STar Drive MA recallZ-2351-2018 · FHC, Inc. · initiated May 10, 2018
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from FHC, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.
FDA root cause category: Device Design
What the company is doing
Urgent Medical Device Recall notification letters were sent to customers on 5/10/18. The letters included the following instructions: FHC, Inc. Field Representatives will review your inventory and replace effected Depth Stop Adapters you have on hand with an improved design. In the interim, we advise that you do not over-tighten the Depth Stop Adapter screw. If low impedances or a short circuit is found, physicians should replace the lead prior to use and report the issue to FHC, Inc. or your local field rep.
Recall details
- Recall number
- Z-2351-2018
- Status
- Terminated
- Initiated by the company
- May 10, 2018
- Recalling firm
- FHC, Inc.
- Quantity
- 2952 total
- Product code
- HAW
- FDA submissions
- K092562
- Distribution
- The devices were distributed to the following US states: AL, AR, AZ, CA, CO, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The devices were distributed to the following foreign countries: Argentina, Australia, Canada, Columbia, Denmark, Ecuador, France, Germany, Great Britain, Italy, Japan, Norway, Romania, South Africa, Switzerland, Saudi Arabia, Taiwan.
Product description
microTargeting" STar" Drive System (Manual), Product Number 70-ZD-MA
Code information (models, lots, serial numbers)
UDI: 00873263002162
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- STar™ Drive MA 70-ZD-MA-00216Fhc, Inc. · Neurological stereotaxic Instrument1 recall1 death1 report
Product lines
Other devices under the same FDA submission
8 devices cleared under K092562 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Star™ Drive M/E FC8002-00149Fhc, Inc. · Neurological stereotaxic Instrument2 recalls0 reports
- Star™ Drive M/E FC8002-00637Fhc, Inc. · Neurological stereotaxic Instrument2 recalls0 reports
- STar™ Drive MA FC8001-00148Fhc, Inc. · Neurological stereotaxic Instrument1 recall1 report
- STar™ Drive M/E 70-ZD-ME-00215Fhc, Inc. · Neurological stereotaxic Instrument1 recall0 reports
- STar™ Drive Lead Adapter 70-CN-DB-00264Fhc, Inc. · Neurological stereotaxic Instrument1 recall0 reports
- STar™ Drive MA System ST-DS-MA-00441Fhc, Inc. · Neurological stereotaxic Instrument1 recall0 reports
- STar™ Drive M/E System ST-DS-ME-00443Fhc, Inc. · Neurological stereotaxic Instrument1 recall0 reports
- STar™ Drive MA FC8001-00635Fhc, Inc. · Neurological stereotaxic Instrument1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 80148).
- Z-2345-2018 · TerminatedmicroTargeting" Drive System, Product Number MT-DS
- Z-2346-2018 · TerminatedmicroTargeting" Drive System, Product Number MT-DS-01
- Z-2347-2018 · TerminatedmicroTargeting" Drive System, Product Number 66-ZD-MD
- Z-2348-2018 · TerminatedmicroTargeting" Drive System, Product Number 66-ZD-MD-01
- Z-2349-2018 · TerminatedmicroTargeting" STar" Drive System (Manual), Product Number ST-DS-MA
- Z-2350-2018 · TerminatedmicroTargeting" STar" Drive System (Motorized), Product Number ST-DS-ME
- Z-2352-2018 · TerminatedmicroTargeting" STar" Drive System (Motorized), Product Number 70-ZD-ME
- Z-2353-2018 · TerminatedmicroTargeting" Lead Adapter for microTargeting" Drive, Product Number 66-CN-DB
- Z-2354-2018 · TerminatedmicroTargeting" Lead Adapter for STar" Drive, Product Number 70-CN-DB
- Z-2355-2018 · TerminatedmicroTargeting" Drive System Distributed by Medtronic, Product Number FC1006
- Z-2356-2018 · TerminatedmicroTargeting" Drive System Distributed by Medtronic, Product Number 9033G0601
- Z-2357-2018 · TerminatedmicroTargeting" STar" Drive System (Manual) Distributed by Medtronic, Product Number FC8001
- Z-2358-2018 · TerminatedmicroTargeting" STar" Drive System (Motorized) Distributed by Medtronic, Product Number FC8002
- Z-2359-2018 · Terminatedmicrotargeting" Depth Stop Adapter, Product Model Number, Product Number 66-AC-DS(1.8),
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page