Skip to content

Centricity recallZ-1910-2019 · GE Healthcare, LLC · initiated May 16, 2019

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from GE Healthcare, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

There is the possibility of viewing studies directly from the Enterprise Archive or VNA with incorrect patient images because the updated series or study is not present in the archive.

FDA root cause category: Device Design

What the company is doing

On May 20, 2019, the firm, GE Healthcare, sent an " URGENT MEDICAL DEVICE CORRECTION" letter date 5/20/2019 to their consignees. The letter described the product, problem and actions to be taken. The customers were instructed to do the following: discontinue use of the UV study management functionality for study split/study info updates until a correction is available and to wait for a GE representative to contact them to arrange for the correction. If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.

Recall details

Recall number
Z-1910-2019
Status
Terminated
Initiated by the company
May 16, 2019
Posted by the FDA
Jul 1, 2019
Recalling firm
GE Healthcare, LLC
Quantity
145 (87 US 103 OUS)
Product code
LLZ
FDA submissions
K150420, K182419
Distribution
Worldwide Distribution: US (nationwide) to states of: FL, GA, IL, IN, LA, MA, MO, NJ, NY, OR, PA, SC, SD, TX, and WA; and countries of : Brazil, Canada, Chile, France, French Polynesia, Germany, India, Italy, Japan, Kuwait, Lebanon, Singapore, Spain and United Kingdom.

Product description

Centricity Universal Viewer 6.0 Is a device that displays medical images and data from various imaging sources, and from other healthcare information sources.

Code information (models, lots, serial numbers)

UDI # 840682103800  Model # 2098249-002 Software Version: 6.0.9, 6.0SP9.01, CPACS 6.0 SP9.0.1, UV 6.0 SP9.0.1, UV 6.0 SP9.01.1, UV 6.0 SP9.0.1.2, UV 6.0 SP9.01.3, UV6.0 SP9.01.4, UV 6.0 SP9.01.5, UV 6.0 SP9.0.1.6, UV Version 6.0 SP5.0.4.1, UV Version 6.0 SP9, UV Version 6.0 SP9.0.0.1, UV Version 6.0 SP901, UV Version 6.0 SP9011, and UV Version 6.0 SP902

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 82998).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page