Biomet Knee System recallZ-2177-2018 · Zimmer Biomet, Inc. · initiated May 18, 2018
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Zimmer Biomet, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
One lot of 75mm tibial tray is potentially etched and labeled as a 79 mm tibial tray. This may lead to delay in surgery, or immediate revision surgery to remove an incorrectly sized tibial component after it has been cemented.
FDA root cause category: Under Investigation by firm
What the company is doing
On May 18, 2018, the firm notified affected distributors and user facilities of the recall through Urgent Medical Device Recall letters sent via email and FedEx. The letters informed affected consignees of the issue and stated that the affected units were distributed in March 2018. Distributor Responsibilities: 1. Review this notification and ensure that affected team members are aware of the contents. 2. Immediately locate and quarantine affected product in your inventory. 3. Immediately return all affected product from your distributorship and from affected hospitals within your territory. a. Complete Attachment 1 Inventory Return Certification Form and send to [email protected] within three (3) days. b. For each return, send a copy of Attachment 1 to [email protected]. c. Include a hardcopy of Attachment 1 in each carton of your return shipment for immediate processing. d. Mark RECALL on the outside of the returned cartons. 4. Return the Additional Accounts form to [email protected]. a. Review the list of hospitals and/or surgeons included with the email notification sent to your facility, which includes a list of hospitals and/or surgeons that have already been notified of this recall. b. Identify whether there are any additional hospitals and/or surgeons that Zimmer Biomet has not notified and list these accounts on the Additional Accounts form. Please provide the form in Excel format. c. If there are no additional accounts or surgeons to notify, please indicate that there are no additional accounts, or indicate None or NA on the form. Risk Manager Responsibilities: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected
Recall details
- Recall number
- Z-2177-2018
- Status
- Terminated
- Initiated by the company
- May 18, 2018
- Posted by the FDA
- Jun 1, 2018
- Recalling firm
- Zimmer Biomet, Inc.
- Quantity
- 16
- Product code
- JWH
- FDA submissions
- K010212, K142933, K171054, K915132
- Distribution
- US Distribution to the states of : FL, GA, IN, MA, NM, UT, and WI. Foreign distribution to Canada and Mexico
Product description
Biomet Modular Primary Tibial Tray Implants; Modular Interlok Primary Tray (size 75 mm); Item No. 141215; UDI: (01) 00880304005310 (17) 280306 (10) 385340 Product is used in Total Knee Arthroplasty
Code information (models, lots, serial numbers)
Lot 385340 UDI: (01) 00880304005310 (17) 280306 (10) 385340
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- BIOMET KNEE SYSTEM 141215Biomet Orthopedics, LLC · Prosthesis, knee, patello/femorotibial, semi-constrained, uhmwpe, pegged, cemented, polymer/metal/polymer1 recall3 reports
Product lines
Other devices under the same FDA submission
213 devices cleared under K010212, K142933, K171054, K915132 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- VANGUARD KNEE SYSTEM 184764Biomet Orthopedics, LLC · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER2 recalls5 reports
- Vanguard® Knee System 183108Biomet Orthopedics, LLC · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER2 recalls3 reports
- Vanguard® Knee System 183126Biomet Orthopedics, LLC · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER2 recalls3 reports
- Vanguard® Knee System 189082Biomet Orthopedics, LLC · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER2 recalls2 reports
- Vanguard® Knee System 183620Biomet Orthopedics, LLC · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 recall5 reports
- Vanguard® Knee System 183106Biomet Orthopedics, LLC · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 recall4 reports
- Vanguard® Knee System 183104Biomet Orthopedics, LLC · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 recall3 reports
- Vanguard® Knee System 183122Biomet Orthopedics, LLC · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 recall3 reports
- VANGUARD KNEE SYSTEM 184762Biomet Orthopedics, LLC · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 recall3 reports
- VANGUARD KNEE SYSTEM 184766Biomet Orthopedics, LLC · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 recall3 reports
- Vanguard® Knee System 183124Biomet Orthopedics, LLC · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 recall2 reports
- Vanguard® Knee System 183622Biomet Orthopedics, LLC · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 recall1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page