1)BOX PST01199 BARIATRIC TRISTAPLE (Item Number:… recallZ-1849-2018 · Covidien Medtronic · initiated Apr 26, 2018
Class II Terminated All recalls posted in May 2018
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Covidien Medtronic or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
FDA root cause category: Process control
What the company is doing
On April 26, 2108, all consignees were issued were issued Urgent Medical Device Recall notices and Return Forms via Federal Express or certified mail. On May 22, 2018, all consignees received an UPDATED URGENT MEDICAL DEVICE RECALL notice. The updated notice included NEW product which has been processed under FDA Ref # 80423. The notice informs customers of the potential for use of a device with a missing sled and the actions they should take to discontinue use of the product and return all inventory of the affected lots to Medtronic (Covidien), 195 McDermott Road, North Haven, CT 06473, Attention: Field Returns Department. All in-control inventory and devices returned from customers will be reworked. **FOR INTERNATIONAL CUSTOMERS** Please be aware that the recalled units are packed within various convenience kits as identified on the Attachment B provided with your customer letter. Customers are also encouraged to EMAIL OR FAX THE ACKNOWLEDGEMENT FORM TO: - "Product purchased directly from Medtronic": [email protected] or fax to (800) 895-6140. - "Product purchased through distributor": [email protected] or fax it to (203) 492-7719.
Recall details
- Recall number
- Z-1849-2018
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Apr 26, 2018
- Posted by the FDA
- May 16, 2018
- Recalling firm
- Covidien Medtronic
- Quantity
- 169,775 units in total
- Product code
- GDW
- FDA submissions
- K111825
- Distribution
- Worldwide and US Nationwide
Product description
(1)BOX PST01199 BARIATRIC TRISTAPLE (Item Number: PST01199) (2) BOX PST03403 ALTERNATIVE APPENDIX KIT (Item Number: PST03403) (3) BOX PST03533 KIT APP STAPLER (Item Number: PST03403) Product Usage: The Endo GIA universal and Endo GIA universal XL staplers place two, triple-staggered rows of titanium staples and simultaneously divides the tissue between the two, triple-staggered rows. The size of the staples is determined by the selection of the 2.0 mm, 2.5 mm, 3.5 mm or 4.8 mm single use loading unit. The Endo GIA universal staplers will accommodate any of the single use loading unit sizes that are available in the 30 mm, 45 mm and 60 mm lines.
Code information (models, lots, serial numbers)
Lot Numbers: (1) 0214109631, 0214909770 (2) 0214975597 (3) 0214860061
Other devices under the same FDA submission
77 devices cleared under K111825 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Endo GIA EGIA60AMTCovidien LP · STAPLE, IMPLANTABLE19 deaths10,680 reports
- Endo GIA Ultra EGIAUSTNDCovidien LP · STAPLE, IMPLANTABLE5 deaths9,730 reports
- Endo GIA Ultra EGIAUSHORTCovidien LP · STAPLE, IMPLANTABLE2 deaths3,011 reports
- Endo GIA EGIA45AMTCovidien LP · STAPLE, IMPLANTABLE10 deaths2,965 reports
- Endo GIA Ultra EGIAUXLCovidien LP · STAPLE, IMPLANTABLE2,909 reports
- Endo GIA EGIA45AVMCovidien LP · STAPLE, IMPLANTABLE7 deaths1,583 reports
- Endo GIA EGIA60AVMCovidien LP · STAPLE, IMPLANTABLE9 deaths1,438 reports
- GIA GIA8038SCovidien LP · STAPLE, IMPLANTABLE12 deaths1,237 reports
- Endo GIA 030458Covidien LP · STAPLE, IMPLANTABLE1 recall1 death948 reports
- Endo GIA EGIA60AXTCovidien LP · STAPLE, IMPLANTABLE4 deaths987 reports
- Endo GIA 030415Covidien LP · STAPLE, IMPLANTABLE973 reports
- Endo GIA 030449Covidien LP · STAPLE, IMPLANTABLE2 deaths716 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 79974).
- Z-1814-2018 · TerminatedEndo GIA" Auto Suture" Universal Articulating Loading Unit 30mm- 2.0mm ( Item Code 030450) Product Usage: The Endo GIA universal and Endo GIA universal XL staplers place two, t…
- Z-1815-2018 · TerminatedEndo GIA" Auto Suture" Universal Articulating Loading Unit 30mm - 2.5mm ( Item Code 030451) Product Usage: The Endo GIA universal and Endo GIA universal XL staplers place two, …
- Z-1816-2018 · TerminatedEndo GIA" Auto Suture" Universal Articulating Loading Unit 30mm - 3.5mm ( Item Code 030452) Product Usage: The Endo GIA universal and Endo GIA universal XL staplers place two, …
- Z-1817-2018 · TerminatedEndo GIA" Auto Suture" Universal Articulating Loading Unit 45mm - 2.5mm( Item Code 030454) Product Usage: The Endo GIA universal and Endo GIA universal XL staplers place two,…
- Z-1818-2018 · TerminatedEndo GIA" Auto Suture" Universal Articulating Loading Unit 45mm- 3.5mm( Item Code 030455) Product Usage: The Endo GIA universal and Endo GIA universal XL staplers place two, tr…
- Z-1819-2018 · TerminatedEndo GIA" Auto Suture" Universal Articulating Loading Unit 60mm -2.5mm ( Item Code 030457) Product Usage: The Endo GIA universal and Endo GIA universal XL staplers place two, t…
- Z-1820-2018 · TerminatedEndo GIA" Auto Suture" Universal Articulating Loading Unit 60mm - 3.5mm ( Item Code 030458) Product Usage: The Endo GIA universal and Endo GIA universal XL staplers place two, …
- Z-1821-2018 · TerminatedEndo GIA" 30mm Gray Curved Tip Articulating Vascular Reload with Tri-Staple" Technology( Item Code EGIA30CTAV) Product Usage: The Endo GIA universal and Endo GIA universal XL st…
- Z-1822-2018 · TerminatedEndo GIA" 30mm Gray Curved Tip Articulating Vascular Reload with Tri-Staple" Technology( Item Code EGIA30AV) Product Usage: The Endo GIA universal and Endo GIA universal XL stap…
- Z-1823-2018 · TerminatedEndo GIA" 45mm Gray Curved Tip Articulating Vascular Reload with Tri-Staple" Technology( Item Code EGIA45CTAV) Product Usage: The Endo GIA universal and Endo GIA universal XL st…
- Z-1824-2018 · TerminatedCovidien" Best Practices" Procedure Kits containing Auto Suture" Endo GIA" Devices Product Usage: The Endo GIA universal and Endo GIA universal XL staplers place two, triple-s…
- Z-1825-2018 · TerminatedBOX KIT1095V KIT OBESITAS (Item Code KIT1095V) Product Usage: The Endo GIA universal and Endo GIA universal XL staplers place two, triple-staggered rows of titanium staples and s…
- Z-1826-2018 · Terminated(1) BOX 01KBAR006116V PROC BYPASS 01KBAR0061(Item Code 01KBAR006116V PROC BYPAS) (2) BOX 1073076GASB0 BBR STVEIT GASTRIC B (Item Code BOX1073076GASB0) (3) BOX BOX1073076GASB…
- Z-1828-2018 · TerminatedBOX1073076GASB1 BBR GASTRIC BYPASS S(Item Code BOX1073076GASB1 BBR GASTRIC BYPASS S) Product Usage: The Endo GIA universal and Endo GIA universal XL staplers place two, triple-st…
- Z-1829-2018 · Terminated(1) BOX BOX1073180 APPE1 APPENDIX KIT ZAMS (Item Code BOX1073180) (2)BOX BOX1073180APPE1 APPENDIX KIT ZAMS (Item Code BOX1073180APPE1 APPENDIX KIT ZAMS) Product …
- Z-1830-2018 · Terminated(1)BOX HEMIGX1 ZESTAW DO HEMICOLECTOMII (Item Number HEMIGX1 ZESTAW DO HEMICOLECTOMII) (2) BOX LAPPH ZASTAW DO PRAWEJ HEMIKOLEKTOMI (item number: LAPPH ZASTAW DO PRAWEJ HEMIKO…
- Z-1831-2018 · TerminatedKIT00486R OBESIDAD D. WORWARLD 2 X1 (Item Number KIT00486R) Product Usage: The Endo GIA universal and Endo GIA universal XL staplers place two, triple-staggered rows…
- Z-1832-2018 · TerminatedBOX KIT00713 KIT BY PASS Product Usage: The Endo GIA universal and Endo GIA universal XL staplers place two, triple-staggered rows of titanium staples and simultaneously divides…
- Z-1833-2018 · TerminatedBOX KIT00714 KIT SLEEVE Product Usage: The Endo GIA universal and Endo GIA universal XL staplers place two, triple-staggered rows of titanium staples and simultaneously divides …
- Z-1834-2018 · TerminatedBOX KIT1171C KIT LAP CYSTECTOMY Product Usage: The Endo GIA universal and Endo GIA universal XL staplers place two, triple-staggered rows of titanium staples and simultaneously …
- Z-1837-2018 · TerminatedKIT-DE-0080 BYPASS KIT LOHNE X1( Item KIT-DE-0080) Product Usage: The Endo GIA universal and Endo GIA universal XL staplers place two, triple-staggered rows of titanium staples …
- Z-1838-2018 · Terminated(1)KIT4184 APEN LOBECTOMY KIT X0 Item Number: KIT4184 APEN LOBECTOMY KIT XO (2) BOX VATSLOB11 ZESTAW DO VATS LOBECTOMII (Item Number: VATSLOB11, VATSLOB11 ZESTAW DO VAT…
- Z-1839-2018 · TerminatedBOX KITDE0108 BYPASS KIT LUEBECK ( KITDE0108) Product Usage: The Endo GIA universal and Endo GIA universal XL staplers place two, triple-staggered rows of titanium staples and s…
- Z-1840-2018 · Terminated(1) BOX KITDE0143 BYPASS BOX (Item Number: KITDE0143, KITDE0143 BYPASS BOX) (2) BOX KITDE0154 BYPASS KH NORDWEST (Item Number: KITDE0154, KITDE0154 BYPASS KH NORDWES…
- Z-1842-2018 · Terminated(1)BOX KITDE0183 APPENDIX LIPPE DETMOLD (KITDE0183) (2) KIT-DE-0083 APPENEKTOMIE KIT NAUENX1 (Item Number: KIT-DE-0083,KIT-DE-0083 APPENEKTOMIE KIT NAUENX1 Product Usage: The En…
- Z-1843-2018 · Terminated(1)LAPBARLECZNA ZESTAW BARIATRYCZNY LECZNA (2) KITBAR021- BARIATRIC KIT X1 Item Code:KITBAR021- BARIATRIC KIT X0 (3)BOX SGGLOGX1 ZESTAW DO BARIATRII Item Code: SGGLOGX1 ZESTA…
- Z-1844-2018 · TerminatedBOX LAPIMBGP1 ZESTAW DO MGBP (LAPIMBGP1) Product Usage: The Endo GIA universal and Endo GIA universal XL staplers place two, triple-staggered rows of titanium staples and simul…
- Z-1847-2018 · Terminated(1) BOX LAPZD ZESTAW DO SG (2) BOX LAPISG1 ZESTAW DO SG (3) BOX LAPISGW ZESTAW DO SG Product Usage: The Endo GIA universal and Endo GIA universal XL staplers place …
- Z-1848-2018 · TerminatedBOX MGBGLOGX1 ZESTAW MINI GASTRIC BYPASS Product Usage: The Endo GIA universal and Endo GIA universal XL staplers place two, triple-staggered rows of titanium staples and simult…
- Z-1850-2018 · TerminatedMGBX ZESTAW GB LS X1 Product Usage: The Endo GIA universal and Endo GIA universal XL staplers place two, triple-staggered rows of titanium staples and simultaneously divides the…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page