PRESTIGE® LP Cervical Disc System 6973060Recalls & reported problems

Medtronic Sofamor Danek, Inc. · INNER SHAFT 6973060 6MM INSERTER US · Class 3

All models of PRESTIGE LP Cervical Disc System (11)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00613994714848), received between Jan 13, 2020 and Aug 31, 2026 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Mechanical Problem” (3 reports).

1 other recall cover devices cleared under the same FDA submission (P090029); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Mechanical Problem3
  • Degraded1
  • Material Twisted/Bent1
  • Mechanics Altered1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Known Impact Or Consequence To Patient2

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 31, 2026Mechanical Problem; Material Twisted/Bent

    Information was received from multiple sources (Manufacturer Representative, Healthcare Provider) regarding a device used for cervic al arthroplasty Spinal therapy at C5/6. It was reported that the inserters had difficulty seating all four inserter dimples into the interbody. Three of four dimples could be seated, but a snug fit could not be achieved before tightening the locking knob to secure the implant. There was difficulty attaching and locking the implant, including tightening the locking knob. The inserter prongs or dimples were suspected to be bent or unaligned; however, no visible damage was confirmed. There was no patient involved.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50516719 ↗
  • MalfunctionFeb 8, 2026Degraded; Mechanical Problem

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING ANTERIOR CERVICAL DISCECTOMY AND FUSION (ACDF) PROCEDURE AT LEVELS 5-7 FOR A PRE-OPERATIVE DIAGNOSIS OF HERNIATED NUCLEUS PULPOSUS. IT WAS REPORTED THAT THREE INSERTERS HAVE BEEN BROKEN AND NO LONGER ATTACHED TO THE IMPLANTS. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. ADDITIONAL INFORMATION WAS RECEIVED FROM THE MANUFACTURER REPRESENTATIVE THAT THERE WAS AN INSTRUMENT INCOMPATIBILITY WHICH MIGHT BE DUE TO WEAR ISSUE AND NOT BREAKAGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24288233 ↗
  • MalfunctionNov 3, 2020Mechanics Altered; Mechanics Altered

    INFORMATION WAS RECEIVED FROM THE HEALTHCARE PROFESSIONAL (HCP) VIA THE MANUFACTURER REPRESENTATIVE REGARDING AN EVENT THAT OCCURRED DURING A SPINAL ADR PROCEDURE IN A PATIENT. IT WAS REPORTED THAT THE IMPLANT FELL OFF THE INSERTER. NO PATIENT INJURY / COMPLICATION WAS REPORTED. PRODUCT WILL BE RETURNED, AND IT WAS REPLACED WITH A MEDTRONIC PRODUCT.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient

    MAUDE report 10776562 ↗
  • MalfunctionJan 13, 2020Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THAT THE PATIENT UNDERWENT CERVICAL DISC ARTHROPLASTY AT C6-7 DUE TO DEGENERATIVE DISK DISEASE(DDD). INTRA OP, INSERTER DID NOT RETAIN THE IMPLANT PROPERLY WHEN IMPACTED WITH THE MALLET. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9579882 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P090029. One submission often covers many models, so check the product and lot details in each record.

  • Z-1641-2018 · Mar 13, 2018 · Terminated
    Medtronic Prestige LP(TM) Cervical Disc System with Streamlined Instruments, DRILL GUIDES: a) REF 6975205, 5mm b) REF 6975206, 6mm c) REF 6975207, 7mm Product Usage: The PRESTIGE

Other PROSTHESIS, INTERVERTEBRAL DISC devices

All PROSTHESIS, INTERVERTEBRAL DISC devices →