PRESTIGE® LP Cervical Disc System 6973060Recalls & reported problems
Medtronic Sofamor Danek, Inc. · INNER SHAFT 6973060 6MM INSERTER US · Class 3
All models of PRESTIGE LP Cervical Disc System (11)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00613994714848), received between Jan 13, 2020 and Aug 31, 2026 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Mechanical Problem” (3 reports).
1 other recall cover devices cleared under the same FDA submission (P090029); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Mechanical Problem3
- Degraded1
- Material Twisted/Bent1
- Mechanics Altered1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- No Known Impact Or Consequence To Patient2
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 31, 2026Mechanical Problem; Material Twisted/Bent
Information was received from multiple sources (Manufacturer Representative, Healthcare Provider) regarding a device used for cervic al arthroplasty Spinal therapy at C5/6. It was reported that the inserters had difficulty seating all four inserter dimples into the interbody. Three of four dimples could be seated, but a snug fit could not be achieved before tightening the locking knob to secure the implant. There was difficulty attaching and locking the implant, including tightening the locking knob. The inserter prongs or dimples were suspected to be bent or unaligned; however, no visible damage was confirmed. There was no patient involved.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50516719 ↗MalfunctionFeb 8, 2026Degraded; Mechanical Problem
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING ANTERIOR CERVICAL DISCECTOMY AND FUSION (ACDF) PROCEDURE AT LEVELS 5-7 FOR A PRE-OPERATIVE DIAGNOSIS OF HERNIATED NUCLEUS PULPOSUS. IT WAS REPORTED THAT THREE INSERTERS HAVE BEEN BROKEN AND NO LONGER ATTACHED TO THE IMPLANTS. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. ADDITIONAL INFORMATION WAS RECEIVED FROM THE MANUFACTURER REPRESENTATIVE THAT THERE WAS AN INSTRUMENT INCOMPATIBILITY WHICH MIGHT BE DUE TO WEAR ISSUE AND NOT BREAKAGE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24288233 ↗MalfunctionNov 3, 2020Mechanics Altered; Mechanics Altered
INFORMATION WAS RECEIVED FROM THE HEALTHCARE PROFESSIONAL (HCP) VIA THE MANUFACTURER REPRESENTATIVE REGARDING AN EVENT THAT OCCURRED DURING A SPINAL ADR PROCEDURE IN A PATIENT. IT WAS REPORTED THAT THE IMPLANT FELL OFF THE INSERTER. NO PATIENT INJURY / COMPLICATION WAS REPORTED. PRODUCT WILL BE RETURNED, AND IT WAS REPLACED WITH A MEDTRONIC PRODUCT.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 10776562 ↗MalfunctionJan 13, 2020Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT THE PATIENT UNDERWENT CERVICAL DISC ARTHROPLASTY AT C6-7 DUE TO DEGENERATIVE DISK DISEASE(DDD). INTRA OP, INSERTER DID NOT RETAIN THE IMPLANT PROPERLY WHEN IMPACTED WITH THE MALLET. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9579882 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P090029. One submission often covers many models, so check the product and lot details in each record.
- Z-1641-2018 · Mar 13, 2018 · TerminatedMedtronic Prestige LP(TM) Cervical Disc System with Streamlined Instruments, DRILL GUIDES: a) REF 6975205, 5mm b) REF 6975206, 6mm c) REF 6975207, 7mm Product Usage: The PRESTIGE
Other PROSTHESIS, INTERVERTEBRAL DISC devices
- Prodisc C SK PDSM5Centinel Spine, LLC · Prosthesis, Intervertebral Disc2 recalls4 reports
- Simplify® Cervical Artificial DiscNuvasive, Inc. · Prosthesis, Intervertebral Disc1 recall12 reports
- Prodisc C 09.820.055SCentinel Spine, LLC · Prosthesis, Intervertebral Disc1 recall9 reports
- Simplify® Cervical Artificial DiscNuvasive, Inc. · Prosthesis, Intervertebral Disc1 recall4 reports
- Prodisc C SK PDSXL5Centinel Spine, LLC · Prosthesis, Intervertebral Disc1 recall3 reports
- Simplify® Cervical Artificial DiscNuvasive, Inc. · Prosthesis, Intervertebral Disc1 recall3 reports
- Simplify® Cervical Artificial DiscNuvasive, Inc. · Prosthesis, Intervertebral Disc1 recall3 reports
- Simplify® Cervical Artificial DiscNuvasive, Inc. · Prosthesis, Intervertebral Disc1 recall3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page