Skip to content

STANDOP VOLISTA® ARDVST229002ARecalls & reported problems

Maquet Sas · VST66SF AIM 1011 · Class 2

All models of STANDOP VOLISTA (23)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 03700712401764), received between Sep 19, 2022 and Sep 17, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Corroded” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

18 other recalls cover devices cleared under the same FDA submission (K130513); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 6, 2022Class IIOpen, Classified

    Paint particles may detach and fall in the operating field. In the worst case scenario, infection, organ irritation, hypersensitivity reaction, and tissue granuloma may result if paint particles fall during surgical operation into the patient body or in the operating field.

    Z-1308-2022 · Root cause: Process change control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Corroded2
  • Peeled/Delaminated2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 17, 2024Corroded; Peeled/Delaminated; Corroded; Peeled/Delaminated

    ON 2ND SEPTEMBER, 2024 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS - STANDOP VOLISTA®. IT WAS STATED AND CONFIRMED BY PHOTOGRAPHIC EVIDENCE THAT THE CORROSION HAS BUILT UP AND THE PAINT WAS PEELING FROM SPRING ARM, MAIN TUBE, SUSPENSION ARM. THERE WAS NO INJURY REPORTED, HOWEVER, WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTICLES FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE CONTAMINATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20236919 ↗
  • MalfunctionSep 19, 2022Corroded; Peeled/Delaminated; Corroded; Peeled/Delaminated

    ON (B)(6) 2022 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS, VOLISTA STANDOP. ACCORDING TO PHOTOGRAPHIC EVIDENCE, THE PAINT WAS PEELING AND RUST OCCURED ON THE SUSPENSION ARM. THERE WAS NO INJURY REPORTED, HOWEVER WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTICLES FALLING OFF INTO STERILE FIELD OR DURING PROCEDURE MAY CAUSE CONTAMINATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15445168 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K130513. One submission often covers many models, so check the product and lot details in each record.

  • Z-0758-2024 · Nov 8, 2023 · Open, Classified
    Maquet Volista, Volista Access, Volista Access II, Volista Standop OR Light System, Model Numbers and Catalog Numbers ROL B FOR VLT400SF STP - ARDROL309001A, ROL B FOR VLT400SF STP
  • Z-1863-2019 · May 28, 2019 · Terminated
    Maquet VOLISTA StandOP Surgical Lights- Product Code: ARDVST229036A Model:VST66SF
  • Z-1852-2019 · May 28, 2019 · Terminated
    Maquet VOLISTA StandOP Surgical Lights- Product Code: ARD568811901 Model: VLT600SF AIM STP
  • Z-1858-2019 · May 28, 2019 · Terminated
    Maquet VOLISTA StandOP Surgical Lights- Product Code: ARDVST229001A Model: VST60DF AIM
  • Z-1856-2019 · May 28, 2019 · Terminated
    Maquet VOLISTA StandOP Surgical Lights- Product Code: ARD568812910 Model: VLT400SF STP
  • Z-1859-2019 · May 28, 2019 · Terminated
    Maquet VOLISTA StandOP Surgical Lights- Product Code: ARDVST229002A Model: VST66SF AIM
  • Z-1857-2019 · May 28, 2019 · Terminated
    Maquet VOLISTA StandOP Surgical Lights- Product Code: ARDVST229000A Model: VST60SF AIM
  • Z-1855-2019 · May 28, 2019 · Terminated
    Maquet VOLISTA StandOP Surgical Lights- Product Code: ARD568811961 Model: VLT600DF AIM STP
  • Z-1860-2019 · May 28, 2019 · Terminated
    Maquet VOLISTA StandOP Surgical Lights- Product Code: ARDVST229003A Model: VST66DF AIM
  • Z-1854-2019 · May 28, 2019 · Terminated
    Maquet VOLISTA StandOP Surgical Lights- Product Code: ARD568811911 Model: VLT600SF AIM STP
  • Z-1861-2019 · May 28, 2019 · Terminated
    Maquet VOLISTA StandOP Surgical Lights- Product Code: ARDVST229034A Model: VST60SF
  • Z-1864-2019 · May 28, 2019 · Terminated
    Maquet VOLISTA StandOP Surgical Lights- Product Code: ARDVST229037A Model: VST66DF
  • Z-1853-2019 · May 28, 2019 · Terminated
    Maquet VOLISTA StandOP Surgical Lights- Product Code: ARD568811910 Model:VLT600SF STP
  • Z-1862-2019 · May 28, 2019 · Terminated
    Maquet VOLISTA StandOP Surgical Lights- Product Code: ARDVST229035A Model:VST60DF
  • Z-2502-2018 · Jun 14, 2018 · Terminated
    VOLISTA StandOP Surgical Light, ARD568811961 VLT600DF AIM STP; ARD568821961 VLT600DF AIM STP QL; ARD568811911 VLT600SF AIM STP; ARD568821911 VLT600SF AIM STP QL; ARD568821910 VLT6

…and 3 more.

Other Light, surgical, ceiling mounted devices

All Light, surgical, ceiling mounted devices →

Sources and updates