AltiVate Reverse INSTRUMENTATION, HUMERAL SOCKET SHELL… recallZ-1241-2018 · Encore Medical, Lp · initiated Jan 30, 2018

Class II Terminated All recalls posted in March 2018

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Encore Medical, Lp or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Broken or missing screws and retaining clips. This may be due to screw tolerance and material strength of the trial.

FDA root cause category: Device Design

What the company is doing

The firm disseminated their field safety notices on 01/30/2018 by email. The letter identified the affected product, problem and actions to be taken. The notices requested the trails be returned for replacement.

Recall details

Recall number
Z-1241-2018
Recall class
Class II
Status
Terminated
Initiated by the company
Jan 30, 2018
Posted by the FDA
Mar 5, 2018
Recalling firm
Encore Medical, Lp
Quantity
387 units
Product code
PHX
FDA submissions
K141990
Distribution
US Nationwide Distribution

Product description

AltiVate Reverse INSTRUMENTATION, HUMERAL SOCKET SHELL TRIAL, STANDARD, REF 804-06-152, Qty: 01, Encore Medical, LP Orthopedic surgical instrument. Product Usage: The humeral socket shell trial is used to replicate/trial the shell geometry of the AltiVate Reverse Standard Shell Humeral Stem implants. It assembles to the humeral broaches by a screw in order to replicate/trial the entire humeral stem implant geometry prior to implantation.

Code information (models, lots, serial numbers)

Lot Codes: 183312L03, 200006L01, 224643L01

Other devices under the same FDA submission

12 devices cleared under K141990 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page