Capio recallZ-1620-2018 · Boston Scientific Corporation · initiated Feb 12, 2018
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Boston Scientific Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential for Capio sutures to break and /or detach.
FDA root cause category: Device Design
What the company is doing
Boston Scientific sent Urgent Medical Device Correction letters, dated February 12, 2018, to customers via FED-EX two day delivery. Customers were informed of the updates to the affected product instructions for use and advised to read the notification carefully and immediately post the information in a visible location near the product to ensure it is easily accessible to all users of the device. Also, the notification should be passed on to any health professional of the customer's organization that needs to be aware and to any organization where the potentially affected devices have been transferred. Customers with questions should contact their local BSC (Boston Scientific) Representative. **No product is being recalled and you are not required to return product to Boston Scientific** There is no impact to previously implanted mesh devices.
Recall details
- Recall number
- Z-1620-2018
- Status
- Terminated
- Initiated by the company
- Feb 12, 2018
- Posted by the FDA
- May 4, 2018
- Recalling firm
- Boston Scientific Corporation
- Quantity
- 4,144
- Product code
- PWI
- FDA submissions
- K932553
- Distribution
- Worldwide Distribution -- US, Brazil, Guatemala, Ecuador, Malaysia, Czech Republic, United Kingdom, Argentina, Honduras, Peru, Singapore, Denmark, Iceland, Chile, Nicaragua, Venezuela, Japan, Egypt, Israel, Canada, Panama, Taiwan, United Arab Emirates, Finland, Ireland, Mexico, Dominican Republic, Hong Kong, Austria, France, Italy, Costa Rica, Trinidad and Tobago, Australia, Bahrain, Georgia, Islamic Republic of Iran, El Salvador, Colombia, New Zealand, Belgium, Germany, Jordan, Norway, Russia, Pakistan, Slovakia, Portugal, South Africa, Oman, Saudi Arabia, Poland, Slovenia, Qatar, Spain, Kenya, Lebanon, Mauritius, Netherlands, Sweden, Turkey, Kuwait, Malta, Namibia, Romania, Switzerland, and Greece.
Product description
Capio Standard Box 4 (Capio PC); Model #M0068312321. Intended for use in general suturing applications during open or endoscopic surgery to assist in the placement of suture material in tissues at the operative site.
Code information (models, lots, serial numbers)
UDI:08714729257622 Batch numbers: 17817547 19050231 20639026 17925410 19225758 20717524 18139200 19335465 21081753 18363737 19437994 21258781 18615714 19698427 21386211 18826587 19983956 20341449 18955689 20341449
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Capio M0068312320Boston Scientific Corporation · Laparoscope, general & plastic surgery1 recall14 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 79608).
- Z-1616-2018 · TerminatedCapio SLIM Suture Capturing Device, Model #M0068318250. Intended for use in general suturing applications during open surgery to assist in the placement of suture material in tiss…
- Z-1617-2018 · TerminatedCapio SLIM Box 5, Model #M0068318261; Intended for use in general suturing applications during open surgery to assist in the placement of suture material in tissues at the operat…
- Z-1618-2018 · TerminatedCapio CL; Model #M0068311350. Indicated for the placement of suture in a variety of Transvaginal procedures.
- Z-1619-2018 · TerminatedCapio Open Access Box 4 (Capio OPC); Model #M0068311251. Intended for use in general suturing applications during open or endoscopic surgery to assist in the placement of suture m…
- Z-1621-2018 · TerminatedCapio RP; Model #M0068321010. Intended for use in general suturing applications during open surgery to assist in the placement of suture material in tissues at the operative site.
- Z-1622-2018 · TerminatedUphold LITE w/ Capio SLIM; Model #M0068318170. Indicated for tissue reinforcement in women with pelvic organ prolapse.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page