Fabius MRI, anesthesia machine, catalog no. 8607300 recallZ-1481-2018 · Draeger Medical, Inc. · initiated Mar 1, 2018

Class I Terminated All recalls posted in May 2018

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Draeger Medical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The automatic ventilation may fail if the position detection of the motor is disturbed. If the ventilator cover is damaged, manual ventilation may also fail.

FDA root cause category: Process control

What the company is doing

On March 1, 2018, the recalling firm notified affected customers via an Urgent Medical Device Recall letter. The recalling firm stated that customers would receive replacement motors for potentially affected devices. Until the replacement occurred, customers were instructed to continue to operate their devices with the usual attention and make sure that a manual resuscitator for emergency ventilation is kept ready according to the instructions for use. Customers were also supplied with a reply card to indicate that they had received the notification. Customers with questions about the letter were directed to call Michael Kelhart between 8AM-4:30PM EST at 1-800-437-2437 (press 1 at the prompt, then 32349). Questions regarding the operation and/or servicing of the Draeger Fabius anesthesia machine were directed to DraegerService Technical Support between the hours of 8:00 AM 8:00 PM EST at 1-800-437-2437 (press 4 at the prompt then 4 again).

Recall details

Recall number
Z-1481-2018
Recall class
Class I (most serious)
Status
Terminated
Initiated by the company
Mar 1, 2018
Posted by the FDA
May 8, 2018
Recalling firm
Draeger Medical, Inc.
Quantity
8
Product code
BSZ
FDA submissions
K072884
Distribution
US Nationwide Distribution to accounts in 21 states: AZ, CA, CO, ID, IL, KS, MA, MI, MS, NC, NY, OH, OK, PA, RI, TN, TX, UT, VA, WA, WI; and PR.

Product description

Fabius MRI, anesthesia machine, catalog no. 8607300 Product Usage: The products are inhalation anesthesia machines for use in operating, induction and recovery rooms. They may be used with O2, N2O, and AIR supplied by a medical gas pipeline system or by externally mounted gas cylinders. Fabius series anesthesia machines are equipped with an electrically driven and electronically controlled ventilator and monitors for airway pressure (P), volume (V), and inspiratory oxygen concentration (FiO2). This device can also be used in MRI scanner rooms with magnets 1.5 tesla and 3 tesla at a fringe field strength of 40 mtesla or less.

Code information (models, lots, serial numbers)

Serial no: ASKM-0020 ASKM-0011 ASKL-0005 ASKL-0007 ASKL-0010 ASKM-0031 ASKM-0032 ASKM-0023

Other devices under the same FDA submission

1 device cleared under K072884 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.