTA Lavender, Tan, and Pink Top Tubes and BD Vacutainer Lithium Heparin Green Top Tubes can cause an underestimation of lead in blood samples when used with Magellan Diagnostics LeadCare assays, employing the Anodic Stripping Voltammetry (ASV) methodology, or any other assay employing ASV methodology.
FDA root cause category: Device Design
What the company is doing
On March 22, 2018, BD issued a press release. BD intends to distribute URGENT MEDICAL DEVICE CORRECTION notices to their customers in the upcoming days via courier service. The URGENT MEDICAL DEVICE CORRECTION notification has been posted on BD's website along with a link to their press release. BD has updated the IFU for BD Vacutainer EDTA and BD Vacutainer Lithium Heparin Tubes to include the noted precaution for use of Lavender, Tan, Pink and Green Top tubes on Magellan Diagnostics Leadcare assays using ASV methodology. **Distributors are advised to Take the Following Actions: 1) This is a Medical Device Correction and there is no need to return or discard product. The product can continue to be used with other non-ASV blood lead level test technologies such as GFAAS and ICP-MS and all assays which do not employ ASV methodology. 2) If you have distributed this product, please identify your customers and notify them of this recall by sharing a copy of the correction notice. 3) Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the covered product. 4) If you would like BD to conduct the notification to your customers, please email your customer list within 3 business days to [email protected]. Please include the contact name, address, phone number, email address, and/or fax numbers for each customer. Distributors who require further assistance, please contact BD at 1-888-237-2762 (select Option #3 and then Option # 4) between 8AM and 5 PM, CT Monday through Friday. **Medical Directors, Risk Managers, Medical Device Safety Officers & Lab Managers are advised to Take the Following Actions: 1) This is a Medical Device Correction and there is no need to return or discard product. The product can continue to be used with other non-ASV blood lead level test technologies such as GFAAS and ICP-MS and all assays which do not employ ASV methodology. 2) Continue to follow FDA's safety
Recall details
- Recall number
- Z-1142-2018
- Recall class
- Class I (most serious)
- Status
- Terminated
- Initiated by the company
- Mar 22, 2018
- Posted by the FDA
- Mar 22, 2018
- Recalling firm
- Becton Dickinson & Company
- Quantity
- 117,816,603
- Product code
- JKA
- FDA submissions
- K160657
- Distribution
- Distributed in 46 states: AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, ND, NE, NH, NJ, NM, NY, NV, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and Puerto Rico.
Product description
BD Vacutainer Lithium Heparin Green Top Tube; Catalog Numbers: 1. 366664 (UDI: 50382903666640); 2. 367884 (UDI: 50382903678841); 3. 367886 (UDI: 50382903678865); 4. 368040 (UDI: 50382903680400); 5. 366667 (UDI: 50382903666671); 6. 367880 (UDI: 50382903678803)
Code information (models, lots, serial numbers)
All lots. Catalog Number (Shelf Life): 1. 366664 (Requested); 2. 367884 (Requested); 3. 367886 (Requested); 4. 368040 (Requested); 5. 366667 (Requested); 6. 367880 (Requested)
Other devices under the same FDA submission
4 devices cleared under K160657 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- BD Vacutainer® 365045Becton, Dickinson U.K. Limited · Tubes, vials, systems, serum separators, blood collection1 recall6 reports
- BD Vacutainer® 365043Becton, Dickinson U.K. Limited · Tubes, vials, systems, serum separators, blood collection1 recall5 reports
- BD Vacutainer® 365047Becton, Dickinson U.K. Limited · Tubes, vials, systems, serum separators, blood collection9 reports
- BD Vacutainer® 365044Becton, Dickinson U.K. Limited · Tubes, vials, systems, serum separators, blood collection6 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 79223).
- Z-1139-2018 · TerminatedBD Vacutainer EDTA Lavender Top Tubes: Catalog Numbers: 1. 366401 (UDI: 50382903664011); 2. 366450 (UDI: 50382903664509); 3. 366643 (UDI: 50382903666435); 4. 367835 (UDI: 503829036…
- Z-1141-2018 · TerminatedBD Vacutainer EDTA Pink Top Tube; Catalog Numbers: 1. 367842 (UDI: 50382903678421); 2. 367899 (UDI: 50382903678995); 3. 368049 (UDI: 50382903680493); 4. 368589 (UDI: 50382903685894…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page