Uphold LITE w/ Capio SLIM recallZ-1622-2018 · Boston Scientific Corporation · initiated Feb 12, 2018

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Boston Scientific Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential for Capio sutures to break and /or detach.

FDA root cause category: Device Design

What the company is doing

Boston Scientific sent Urgent Medical Device Correction letters, dated February 12, 2018, to customers via FED-EX two day delivery. Customers were informed of the updates to the affected product instructions for use and advised to read the notification carefully and immediately post the information in a visible location near the product to ensure it is easily accessible to all users of the device. Also, the notification should be passed on to any health professional of the customer's organization that needs to be aware and to any organization where the potentially affected devices have been transferred. Customers with questions should contact their local BSC (Boston Scientific) Representative. **No product is being recalled and you are not required to return product to Boston Scientific** There is no impact to previously implanted mesh devices.

Recall details

Recall number
Z-1622-2018
Status
Terminated
Initiated by the company
Feb 12, 2018
Posted by the FDA
May 4, 2018
Recalling firm
Boston Scientific Corporation
Quantity
24,126
Product code
PWI
FDA submissions
K122459
Distribution
Worldwide Distribution -- US, Brazil, Guatemala, Ecuador, Malaysia, Czech Republic, United Kingdom, Argentina, Honduras, Peru, Singapore, Denmark, Iceland, Chile, Nicaragua, Venezuela, Japan, Egypt, Israel, Canada, Panama, Taiwan, United Arab Emirates, Finland, Ireland, Mexico, Dominican Republic, Hong Kong, Austria, France, Italy, Costa Rica, Trinidad and Tobago, Australia, Bahrain, Georgia, Islamic Republic of Iran, El Salvador, Colombia, New Zealand, Belgium, Germany, Jordan, Norway, Russia, Pakistan, Slovakia, Portugal, South Africa, Oman, Saudi Arabia, Poland, Slovenia, Qatar, Spain, Kenya, Lebanon, Mauritius, Netherlands, Sweden, Turkey, Kuwait, Malta, Namibia, Romania, Switzerland, and Greece.

Product description

Uphold LITE w/ Capio SLIM; Model #M0068318170. Indicated for tissue reinforcement in women with pelvic organ prolapse.

Code information (models, lots, serial numbers)

UDI:08714729839200 Batch numbers:  27019 56827 21250249 27898 56829 21271669 28713 57715 21336724 28974 57716 21377535 29299 58081 21402115 30503 58083 21432026 31359 58944 21447170 31516 58945 21496403 32828 60407 ML00002769 32829 60409 ML00002810 33791 61098 ML00002857 33792 61420 ML00002913 34053 61421 ML00002964 36070 61422 ML00002999 37918 61425 ML00003034 39009 61426 ML00003091 39700 62156 ML00003142