Alinity recallZ-2208-2019 · Abbott Ireland Diagnostics Division · initiated Nov 21, 2018

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Abbott Ireland Diagnostics Division or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Abbott has become aware of a manufacturing issue that can result in damage to a bottle neck in some of the reagent cartridges found in the lots listed in Appendix A. If a bottle neck is damaged, it may cause a failure during the reagent loading process on the Alinity i instrument.

FDA root cause category: Device Design

What the company is doing

The firm, Abbott, initiated the recall by a "Product Recall Urgent - Immediate Action Required" letter dated 11/21/2018 to its consignees. The letter explained the product, problem and provided instruction on how to inspect the reagent cartridges for defect. The consignee was directed to discard the defective units, use an alternate lot of the product if available; manually inspect each reagent cartridge in inventory before use; complete and return the Customer Reply Form via FAX #: 18007770051 or email: [email protected]. Even if you no longer have the instrument(s), please return the completed Customer Reply Form so that we can appropriately update our customer database; If you have forwarded the product to other laboratories, please inform them of this Product Recall and provide to them a copy of this letter; and retain this letter for your laboratory records. The units will be replaced. If you or any of the health care providers you serve have any questions regarding this information, U.S. Customers please contact Customer Service at 18774ABBOTT (available 24 hours a day, 7 days a week). Customers outside the U.S., please contact your local area Customer Service.

Recall details

Recall number
Z-2208-2019
Status
Terminated
Initiated by the company
Nov 21, 2018
Recalling firm
Abbott Ireland Diagnostics Division
Quantity
216 units
Product code
DBF
FDA submissions
K983759
Distribution
US Distribution to states of: AR, FL, GA, MD, OK and VA.

Product description

Alinity i Ferritin Reagent Kit, List Number 07P6530 The Alinity i Ferritin assay is a chemiluminescent microparticle immunoassay (CMIA) used for the quantitative determination of ferritin in human serum and plasma on the Alinity i analyzer.

Code information (models, lots, serial numbers)

UDI/ Lot Numbers: (01)00380740131487 (17)190730(10)91532UI00/91532UI00, (01)00380740131487 (17)190905(10)93238UI00/93238UI00