Affixus Hip Fracture Nail Left 125 9 mm x 380 mm, Item… recallZ-3084-2018 · Zimmer Biomet, Inc. · initiated Jul 17, 2018
Class II Terminated All recalls posted in September 2018
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Zimmer Biomet, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The set screw is not able to advance or reverse from the original position in the nail.
FDA root cause category: Under Investigation by firm
What the company is doing
On 7/17/18, Urgent Medical Device Recall letters were sent to Risk Managers and Distributors. The letters instructed customers to do the following: Risk Manager Responsibilities: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. Complete Attachment 1 Certificate of Acknowledgement and send to [email protected]. This form must be returned even if you do not have affected products at your facility. 4. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facilitys documentation. 5. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to [email protected]. Distributor Responsibilities: 1. Review this notification and ensure that affected team members are aware of the contents. 2. Immediately locate and quarantine affected product in your inventory. 3. Immediately return all affected product from your distributorship and from affected hospitals within your territory. a. Complete Attachment 1 Inventory Return Certification Form and send to [email protected] within three (3) days. b. Request a Return Authorization Number through Fast SMS or by emailing [email protected] c. For each return, send a copy of Attachment 1 to [email protected]. d. Include a hardcopy of Attachment 1 in each carton of
Recall details
- Recall number
- Z-3084-2018
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Jul 17, 2018
- Posted by the FDA
- Sep 13, 2018
- Recalling firm
- Zimmer Biomet, Inc.
- Quantity
- 5947 total
- Product code
- HSB
- FDA submissions
- K100238
- Distribution
- The products were distributed to the following US states: AL, AR, AZ, CA, CO, FL, GA, HI, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WY, and Puerto Rico. The products were distributed to the following foreign countries: Chile, Ecuador, Japan, Netherlands.
Product description
Affixus Hip Fracture Nail Left 125 9 mm x 380 mm, Item Number: 814409380
Code information (models, lots, serial numbers)
053790 155180 762720 790780 813530
Other devices under the same FDA submission
89 devices cleared under K100238 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- AFFIXUS®HIP FRACTURE NAIL 8145-11-180Biomet Orthopedics, LLC · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall2 reports
- AFFIXUS®HIP FRACTURE NAIL 8143-09-180Biomet Orthopedics, LLC · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall1 report
- AFFIXUS®HIP FRACTURE NAIL - RIGHT 8143-09-340Biomet Orthopedics, LLC · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall1 report
- AFFIXUS®HIP FRACTURE NAIL 8143-11-180Biomet Orthopedics, LLC · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall1 report
- AFFIXUS®HIP FRACTURE NAIL - LEFT 8144-09-440Biomet Orthopedics, LLC · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall1 report
- AFFIXUS®HIP FRACTURE NAIL - LEFT 8144-11-440Biomet Orthopedics, LLC · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall1 report
- AFFIXUS®HIP FRACTURE NAIL 8145-09-180Biomet Orthopedics, LLC · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall1 report
- AFFIXUS®HIP FRACTURE NAIL - RIGHT 8145-09-380Biomet Orthopedics, LLC · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall1 report
- AFFIXUS®HIP FRACTURE NAIL - RIGHT 8145-11-280Biomet Orthopedics, LLC · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall1 report
- AFFIXUS®HIP FRACTURE NAIL - RIGHT 8145-11-380Biomet Orthopedics, LLC · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall1 report
- AFFIXUS®HIP FRACTURE NAIL - LEFT 8146-11-340Biomet Orthopedics, LLC · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall1 report
- AFFIXUS®HIP FRACTURE NAIL - LEFT 8146-11-400Biomet Orthopedics, LLC · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 80755).
- Z-3046-2018 · TerminatedAffixus Hip Fracture Nail Right 125 9 mm x 280 mm, Item Number: 814309280
- Z-3047-2018 · TerminatedAffixus Hip Fracture Nail 125 9 mm x 180 mm, Item Number: 814309180
- Z-3048-2018 · TerminatedAffixus Hip Fracture Nail Right 125 9 mm x 300 mm, Item Number 814309300
- Z-3049-2018 · TerminatedAffixus Hip Fracture Nail Right 125 9 mm x 320 mm, Item Number: 814309320
- Z-3050-2018 · TerminatedAffixus Hip Fracture Nail Right 125 9 mm x 360 mm, Item Number: 814309360
- Z-3051-2018 · TerminatedAffixus Hip Fracture Nail Right 125 9 mm x 380 mm, Item Number: 814309380
- Z-3052-2018 · TerminatedAffixus Hip Fracture Nail Right 125 9 mm x 420 mm, Item Number: 814309420
- Z-3053-2018 · TerminatedAffixus Hip Fracture Nail Right 125 9 mm x 440 mm, Item Number: 814309440
- Z-3054-2018 · TerminatedAffixus Hip Fracture Nail Right 125 9 mm x 460 mm, Item Number: 814309460
- Z-3055-2018 · TerminatedAffixus Hip Fracture Nail 125 11 mm x 180 mm, Item Number: 814311180
- Z-3056-2018 · TerminatedAffixus Hip Fracture Nail Right 125 11 mm x 280 mm, Item Number: 814311280
- Z-3057-2018 · TerminatedAffixus Hip Fracture Nail Right 125 11 mm x 300 mm, Item Number: 814311300
- Z-3058-2018 · TerminatedAffixus Hip Fracture Nail Right 125 11 mm x 320 mm, Item Number: 814311320
- Z-3059-2018 · TerminatedAffixus Hip Fracture Nail Right 125 11 mm x 340 mm, Item Number: 814311340
- Z-3060-2018 · TerminatedAffixus Hip Fracture Nail Right 125 11 mm x 360 mm, Item Number: 814311360
- Z-3061-2018 · TerminatedAffixus Hip Fracture Nail Right 125 11 mm x 380 mm, Item Number: 814311380
- Z-3062-2018 · TerminatedAffixus Hip Fracture Nail Right 125 11 mm x 400 mm, Item Number: 814311400
- Z-3063-2018 · TerminatedAffixus Hip Fracture Nail Right 125 11 mm x 420 mm, Item Number: 814311420
- Z-3064-2018 · TerminatedAffixus Hip Fracture Nail Right 125 11 mm x 440 mm, Item Number: 814311440
- Z-3065-2018 · TerminatedAffixus Hip Fracture Nail Right 125 11 mm x 460 mm, Item Number: 814311460
- Z-3066-2018 · TerminatedAffixus Hip Fracture Nail Right 125 13 mm x 180 mm, Item Number: 814313180
- Z-3067-2018 · TerminatedAffixus Hip Fracture Nail Right 125 13 mm x 260 mm, Item Number: 814313260
- Z-3068-2018 · TerminatedAffixus Hip Fracture Nail Right 125 13 mm x 280 mm, Item Number: 814313280
- Z-3069-2018 · TerminatedAffixus Hip Fracture Nail Right 125 13 mm x 300 mm, Item Number: 814313300
- Z-3070-2018 · TerminatedAffixus Hip Fracture Nail Right 125 13 mm x 320 mm, Item Number: 814313320
- Z-3071-2018 · TerminatedAffixus Hip Fracture Nail Right 125 13 mm x 340 mm, Item Number: 814313340
- Z-3072-2018 · TerminatedAffixus Hip Fracture Nail Right 125 13 mm x 360 mm, Item Number: 814313360
- Z-3073-2018 · TerminatedAffixus Hip Fracture Nail Right 125 13 mm x 380 mm, Item Number: 814313380
- Z-3074-2018 · TerminatedAffixus Hip Fracture Nail Right 125 13 mm x 400 mm, Item Number: 814313400
- Z-3075-2018 · TerminatedAffixus Hip Fracture Nail Right 125 13 mm x 420 mm,Item Number: 814313420
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page