Skip to content

Prestige LP Streamlined Instruments recallZ-1641-2018 · Medtronic Sofamor Danek USA Inc · initiated Mar 13, 2018

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medtronic Sofamor Danek USA Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The firm received complaints of drill bits breaking during use with the Prestige LP(TM) Cervical Disc System. Subsequent investigation demonstrated the interaction of the drill guide and the drill bit may lead to drill bit breakage.

FDA root cause category: Under Investigation by firm

What the company is doing

Medtronic distributed a voluntary removal notice via FedEx courier service to impacted U.S. consignees on 13MARCH 2018. The notice requested the following: "Instructions to Customers: 1.) Please locate and remove the impacted product from normal storage locations. Do not use this product. 2.) Complete the enclosed Customer Confirmation Form and return via fax to 763-367-8134, or via e-mail to [email protected]. 3.) As applicable, contact your Medtronic Sales Representative to coordinate the return of affected product."

Recall details

Recall number
Z-1641-2018
Status
Terminated
Initiated by the company
Mar 13, 2018
Recalling firm
Medtronic Sofamor Danek USA Inc
Quantity
231 units
Product code
MJO
FDA submissions
P090029
Distribution
US Nationwide Distribution

Product description

Medtronic Prestige LP(TM) Cervical Disc System with Streamlined Instruments, DRILL GUIDES: a) REF 6975205, 5mm b) REF 6975206, 6mm c) REF 6975207, 7mm Product Usage: The PRESTIGE LP reusable instruments are intended for the surgical implantation of the PRESTIGE LP Cervical Disc. Medtronic reusable instruments must be sterilized prior to initial use and must be cleaned and sterilized prior to each re-use.

Code information (models, lots, serial numbers)

a) REF 6975205, 5mm; UDI 00643169752979 Lot Number ID17K018 b) REF 6975206, 6mm; UDI 00643169752962  Lot Number ID17J009 c) REF 6975207, 7mm; UDI 00643169752955 Lot Numbers: ID17K007, ID17L035

Devices named in this recall

3 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

14 devices cleared under P090029 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates