Skip to content

Maquet VOLISTA StandOP Surgical Lights recallZ-1859-2019 · Maquet Cardiovascular Us Sales, Llc · initiated May 28, 2019

Class II Terminated All recalls posted in June 2019

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Maquet Cardiovascular Us Sales, Llc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential that the coating of the light head forks on the VOLISTA Surgical Lights may crack, posing the risk of particles falling into the surgical field during surgery that may result in patient infection

FDA root cause category: Device Design

What the company is doing

Getinge/Maquet SAS issued on 5/28/19 Urgent Medical Correction letter via FedEx 2 day Delivery with Signature Proof of Delivery (SPOD). Letter provides identification of issue, health risk and action to take: Before using the VOLISTA StandOP, please inspect the forks and check if cracks on the coating are visible as instructed in the VOLISTA StandOP manual; If a coating crack is detected, please STOP USING the VOLISTA StandOP and contact your Getinge representative. Complete and sign the attached MEDICAL DEVICE CORRECTION RESPONSE FORM . Return the completed form to Maquet/Getinge by e-mailing a scanned copy to [email protected] or by faxing the form to 1-707-202-7275;A Getinge Sales or Service representative will contact you to schedule inspection and correction if necessary of the VOLISTA StandOP SURGICAL LIGHTS at your facility. For technical questions, contact Technical Support Department at 1-888-627-8383 (press option 3 then option 1), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. EST. ***Updated 3/30/2020*** On March 30, 2020, the firm sent an new letter to customers, informing them of an update to the corrective actions previously communicated. The firm has developed an interim solution for affected devices. The solution consists of a validated adhesive strip that is resistant to mechanical stresses and compatible with typical chemical agents used in the operating rooms. The solution allows customers to continue using their lights until their defective forks are replaced. The interim solution will be applied by a Getinge representative. A Getinge Sales or Service representative will contact customers to schedule inspection and correction (if necessary) of affected devices at each facility. If your site has already been contacted to schedule an inspection, the work above will be performed during that scheduled visit.

Recall details

Recall number
Z-1859-2019
Recall class
Class II
Status
Terminated
Initiated by the company
May 28, 2019
Posted by the FDA
Jun 21, 2019
Recalling firm
Maquet Cardiovascular Us Sales, Llc
Quantity
12 units
Product code
FTD
FDA submissions
K130513
Distribution
Worldwide distribution. US Nationwide, ALBANIA GHANA PORTUGAL ARGENTINA HONG KONG ROMANIA AUSTRIA HUNGARY RUSSIA AUSTRALIA INDIA SAUDI ARABIA AZERBAIJAN IRAN SINGAPORE BAHAMAS IRELAND SLOVAKIA BANGLADESH IRAQ SLOVENIA BELGIUM ISRAEL SOUTH AFRICA BOLIVIA ITALY SOUTH KOREA BRAZIL JAPAN SPAIN BULGARIA JORDAN SWEDEN CAMEROON LATVIA SWITZERLAND CANADA LEBANON TAIWAN CHILE LUXEMBOURG TANZANIA CHINA MALAYSIA THAILAND COLOMBIA MEXICO TRINIDAD and TOBAGO COSTA RICA MOLDAVA TURKEY CROATIA MOZAMBIQUE UNITED ARAB EMIRATES CUBA MYANMAR UNITED KINGDOM CZECH REPUBLIC NETHERLANDS VIETNAM DENMARK NAMIBIA VENEZUELA DOMINICAN REPUBLIC NIGERIA YEMEN ECUADOR NORWAY ZIMBABWE EGYPT NEW ZEALAND ESTONIA OMAN ESWATINI PAKISTAN FINLAND PANAMA FRANCE PARAGUAY GERMANY POLAND

Product description

Maquet VOLISTA StandOP Surgical Lights- Product Code: ARDVST229002A Model: VST66SF AIM

Code information (models, lots, serial numbers)

Serial Numbers:  500005 500006 500007 500008 500009 500020 500021 500022 500025 500026 500027 500028

Other devices under the same FDA submission

53 devices cleared under K130513 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 82978).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page