VOLISTA StandOP Surgical Light recallZ-1349-2018 · Getinge US Sales LLC · initiated Feb 7, 2018
Class II Terminated All recalls posted in April 2018
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Getinge US Sales LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The manufacturer received complaints indicating that the device's central handle holder detaches easily from the handle holder ring on the light head during clinical use. In the event of such detachment, particles from the handle holder ring could fall into the sterile field during a procedure.
FDA root cause category: Component design/selection
What the company is doing
The firm, Getinge, sent an "URGENT MEDICAL DEVICE RECALL" Notice dated 2/7/2018 to all affected consignees via mail. The notice described the product, problem and actions to be taken. The consignees were instructed to do the following: 1) Ensure that all users of the device are made aware of this field correction. 2) Check the device serial number against the list of affected serial numbers. 3) Contact the recalling firm to arrange for servicing of all affected units. Getinge Customer Service: (800) 950-9912, 8AM - 5PM (EST) Monday to Friday. 4) Complete and sign the enclosed customer response form, and return it by fax to: (973) 860-1675 or by email to: [email protected]. If you have any questions, please contact your Getinge representative or call Getinge Customer Service at (800)950-9912 (Select option 1) between 8:00 AM and 5:00 PM (EST).
Recall details
- Recall number
- Z-1349-2018
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Feb 7, 2018
- Posted by the FDA
- Apr 10, 2018
- Recalling firm
- Getinge US Sales LLC
- Quantity
- 5214
- Product code
- FSY
- FDA submissions
- K130513
- Distribution
- Worldwide Distribution: US (nationwide) to states of: AR, AZ, CA, CO, FL, ID, IA, KS, KY, ME, MO, NE, NY, OR, TN, TX, UT, and WA; and countries of: Argentina, Australia, Austria, Belgium, Bolivia, Brazil, Cameroon, Canada, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, Germany, Ghana, Hong Kong, Hungary, India, Iran, Iraq, Italy, Japan, Jordan, Kenya, Latvia, Lebanon, Luxembourg, Malaysia, Mexico, Namibia, Netherlands, Nigeria, Oman, Pakistan, Panama, Paraguay, Poland, Portugal, South Korea, Romania, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Trinidad-Tobago, Turkey, UAE, UK, Yemen and Zimbabwe.
Product description
VOLISTA StandOP Surgical Light; Model numbers VLT600 SF AIM STP, VLT600 DF AIM STP, Ref. codes ARD568811901, ARD568811911 ARD568811951, ARD568811961 The device is designed to illuminate the body of a patient during surgical operations, diagnostics and treatment. These surgical lights may be installed in surgical suites, examining rooms, doctors' surgeries and outpatient consultations.
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
Model VLT600 SF AIM STP: Ref Code: ARD568811901: Serial Numbers: 21001 21002 22000 Ref Code: ARD568811911: Serial Numbers: 25002 31000 31001 31002 31003 31004 31006 31008 31009 31010 31012 31017 31018 31019 31020 31021 31022 31023 31024 31025 31026 35000 35001 35011 35012 35013 35014 35015 35017 35018 35019 35020 35021 35022 35023 35024 35025 35026 35027 35028 35029 35030 35031 35032 35033 38000 38001 38002 38003 38005 38009 38011 38012 38013 38014 38015 38016 38017 38018 38019 38020 38021 38022 38023 38024 38025 38026 38027 500000 500001 500002 500003 500004 500005 Model VLT600 DF AIM STP: Ref Code: ARD568811951 (Not in distribution in the U.S.) Serial Numbers: 20502 20503 Ref Code: ARD568811961 Serial Numbers: 25005 25006 25007 28003 28004 31000 31001 31002 31003 31004 31005 31006 31008 31009 31011 31012 31013 31014 31015 31016 31017 31019 31020 31030 31031 31039 31040 31041 31042 31043 31044 31045 31046 31047 31048 31049 31050 31051 31052 31053 31054 31055 31056 38025 38026 38027 38028 500001 500002
Other devices under the same FDA submission
53 devices cleared under K130513 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- ROLITE VOLISTA STANDOP ARDROL309022AMaquet Sas · Light, surgical, ceiling mounted2 recalls0 reports
- ROLITE VOLISTA STANDOP ARDROL309018AMaquet Sas · Lamp, surgical2 recalls0 reports
- ROLITE VOLISTA STANDOP ARDROL309023AMaquet Sas · Lamp, surgical2 recalls0 reports
- STANDOP VOLISTA® ARDVST229003AMaquet Sas · Light, surgical, ceiling mounted1 recall2 reports
- STANDOP VOLISTA® ARDVST229002AMaquet Sas · Light, surgical, ceiling mounted1 recall2 reports
- STANDOP VOLISTA® ARDVST229037AMaquet Sas · Lamp, surgical1 recall1 report
- ROLITE VOLISTA ACCESS ARDROL309024AMaquet Sas · Light, surgical, ceiling mounted1 recall0 reports
- ROLITE VOLISTA ACCESS ARDROL309028AMaquet Sas · Light, surgical, ceiling mounted1 recall0 reports
- ROLITE VOLISTA ACCESS ARDROL309025AMaquet Sas · Lamp, surgical1 recall0 reports
- ROLITE VOLISTA ACCESS ARDROL309029AMaquet Sas · Light, surgical, ceiling mounted1 recall0 reports
- ROLITE VOLISTA STANDOP ARDROL309016AMaquet Sas · Lamp, surgical1 recall0 reports
- ROLITE VOLISTA STANDOP ARDROL309017AMaquet Sas · Lamp, surgical1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page